Skip to content

CD45 MMab Concentrate, 1 mlRecalls & reported problems

Bio Sb, Inc. · CD45 is a mouse monoclonal antibody derived from cell culture supernatant that is concentrated, dialyzed, filter sterilized and diluted in buffer pH 7.5 · Class 1

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI B502BSB52510), received between May 19, 2025 and May 19, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 19, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    A CUSTOMER COMPLAINT WAS RECEIVED REGARDING CD45 (2B11 & PD7/26) ANTIBODY COCKTAIL PERFORMANCE IN IHC STAINING. INTERNAL INVESTIGATION REVEALED THAT CERTAIN LOTS WERE MISSING ONE OF THE LABELED CLONES, WHICH MAY LEAD TO FALSE-NEGATIVE RESULTS. THE PRODUCT LABELING INDICATES THE PRESENCE OF BOTH CLONES 2B11 AND PD7/26, BUT ONLY ONE WAS CONFIRMED IN THE AFFECTED LOTS. ALTHOUGH NO ADVERSE HEALTH CONSEQUENCES WERE REPORTED, THE POTENTIAL FOR DIAGNOSTIC MISINTERPRETATION EXISTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22041562 ↗

Other Immunohistochemistry Reagents And Kits devices

All Immunohistochemistry Reagents And Kits devices →