Shiley 6FENRecalls & reported problems
Covidien LP · Tracheostomy Tube Cuffed with Inner Cannula,Fenestrated · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
Quick summary
This Covidien Shiley tracheostomy tube, a breathing tube placed through the neck, has 106 reports, no deaths and 19 injuries. Many reports describe the tube coming out of the airway unintentionally, and some describe air leaking from the cuff or parts missing from the kit. There is 1 recall, still open, started in April 2026 over incorrect labeling on the display box.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 106
- Deaths reported
- 0
- Injuries reported
- 19
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 106 adverse event reports that list this exact device (DI 60884522007028), received between Apr 22, 2017 and Jul 15, 2026 — classified as 19 injuries, 87 malfunctions.
The most frequently reported specific device problem is “Use of Device Problem” (55 reports).
1 FDA recall record name this device's identifier, 1 still open.
23 other recalls cover devices cleared under the same FDA submission (K962173); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Recalls naming this device
- Apr 15, 2026Open, Classified
Due to customer complaint regarding incorrect display box labeling.
Z-2261-2026 · Root cause: Process control · FDA record ↗
Most reported problems
Device problems
- Use of Device Problem55
- Gas/Air Leak23
- Component Missing8
- Missing Information7
- Connection Problem4
- Inadequacy of Device Shape and/or Size4
- Inflation Problem4
- Material Split, Cut or Torn4
Patient problems
- No Clinical Signs, Symptoms or Conditions41
- Unintended Extubation32
- Extubate17
- No Consequences Or Impact To Patient10
- No Patient Involvement10
- Cough1
- Dyspnea1
- Insufficient Information1
Recent adverse event reports
The 25 most recent of 106 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 15, 2026Use of Device Problem
Patient problems: Unintended Extubation
MAUDE report 25825412 ↗MalfunctionJul 13, 2026Material Split, Cut or Torn
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25804194 ↗MalfunctionJun 2, 2026Use of Device Problem
IT WAS REPORTED THAT DURING THE PRE-USAGE INSPECTION, THE TRACH TUBE'S CUFF HAD AN AIR LEAK. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25364892 ↗MalfunctionApr 28, 2026Use of Device Problem
IT WAS REPORTED THAT DURING A PROCEDURE INVOLVING A TRACH CANNULA, THE CANNULA WAS DIFFICULT TO TURN/ADJUST. A REPLACEMENT TRACH CANNULA WAS OPENED AND USED FOR THE PATIENT.
Patient problems: Unintended Extubation
MAUDE report 25023447 ↗MalfunctionApr 3, 2026Component Missing; Missing Information
IT WAS REPORTED THAT THE TRACHEOSTOMY TUBE LABELING HAD AN INCORRECT DESCRIPTION OF THE PRODUCT FROM THE TAG ATTACHED TO THE OTHER SIDE OF THE BOX. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24776214 ↗MalfunctionApr 1, 2026Component Missing; Missing Information
IT WAS REPORTED THAT THE TRACHEOSTOMY TUBE LABELING HAD AN INCORRECT DESCRIPTION OF THE PRODUCT FROM THE TAG ATTACHED TO THE OTHER SIDE OF THE BOX. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24760801 ↗MalfunctionApr 1, 2026Component Missing; Missing Information
IT WAS REPORTED THAT THE TRACHEOSTOMY TUBE LABELING HAD AN INCORRECT DESCRIPTION OF THE PRODUCT FROM THE TAG ATTACHED TO THE OTHER SIDE OF THE BOX. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24760780 ↗MalfunctionApr 1, 2026Component Missing; Missing Information
IT WAS REPORTED THAT THE TRACHEOSTOMY TUBE LABELING HAD AN INCORRECT DESCRIPTION OF THE PRODUCT FROM THE TAG ATTACHED TO THE OTHER SIDE OF THE BOX. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24760692 ↗MalfunctionApr 1, 2026Component Missing; Missing Information
