Shiley 70XLTINRecalls & reported problems
Covidien LP · Inner Cannula XLT,Disposable · Class 2
All models of Shiley (387)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 1
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 60884522006731), received between Jul 24, 2019 and Aug 15, 2025 — classified as 1 death, 2 injuries, 6 malfunctions.
The most frequently reported specific device problem is “Use of Device Problem” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Use of Device Problem5
- Connection Problem1
- Crack1
- Defective Component1
- Insufficient Device Problem Information1
- Material Separation1
- Mechanical Problem1
Patient problems
- Extubate2
- Insufficient Information2
- No Clinical Signs, Symptoms or Conditions2
- Unintended Extubation2
- Death1
- Low Oxygen Saturation1
- Respiratory Failure1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 15, 2025Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT THE PATIENT'S SPOUSE ATTEMPTED TO CHANGE THE INNER CANNULA WHEN THE BLUE PART SNAPPED OFF, ON TWO DIFFERENT INNER CANNULAS WITH THE SAME LOT NUMBER. ADDITIONAL INNER CANNULAS WITH A DIFFERENT LOT NUMBER WERE AVAILABLE AND WAS USED TO CHANGE THE INNER CANNULA. THERE WAS NO REPORTED PATIENT OUTCOME.
Patient problems: Insufficient Information
MAUDE report 22819615 ↗MalfunctionAug 15, 2025Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT THE PATIENT'S SPOUSE ATTEMPTED TO CHANGE THE INNER CANNULA WHEN THE BLUE PART SNAPPED OFF, ON TWO DIFFERENT INNER CANNULAS WITH THE SAME LOT NUMBER. ADDITIONAL INNER CANNULAS WITH A DIFFERENT LOT NUMBER WERE AVAILABLE AND WAS USED TO CHANGE THE INNER CANNULA. THERE WAS NO REPORTED PATIENT OUTCOME.
Patient problems: Insufficient Information
MAUDE report 22819601 ↗MalfunctionJan 17, 2025Crack; Material Separation; Insufficient Device Problem Information; Crack; Material Separation; Insufficient Device Problem Information
IT WAS REPORTED THAT DURING AN INNER CANNULA CHANGE FOR A PATIENT WITH A TRACHEOSTOMY WHO WAS INVASIVELY VENTILATED, THE BLUE LOCKING MECHANISM ON A SIZE 70 XLTIN INNER CANNULA WAS FOUND TO BE BROKEN AND CHIPPED IN TWO PLACES. THERE WAS NO PATIENT SYMPTOMS OR COMPLICATIONS ASSOCIATED WITH THIS EVENT. THE PRODUCT WAS IMMEDIATELY REMOVED FROM SERVICE, AND THE SAME SIZE INNER CANNULA WAS USED WITHOUT FURTHER DEFICIENCIES NOTED.
Patient problems: Unintended Extubation
MAUDE report 21176336 ↗MalfunctionFeb 6, 2023Use of Device Problem; Use of Device Problem
ACCORDING TO THE REPORTER, DURING USE, THE DEVICE INNER CANNULA HAD A DENT. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16303675 ↗MalfunctionFeb 1, 2023Use of Device Problem; Use of Device Problem
ACCORDING TO THE REPORTER, DURING USE, THE DEVICE INNER CANNULA HAD A DENT. THERE WAS NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16277322 ↗MalfunctionNov 10, 2021Defective Component; Defective Component
ACCORDING TO THE REPORTER, DURING USE, THE DEVICE HAD FAILED IN A VERY ILL PATIENT. THE PHYSICIAN WAS UNABLE TO VENTILATE THE PATIENT AND SATURATION LEVELS DROPPED. THE PATIENT PASSED AWAY.
Patient problems: Unintended Extubation
MAUDE report 12791907 ↗InjuryFeb 20, 2020Connection Problem; Connection Problem
ACCORDING TO THE REPORTER, DURING USE, HEALTH CARE PROVIDER TOOK THE NEW INNER CANNULA OUT OF PACKAGE TO REPLACE THE PATIENT'S INNER CANNULA. AS THEY WERE INSERTING THE NEW INNER CANNULA INTO THE TRACH THE FLANGE PART OF THE INNER CANNULA SEPARATED FROM THE WHITE PART OF THE INNER CANNULA. THE PATIENT WAS VENTILATED ON PRESSURE CONTROL 18/5, RATE 18, 30%02. PATIENT EXPERIENCED DESATURATION AND SHORTNESS OF BREATH AS THEY WERE DISCONNECTED FROM THE VENTILATOR AND HCP WAS UNABLE TO CONNECT PATIENT TO VENTILATOR UNTIL A NEW INNER CANNULA WAS OBTAINED.
Patient problems: Extubate; Low Oxygen Saturation
MAUDE report 9733707 ↗InjuryOct 16, 2019Use of Device Problem; Use of Device Problem
ACCORDING TO THE REPORTER, THE DEVICE'S BLUE PIECE WAS DIFFICULT TO SECURE THE INNER CANNULA AND CAME TOO LOOSE FROM THE TRACHEOSTOMY. INNER CANNULA WAS ATTACHED BUT COULDN'T CLICK IT INTO PLACE AND VENT TUBING WAS ATTACHED. THE PATIENT HAD CHRONIC RESPIRATORY FAILURE.
Patient problems: Respiratory Failure
MAUDE report 9197786 ↗DeathJul 24, 2019Mechanical Problem; Mechanical Problem
ACCORDING TO THE REPORTER, THE INNER CANNULA WAS NOT SEATED ON THE TRACH AND UPON EXCHANGE THE REPORTER SAW THAT IT WAS CRUSHED. PATIENT CEASED TO BREATHE.
Patient problems: Death; Extubate
MAUDE report 8822385 ↗
Other Tube tracheostomy and tube cuff devices
- Shiley 8CN85HCovidien LP · Tube tracheostomy and tube cuff1 recall2 deaths354 reports
- Shiley 6CN75HCovidien LP · Tube tracheostomy and tube cuff1 recall3 deaths240 reports
- Shiley 7CN80RCovidien LP · Tube tracheostomy and tube cuff2 recalls148 reports
- Bivona 67PFSS35Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall166 reports
- Bivona 670180Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall140 reports
- Bivona 670170Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall10 deaths132 reports
- Shiley 6CN75RCovidien LP · Tube tracheostomy and tube cuff1 recall1 death121 reports
- Bivona 67PFSS40Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall115 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page