Skip to content

NAMIC 91051602Recalls & reported problems

Navilyst Medical, Inc. · High Pressure Contrast Injection Line · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI H749910516020), received between Aug 8, 2019 and Aug 8, 2019 — classified as 1 injury.

The most frequently reported specific device problem is “Improper Flow or Infusion” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Improper Flow or Infusion1

Patient problems

  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 8, 2019Improper Flow or Infusion; Improper Flow or Infusion

    AS REPORTED BY ANGIODYNAMICS' DISTRIBUTOR IN (B)(4), END USER HOSPITAL WAS UTILIZING A 60" FLEXCIL CONTRAST INJECTION LINE (WITH F-M) FITTINGS ALONG WITH AN EDWARD'S EXTENSION TUBE WHICH WAS CONNECTED TO A NAMIC STOPCOCK. "AS PREPARATION, THE FIRST AIR INSIDE INJECTOR WAS REMOVED AND THE TUBING CONNECTIONS WERE SECURELY TIGHTENED. PRIOR PROCEDURE, CONTRAST MEDIA WAS FLASHED INTO A 10ML SYRINGE OF THE OPERATOR SIDE. THEN ANGIOGRAPHY PROCEDURE BEGAN BY AN INJECTOR (5ML INJECTED, 2ML/SEC). IMMEDIATELY AIR WAS OBSERVED IN THE CATHETER ON THE IMAGE. THEREFORE, THE OPERATOR STOPPED INJECTING THE CONTRAST MEDIA IMMEDIATELY AND STOPPED THE ANGIOGRAPHY PROCEDURE." NO PATIENT INJURY OR COMPLICATIONS WERE REPORTED. THE USED DEVICES WERE DISCARDED AT THE HOSPITAL.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8875673 ↗

Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices

All CATHETER, INTRAVASCULAR, DIAGNOSTIC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page