IT WAS REPORTED THAT THE TRACHEOSTOMY TUBE LABELING HAD AN INCORRECT DESCRIPTION OF THE PRODUCT FROM THE TAG ATTACHED TO THE OTHER SIDE OF THE BOX. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24760678 ↗MalfunctionApr 1, 2026Component Missing; Missing Information
IT WAS REPORTED THAT THE TRACHEOSTOMY TUBE LABELING HAD AN INCORRECT DESCRIPTION OF THE PRODUCT FROM THE TAG ATTACHED TO THE OTHER SIDE OF THE BOX. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24760655 ↗MalfunctionApr 1, 2026Component Missing; Missing Information
IT WAS REPORTED THAT THE TRACHEOSTOMY TUBE LABELING HAD AN INCORRECT DESCRIPTION OF THE PRODUCT FROM THE TAG ATTACHED TO THE OTHER SIDE OF THE BOX. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24760635 ↗MalfunctionOct 23, 2025Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT THE INNER CANNULAS OF THE NINE TRACHEOSTOMY TUBES HAD AN UNEVEN END. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23376181 ↗MalfunctionOct 23, 2025Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT THE INNER CANNULAS OF THE NINE TRACHEOSTOMY TUBES HAD AN UNEVEN END. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23376177 ↗MalfunctionOct 23, 2025Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT THE INNER CANNULAS OF THE NINE TRACHEOSTOMY TUBES HAD AN UNEVEN END. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23376170 ↗MalfunctionOct 23, 2025Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT THE INNER CANNULAS OF THE NINE TRACHEOSTOMY TUBES HAD AN UNEVEN END. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23376165 ↗MalfunctionOct 23, 2025Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT THE INNER CANNULAS OF THE NINE TRACHEOSTOMY TUBES HAD AN UNEVEN END. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23376162 ↗MalfunctionOct 23, 2025Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT THE INNER CANNULAS OF THE NINE TRACHEOSTOMY TUBES HAD AN UNEVEN END. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23376156 ↗MalfunctionOct 23, 2025Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT THE INNER CANNULAS OF THE NINE TRACHEOSTOMY TUBES HAD AN UNEVEN END. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23376148 ↗MalfunctionOct 23, 2025Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT THE INNER CANNULAS OF THE NINE TRACHEOSTOMY TUBES HAD AN UNEVEN END. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23376136 ↗MalfunctionOct 23, 2025Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT THE INNER CANNULAS OF THE NINE TRACHEOSTOMY TUBES HAD AN UNEVEN END. THERE WAS NO PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23376127 ↗MalfunctionSep 9, 2025Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT THE TRACHEOSTOMY TUBE CUFF WAS REACCUMULATING AIR WHILE PLUGGED DESPITE CUFF BEING DEFLATED. THE TUBE WAS REMOVED ON THE PATIENT.
Patient problems: Unintended Extubation
MAUDE report 23001118 ↗MalfunctionAug 14, 2025Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT THE TRACHEOSTOMY TUBE'S PILOT VALVE MALFUNCTIONED CAUSED AN AIR LEAK ON THE FIRST DAY OF PATIENT USED. THE TUBE WAS REPLACED.
Patient problems: Unintended Extubation
MAUDE report 22805931 ↗MalfunctionAug 7, 2025Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT DURING USE, THE PILOT BALLOON OF THE TRACHEOSTOMY TUBE DID NOT INFLATE. THE PHYSICIAN ATTEMPTED TO INFLATE THREE TIMES WITHOUT SUCCESS. IT WAS UNKNOWN IF THERE WAS ANY PATIENT INVOLVEMENT OR COMPLICATIONS AS A RESULT OF THE EVENT.
Patient problems: Insufficient Information
MAUDE report 22745762 ↗InjuryJan 10, 2025Use of Device Problem; Gas/Air Leak; Use of Device Problem; Gas/Air Leak
IT WAS REPORTED THAT A CUFF FENESTRATED TRACHEOSTOMY CANNULA'S CUFF DID NOT INFLATE, CAUSING DISCOMFORT TO THE PATIENT WITH COUGH AND SHORTNESS OF BREATH, EVEN IF THE BARONET WAS ALWAYS INFLATED. THE TRACH WAS REPLACED.
