IMPEDE Embolization Plug IMP-05 (UDI B054IMP051)Recalls & reported problems
Shape Memory Medical Inc. · IMPEDE Embolization Plug 5mm · Class 2
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI B054IMP051), received between Dec 30, 2021 and Dec 30, 2021 — classified as 1 malfunction.
The most frequently reported specific device problem is “Loss of or Failure to Bond” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Loss of or Failure to Bond1
Patient problems
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 30, 2021Loss of or Failure to Bond; Loss of or Failure to Bond
THIS IMP-05 WAS USED TO PERFORM A IIA EMBOLIZATION IN A COMMON ILIAC ARTERY ANEURYSM (CIAA). ACCESS TO THE TARGET SITE WAS THROUGH THE SUPERIOR GLUTEAL ARTERY USING A JUDKINS CATHETER 4FR (0.038 INCH INNER DIAMETER). THE PHYSICIAN USED A RADIFOCUS GUIDEWIRE (0.035 INCH OUTER DIAMETER) TO DELIVER THE IMP-05 THROUGH THE CATHETER. THE IMPEDE PROXIMAL MARKERBAND COULD NOT BE CONFIRMED IN THE CATHETER UNDER FLUOROSCOPY DURING DELIVERY, EVEN WHEN THE RADIFOCUS GUIDEWIRE WAS PULLED BACK FOR BETTER VISIBILITY. AFTER THE ANCHOR COIL PORTION OF THE IMP-05 WAS DEPLOYED, THE PHYSICIAN WAS THEN ABLE TO VISUALIZE THE IMPEDE PROXIMAL MARKERBAND UNDER FLUOROSCOPY, AND FOUND THAT THE FOAM HAD SEPARATED FROM THE ANCHOR COIL. THE PHYSICIAN THEN CHOSE TO DELIVER BOTH THE ANCHOR COIL AND THE FOAM, SEPARATE, INTO THE BLOOD VESSEL. ANOTHER COIL (NOT MANUFACTURED BY SMM) WAS THEN DELIVERED, COMPLETE EMBOLIZATION WAS ACHIEVED AND THE OPERATION WAS COMPLETED. THIS EVENT DID NOT RESULT IN ANY HARM TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13122570 ↗
Other Device, Vascular, For Promoting Embolization devices
- POD RBYPOD8-APenumbra, Inc. · Device, Vascular, For Promoting Embolization195 reports
- CEREPAKTM MDH1Cerenovus, Inc. · Device, Vascular, for Promoting Embolization1 recall72 reports
- Azur CX Detachable 45-750824Microvention, Inc. · Device, Vascular, for Promoting Embolization115 reports
- Tornado G08261Cook Incorporated · Device, Vascular, for Promoting Embolization2 recalls7 reports
- Optima Coil System OPTI0930CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Optima Coil System OPTI0720CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls4 reports
- Prestige Coil System PRES0520HPKPLBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
- Optima Coil System OPTI0413CSFMXBalt USA LLC · Device, Vascular, For Promoting Embolization2 recalls3 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.