AngioDynamics 06597013Recalls & reported problems
Angiodynamics, Inc. · Standard Micro Introducer Kit · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI H787065970130), received between Mar 30, 2023 and Mar 30, 2023 — classified as 1 malfunction.
The most frequently reported specific device problem is “Fracture” (1 report).
9 other recalls cover devices cleared under the same FDA submission (K051655); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Fracture1
- Material Frayed1
Patient problems
- Device Embedded In Tissue or Plaque1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 30, 2023Fracture; Material Frayed; Fracture; Material Frayed
A GUIDEWIRE FROM A STD 4F M.I. KIT 45CM SS/T ECHO S KIT WAS BEING USED FOR A CT GUIDED RIGHT SIDE CHEST TUBE PLACEMENT. A CT GUIDED CHEST TUBE PLACEMENT INTO THE RIGHT PLEURAL SPACE WAS BEING PERFORMED. WHEN ATTEMPTING TO REMOVE THE 0.018 MICROWIRE THROUGH THE MICROPUNCTURE SHEATH, TENSION WAS FELT. THE PROVIDER ATTEMPTED TO REMOVE THE ENTIRE SETUP AS ONE UNIT DUE TO THE TENSION. AT THIS TIME, THE TIP OF THE GUIDEWIRE UNFURLED, AND THE DISTAL PORTION FRACTURED OFF INSIDE THE PLEURAL SPACE. DURING WITHDRAWAL OF THE WIRE, THE TIP OF THE WIRE DETACHED INSIDE THE PATIENT. ATTEMPTS WERE MADE TO REMOVE THE FRAGMENT, HOWEVER THEY WERE UNSUCCESSFUL. THE FRAGMENT REMAINS INSIDE THE PATIENT AT THIS TIME.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 16645608 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K051655. One submission often covers many models, so check the product and lot details in each record.
- Z-0418-2021 · Oct 14, 2020 · TerminatedAngiodynamics 5F Standard Micro-Introducer Kit - indicated for percutaneous introduction of a guidewire or catheter into the vascular system, Catalog Number: 06597035 UPN: H7870
- Z-0776-2017 · Sep 28, 2016 · TerminatedANGIODYNAMICS Percutaneous Introducer Set under the following labels: 1) 4 F Stiffened Micro-Introducer Kit, 45CM SS/T NON-ECHO S PG, Item Number H787065970655, Catalog No.REF 0659
- Z-0773-2017 · Sep 28, 2016 · TerminatedANGIODYNAMICS Percutaneous Introducer Set under the following labels: 1) 4 F Stiffened Micro-Introducer Kit, 45CM SS/T ECHO B PG, Item Number H787065970465, Catalog No.REF 06597046
- Z-0774-2017 · Sep 28, 2016 · TerminatedANGIODYNAMICS Percutaneous Introducer Set under the following labels: 1) 4 F Stiffened Micro-Introducer Kit, 45CM NT/T ECHO B PG, Item Number H787065970505, Catalog No.REF 06597050
- Z-0778-2017 · Sep 28, 2016 · TerminatedANGIODYNAMICS 4F Stiffened Micro-Introducer, Item Number H787065971035, Catalog No.REF 06597103 Indicated for percutaneous introduction of guidewire or catheter into the vascular s
- Z-0777-2017 · Sep 28, 2016 · TerminatedANGIODYNAMICS Percutaneous Introducer Set under the following labels: 1) 5 F Stiffened Micro-Introducer Kit, 45CM NT/T NON-ECHO B PG, Item Number H787065970685, Catalog No.REF 0659
- Z-0775-2017 · Sep 28, 2016 · TerminatedANGIODYNAMICS Percutaneous Introducer Set under the following labels: 1) 5 F Stiffened Micro-Introducer Kit, 45CM NT/T ECHO B PG, Item Number H787065970515, Catalog No.REF 06597051
- Z-0772-2017 · Sep 28, 2016 · TerminatedANGIODYNAMICS Percutaneous Introducer Set under the following labels: 1) 5 F Stiffened Micro-Introducer Kit, 45CM SS/SS ECHO B PG, Item Number H787065970395, Catalog No.REF 0659703
- Z-1405-2013 · Apr 30, 2013 · TerminatedAngioDynamics Micro Introducer Kits under the following labeling: 1) AngioDynamics 4F Standard Micro-Introducer Kit, Sterile, Catalog Number: 06597005, 2) AngioDynamics 5F Standar
Other Trocar devices
- BRK™ G407208St. Jude Medical Cardiovascular Division · Trocar3 deaths75 reports
- BRK™ 407200St. Jude Medical Cardiovascular Division · Trocar4 deaths71 reports
- NA EP003994SMedtronic, Inc. · Trocar60 reports
- Cook G02364Cook Incorporated · Trocar1 recall5 reports
- HeartSpan® 00884450489345Merit Medical Systems, Inc. · Trocar1 recall1 death2 reports
- Cook G01377Cook Incorporated · Trocar1 recall2 reports
- HeartSpan® 00884450489369Merit Medical Systems, Inc. · Trocar1 recall0 reports
- Endrys G19261Cook Incorporated · Trocar1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page