Patient problems: Cough; Dyspnea; Unintended Extubation
MAUDE report 21132289 ↗MalfunctionNov 7, 2024Gas/Air Leak; Gas/Air Leak
ACCORDING TO THE REPORTER, INTRA-OPERATIVELY, THE TRACHEOSTOMY TUBE'S CUFF HAD A RAPID AIR LEAK RIGHT AFTER INTUBATION REPLACED THE SECOND AND THIRD TIMES AND HAD THE SAME ISSUE. A FOURTH TUBE WORKED REPLACING THOSE 3 TUBES.
Patient problems: Unintended Extubation
MAUDE report 20634859 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K962173. One submission often covers many models, so check the product and lot details in each record.
- Z-2174-2012 · Jul 18, 2012 · TerminatedShiley Reusable Cannula Low Pressure Cuffed Tracheostomy Tubes, model numbers (Worldwide) 8LPC, 8FEN; (Outside U.S.) 8FENJ, 8FENJ-S, 8LPC-S, 8FEN-S. Product Usage: The product
- Z-1488-2010 · Apr 13, 2010 · TerminatedShiley Low Pressure Cuffed Tracheostomy Tube with Evacuation Line, 4 LPC-S, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intend
- Z-1467-2010 · Apr 13, 2010 · TerminatedShiley Fenestrated Low Pressure Cuffed Tracheostomy Tube, 6 FEN, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypa
- Z-1462-2010 · Apr 13, 2010 · TerminatedShiley Disposable Cannula Low Pressure Cuffed Tracheostomy Tube, 4 DCT, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended f
- Z-1476-2010 · Apr 13, 2010 · TerminatedShiley Fenestrated Low Pressure Cuffed Tracheostomy Tube with Evacuation Line, 6 FEN-S, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton,
- Z-1480-2010 · Apr 13, 2010 · TerminatedShiley Low Pressure Cuffed Tracheostomy Tube, 8 LPC, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of upper
- Z-1469-2010 · Apr 13, 2010 · TerminatedShiley Fenestrated Low Pressure Cuffed Tracheostomy Tube, 10 FEN, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for byp
- Z-1489-2010 · Apr 13, 2010 · TerminatedShiley Low Pressure Cuffed Tracheostomy Tube with Evacuation Line, 6 LPC-S, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intend
- Z-1464-2010 · Apr 13, 2010 · TerminatedShiley Disposable Cannula Low Pressure Cuffed Tracheostomy Tube, 8 DCT, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended f
- Z-1465-2010 · Apr 13, 2010 · TerminatedShiley Disposable Cannula Low Pressure Cuffed Tracheostomy Tube, 10 DCT, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended
- Z-1470-2010 · Apr 13, 2010 · TerminatedShiley Disposable Cannula Fenestrated Low Pressure Cuffed Tracheostomy Tube, 4 DFEN, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA
- Z-1468-2010 · Apr 13, 2010 · TerminatedShiley Fenestrated Low Pressure Cuffed Tracheostomy Tube, 8 FEN, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypa
- Z-1490-2010 · Apr 13, 2010 · TerminatedShiley Low Pressure Cuffed Tracheostomy Tube with Evacuation Line, 8 LPC-S, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intend
- Z-1466-2010 · Apr 13, 2010 · TerminatedShiley Fenestrated Low Pressure Cuffed Tracheostomy Tube, 4 FEN, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypa
- Z-1479-2010 · Apr 13, 2010 · TerminatedShiley Low Pressure Cuffed Tracheostomy Tube, 6 LPC, Assembled in Mexico, Tyco Healthcare Group LP, Nellcor Puritan Bennett Division, Pleasanton, CA. Intended for bypass of upper
…and 8 more.
Other Tube tracheostomy and tube cuff devices
- Shiley 8CN85HCovidien LP · Tube tracheostomy and tube cuff1 recall2 deaths354 reports
- Shiley 6CN75HCovidien LP · Tube tracheostomy and tube cuff1 recall3 deaths240 reports
- Shiley 7CN80RCovidien LP · Tube tracheostomy and tube cuff2 recalls148 reports
- Bivona 67PFSS35Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall166 reports
- Bivona 670180Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall140 reports
- Bivona 670170Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall10 deaths132 reports
- Shiley 6CN75RCovidien LP · Tube tracheostomy and tube cuff1 recall1 death121 reports
- Bivona 67PFSS40Icu Medical, Inc. · Tube tracheostomy and tube cuff1 recall115 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.