Skip to content

Molded Products, Inc. MPC-130Recalls & reported problems

Molded Products, Inc. · See Luer Cap Set · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI B144MPC1300), received between Feb 11, 2026 and Jun 17, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Device Misassembled During Manufacturing /Shipping” (2 reports).

2 other recalls cover devices cleared under the same FDA submission (K925927); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Device Misassembled During Manufacturing /Shipping2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 17, 2026Device Misassembled During Manufacturing /Shipping

    AN ORIGINAL REPORT WAS CREATED ON (B)(6) 2026; EVENT # (B)(4); A PRODUCT RECALL WAS INITIATED AT THAT TIME. MOLDED PRODUCTS WAS ADVISED (B)(6) 2026 THAT A NEW EVENT HAD OCCURRED THAT INCLUDED THE SAME PRODUCT MANUFACTURED BY MOLDED PRODUCTS. MPC-130 SEE LUER CAP SET LOT# 20389 WAS DEEMED NONCONFORMING DUE TO BEING ASSEMBLED INCORRECTLY. THE LUER PIN WAS NOT SNAPPED INTO THE THREAD CAP WHICH WOULD CAUSE THE TWO PIECES TO DISENGAGE. THIS NONCONFORMANCE IS EASILY SPOTTED AS THE LUER PIN WILL COME OUT AS SOON AS THE CAP IS SCREWED ONTO THE CATHETER OR IT WILL ALREADY BE DISASSEMBLED IN THE PACKAGE. A PROCEDURALIST AT A RADIOLOGY SUITE DID NOT NOTICE THE NONCONFORMANCE AND PROCEEDED TO USE THE PRODUCT ON A PATIENT. THE PRODUCT SHOULD NOT HAVE BEEN AVAILABLE FOR USE AS IT HAD BEEN DOCUMENTED FOR RECALL, AND ALL DISTRIBUTORS HAD BEEN CONTACTED WITH THE RECALL INFORMATION. THE PATIENT HAD TO HAVE AN ADDITIONAL PROCEDURE DONE TO REPLACE A POTENTIALLY CONTAMINATED DIALYSIS LINE NO BLOOD LOSS NOTED AND NO REPORTED INJURY NOTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25535537 ↗
  • MalfunctionFeb 11, 2026Device Misassembled During Manufacturing /Shipping

    CLINIC PERSONNEL NOTED ON (B)(6) 2025 WHILE REMOVING THE SEE LUER CAP SET MPC-130 FROM ITS BAG, THAT THE THREADED SLEEVE WAS NOT ENGAGED WITH THE LUER PIN.THE HEALTH WORKER SET ASIDE THE PART AND OBTAINED A NEW PART. NO DIRECT PATIENT INVOLVEMENT WAS NOTED. THIS EVENT WAS REPORTED TO MOLDED PRODUCTS 12/08/2025 AN INVESTIGATION WAS IMMEDIATELY OPENED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24318216 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K925927. One submission often covers many models, so check the product and lot details in each record.

  • Z-1774-2026 · Feb 17, 2026 · Open, Classified
    See Luer Cap Set, MPC-130, set, administration, intravascular
  • Z-2354-2019 · Jun 10, 2019 · Terminated
    Luer Lock Set - Catalog # MPC-125, for IV administration sets.