Shiley 8CFSRecalls & reported problems
Covidien LP · Tracheostomy Tube Cuffless with Inner Cannula · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 60884522006977), received between Jan 9, 2018 and Oct 20, 2021 — classified as 5 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Material Separation” (2 reports).
23 other recalls cover devices cleared under the same FDA submission (K962173); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Material Separation2
- Disconnection1
- Material Integrity Problem1
- Material Split, Cut or Torn1
- Obstruction of Flow1
- Use of Device Problem1
Patient problems
- Extubate3
- Injury2
- No Consequences Or Impact To Patient2
- Hypoxia1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 20, 2021Obstruction of Flow; Obstruction of Flow
ACCORDING TO THE REPORTER, THE COORDINATING NURSE, WHILE INSPECTING THE TUBE BEFORE INSERTION, NOTICED A JAMMING BLOCKAGE OF THE INNER CANNULA. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12666059 ↗InjuryJul 16, 2020Material Separation; Material Separation
ACCORDING TO THE REPORTER, DURING PROCEDURE, THE DEVICE HAD A BREAKAGE OF THE WHITE JUNCTION WITH INNER CANNULA THAT TURNED IDLE AND COULD NOT BE EXTRACTED. THE PATIENT HAD A MILD BLEEDING AND THE PROCEDURE WAS DELAYED 30 MINUTES LONGER THAN USUAL.
Patient problems: Injury
MAUDE report 10283287 ↗InjuryJul 16, 2020Material Separation; Material Separation
ACCORDING TO THE REPORTER, THE UNIT HAD A BREAKAGE OF INNER CANNULA WHICH, BY TURNING, COULD NOT LONGER BE EXTRACTED. IT WAS INDICATED THAT THE PROCEDURE WAS DELAYED BY 30 MINUTES LONGER THAN USUAL. THE CAUSE OF THE DELAY IN CARRYING OUT THE PROCEDURE WAS THE SUDDEN BLOCK IN THE EXTRACTION OF THE COUNTER CANNULA WHICH WAS REPLACED EVERY DAY, TWICE A DAY.
Patient problems: Injury; No Consequences Or Impact To Patient
MAUDE report 10282169 ↗InjuryMay 13, 2020Material Split, Cut or Torn; Material Split, Cut or Torn
ACCORDING TO THE REPORTER DURING USE, THE DEVICE HAD PHYSICAL DAMAGE. THE CANNULA WAS SUCCESSFULLY PLACED INTO THE PATIENT, HOWEVER DURING THE NIGHT, THE PATIENT SOMETIMES HAS TROUBLE GETTING AIR. THIS MADE THE PATIENT TEAR THE CANNULA OUT IN DESPERATION THREE TIMES AND AT THAT TIME THE PATIENT HAD DESTROYED THE CANNULA. AFTER THE FIRST TIME, THE CANNULA WAS DOUBLE TAPED ON, BUT THE PATIENT MANAGED TO RIP THE CANNULA OUT. AFTERWARDS, THE PATIENT CANNOT REMEMBER WHAT HAPPENED. THE FACT THAT THE CANNULA WAS RIPPED OUT, THE PATIENT HAS GIVEN SOME ACUTE SITUATIONS AND ONCE ENDED UP INTENSIVE AND WAS ABOUT TO GO INTO A CARDIAC ARREST. ADDITIONAL INFORMATION HAS BEEN REQUESTED.
Patient problems: Extubate; Hypoxia
MAUDE report 10050201 ↗MalfunctionSep 29, 2019Use of Device Problem; Use of Device Problem
ACCORDING TO THE REPORTER, DURING A ROUTINE TRACHEOSTOMY CHANGE, THE OUTER CANNULA WAS PLACED IN SITU BUT WHEN THE INNER CANNULA WAS INSERTED, THEY WERE UNABLE TO LOCK IT. THEY WERE ABLE TO SUCCESSFULLY LOCK AN ALTERNATE INNER CANNULA IN PLACE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9132618 ↗InjuryJan 30, 2019Disconnection; Disconnection
ACCORDING TO THE REPORTER, THE DEVICE'S FLANGE BROKE FROM THE OUTER CANNULA. IT WAS REPORTED THAT THE TRACH CHANGE WAS DONE AND AFTER 2 DAYS THE FLANGE CAME APART FROM THE REST OF THE TRACH AND THEY WERE ABLE TO REMOVE PORTION OF TRACH THAT REMAINED IN PATIENT AND INSERT SPARE.
