Skip to content

Soft-Vu 10734301Recalls & reported problems

Angiodynamics, Inc. · Soft-Vu Angiographic Catheter, Kumpe, Non-Braided · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI H787107343010), received between Apr 3, 2023 and Apr 3, 2023 — classified as 1 malfunction.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

1 FDA recall record name this device's identifier, 1 still open.

22 other recalls cover devices cleared under the same FDA submission (K112452, K161596, K173762); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Recalls naming this device

  1. Mar 3, 2026Open, Classified

    The affected devices contain a manufacturing defect which may prevent the appropriately sized guidewire from passing through the inner diameter (ID) of the catheter hub.

    Z-1871-2026 · Root cause: Nonconforming Material/Component · FDA record ↗

Most reported problems

Device problems

  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 3, 2023Detachment of Device or Device Component; Detachment of Device or Device Component

    DURING THE PROCEDURE, AFTER THE SOFT-VU KMP 4F X 40CM 035 NB 0SH CATHETER HAD BEEN THREADED OVER THE WIRE, THE TIP OF THE CATHETER DETACHED INSIDE THE PATIENT. THIS WAS NOTICED WHEN WITHDRAWING THE WIRE FROM THE PATIENT. THE TIP REMAINED ON THE WIRE, AND WAS NOT RETAINED INSIDE THE PATIENT. THE CATHETER WAS REPLACED WITH A NEW OF THE SAME DEVICE AND THE PROCEDURE WAS COMPLETED. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS, HARM, OR REQUIRE MEDICAL INTERVENTION BECAUSE OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16662508 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K112452, K161596, K173762. One submission often covers many models, so check the product and lot details in each record.

  • Z-1868-2026 · Mar 3, 2026 · Open, Classified
    Soft-Vu Angiographic Catheter, Headhunter (1), Non-Braided; SOFT-VU HH1 4F X 100CM 038 NB 0SH; Catalog No.: 10714040; Product/UPN No.: H787107140405 (Box), H787107140400 (Pouch); B
  • Z-1859-2026 · Mar 3, 2026 · Open, Classified
    Soft-Vu Angiographic Catheter, Pigtail, Non-Braided; SOFT-VU PT 4F X 65CM 038 NB 10SH; Catalog No.: 10714002; Product/UPN No.: H787107140025 (Box), H787107140020 (Pouch); Box Quant
  • Z-1864-2026 · Mar 3, 2026 · Open, Classified
    Soft-Vu Angiographic Catheter, Cobra (1), Non-Braided; SOFT-VU CB1 4F X 100CM 038 NB 0SH; Catalog No.: 10714017; Product/UPN No.: H787107140175 (Box), H787107140170 (Pouch); Box Qu
  • Z-1873-2026 · Mar 3, 2026 · Open, Classified
    AccuVu Angiographic Catheter, Straight Flush, Non-Braided, With 10 R/O Markers; ACCU-VU STR FL 4F X 100CM 035 NB 6SH 10 R/O 6@1CM 4@5CM; Catalog No.: 13709604; Product/UPN No.: H78
  • Z-1866-2026 · Mar 3, 2026 · Open, Classified
    Soft-Vu Angiographic Catheter, Kumpe, Non-Braided; SOFT-VU KMP 4F X 100CM 038 NB 0SH; Catalog No.: 10714021; Product/UPN No.: H787107140215 (Box), H787107140210 (Pouch); Box Quanti
  • Z-1865-2026 · Mar 3, 2026 · Open, Classified
    Soft-Vu Angiographic Catheter, Cobra (2), Non-Braided; SOFT-VU CB2 4F X 100CM 038 NB 0SH; Catalog No.: 10714018; Product/UPN No.: H787107140185 (Box), H787107140180 (Pouch); Box Qu
  • Z-1857-2026 · Mar 3, 2026 · Open, Classified
    Soft -Vu Angiographic Catheter, Kumpe, Non-Braided; SOFT-VU KMP 4F X 40CM 038 NB 0SH; Catalog No.: 10710013; Product/UPN No.: H787107100135 (Box), H787107100130 (Pouch); Box Quanti
  • Z-1870-2026 · Mar 3, 2026 · Open, Classified
    Soft-Vu Angiographic Catheter, Omni Flush, Non-Braided; SOFT-VU OF 4F X 65CM 035 NB 6SH; Catalog No.: 10732301; Product/UPN No.: H787107323015 (Box), H787107323010 (Pouch); Box Qua
  • Z-1863-2026 · Mar 3, 2026 · Open, Classified
    Soft-Vu Angiographic Catheter, Berenstein, Non-Braided; SOFT-VU BER 4F X 100CM 038 NB 0SH; Catalog No.: 10714016; Product/UPN No.: H787107140165 (Box), H787107140160 (Pouch); Box Q
  • Z-1860-2026 · Mar 3, 2026 · Open, Classified
    Soft-Vu Angiographic Catheter, Berenstein, Non-Braided; SOFT-VU BER 4F X 65CM 038 NB 0SH; Catalog No.: 10714009; Product/UPN No.: H787107140095 (Box), H787107140090 (Pouch); Box Qu
  • Z-1861-2026 · Mar 3, 2026 · Open, Classified
    Soft-Vu Angiographic Catheter, Cobra (2), Non-Braided; SOFT-VU CB2 4F X 65CM 038 NB 0SH; Catalog No.: 10714011; Product/UPN No.: H787107140115 (Box), H787107140110 (Pouch); Box Qua
  • Z-1874-2026 · Mar 3, 2026 · Open, Classified
    AccuVu Angiographic Catheter, Omni Flush, Non-Braided, With 21 R/O Markers; ACCU-VU OF 4F X 70CM 035 NB 6SH 21 R/O 1CM; Catalog No.: 13709704; Product/UPN No.: H787137097045 (Box),
  • Z-1869-2026 · Mar 3, 2026 · Open, Classified
    Soft-Vu Angiographic Catheter, Straight Art, Non-Braided; SOFT-VU STR ART 4F X 65CM 035 NB 0SH; Catalog No.: 10731401; Product/UPN No.: H787107314015 (Box), H787107314010 (Pouch);
  • Z-1856-2026 · Mar 3, 2026 · Open, Classified
    Soft-Vu Angiographic Catheter, Pigtail, Non-Braided; SOFT-VU PT 4F X 100CM 035 NB 10SH; Catalog No.: 10707103; Product/UPN No.: H787107071035 (Box), H787107071030 (Pouch); Box Quan
  • Z-1858-2026 · Mar 3, 2026 · Open, Classified
    Soft-Vu Angiographic Catheter, Omni Flush, Non-Braided; SOFT-VU OF 4F X 65CM 038 NB 6SH; Catalog No.: 10714001; Product/UPN No.: H787107140015 (Box), H787107140010 (Pouch); Box Qua

…and 7 more.

Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices

All CATHETER, INTRAVASCULAR, DIAGNOSTIC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page