Patient problems: Extubate
MAUDE report 8288552 ↗InjuryJan 9, 2018Material Integrity Problem; Material Integrity Problem
ACCORDING TO THE REPORTER, DURING USE OF THE DEVICE, THE INNER CANNULA DID NOT REMAIN LOCKED ON THE OUTER CANNULA CAUSING LEAKAGE IN THE VENTILATION. ANOTHER CANNULA WAS USED. NO REPORTED PATIENT OUTCOME.
Patient problems: Extubate
MAUDE report 7176924 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K962173. One submission often covers many models, so check the product and lot details in each record.
- Z-2174-2012 · Jul 18, 2012 · TerminatedShiley Reusable Cannula Low Pressure Cuffed Tracheostomy Tubes, model numbers (Worldwide) 8LPC, 8FEN; (Outside U.S.) 8FENJ, 8FENJ-S, 8LPC-S, 8FEN-S. Product Usage: The product
- Z-1479-2010 · Apr 13, 2010 · TerminatedShiley Low Pressure Cuffed Tracheostomy Tube, 6 LPC, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of upper
- Z-1476-2010 · Apr 13, 2010 · TerminatedShiley Fenestrated Low Pressure Cuffed Tracheostomy Tube with Evacuation Line, 6 FEN-S, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton,
- Z-1467-2010 · Apr 13, 2010 · TerminatedShiley Fenestrated Low Pressure Cuffed Tracheostomy Tube, 6 FEN, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypa
- Z-1481-2010 · Apr 13, 2010 · TerminatedShiley Low Pressure Cuffed Tracheostomy Tube, 10 LPC, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of upper
- Z-1465-2010 · Apr 13, 2010 · TerminatedShiley Disposable Cannula Low Pressure Cuffed Tracheostomy Tube, 10 DCT, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended
- Z-1480-2010 · Apr 13, 2010 · TerminatedShiley Low Pressure Cuffed Tracheostomy Tube, 8 LPC, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of upper
- Z-1471-2010 · Apr 13, 2010 · TerminatedShiley Disposable Cannula Fenestrated Low Pressure Cuffed Tracheostomy Tube, 6 DFEN, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA
- Z-1477-2010 · Apr 13, 2010 · TerminatedShiley Fenestrated Low Pressure Cuffed Tracheostomy Tube with Evacuation Line, 8 FEN-S, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton,
- Z-1470-2010 · Apr 13, 2010 · TerminatedShiley Disposable Cannula Fenestrated Low Pressure Cuffed Tracheostomy Tube, 4 DFEN, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA
- Z-1490-2010 · Apr 13, 2010 · TerminatedShiley Low Pressure Cuffed Tracheostomy Tube with Evacuation Line, 8 LPC-S, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intend
- Z-1466-2010 · Apr 13, 2010 · TerminatedShiley Fenestrated Low Pressure Cuffed Tracheostomy Tube, 4 FEN, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypa
- Z-1478-2010 · Apr 13, 2010 · TerminatedShiley Low Pressure Cuffed Tracheostomy Tube, 4 LPC, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of upper
- Z-1472-2010 · Apr 13, 2010 · TerminatedShiley Disposable Cannula Fenestrated Low Pressure Cuffed Tracheostomy Tube, 8 DFEN, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA
- Z-1488-2010 · Apr 13, 2010 · TerminatedShiley Low Pressure Cuffed Tracheostomy Tube with Evacuation Line, 4 LPC-S, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intend
…and 8 more.
Other Tube tracheostomy and tube cuff devices
- Shiley 8CN85HCovidien LP · Tube tracheostomy and tube cuff1 recall2 deaths354 reports
- Shiley 6CN75HCovidien LP · Tube tracheostomy and tube cuff1 recall3 deaths240 reports
- Shiley 7CN80RCovidien LP · Tube tracheostomy and tube cuff2 recalls148 reports
- Bivona 67PFSS35Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall166 reports
- Bivona 670180Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall140 reports
- Bivona 670170Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall10 deaths132 reports
- Shiley 6CN75RCovidien LP · Tube tracheostomy and tube cuff1 recall1 death121 reports
- Bivona 67PFSS40Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall115 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page