AccuVu 13709703Recalls & reported problems
Angiodynamics, Inc. · AccuVu Angiographic Catheter, Omni Flush, Non-Braided, With 21 R/O Markers · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI H787137097030), received between Apr 28, 2016 and Mar 21, 2022 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Crack” (1 report).
21 other recalls cover devices cleared under the same FDA submission (K112452, K161596); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Crack1
- Detachment of Device or Device Component1
- Fracture1
- Insufficient Device Problem Information1
Patient problems
- Device Embedded In Tissue or Plaque2
- No Known Impact Or Consequence To Patient2
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 21, 2022Detachment of Device or Device Component; Detachment of Device or Device Component
AN END USER REPORTED AN ISSUE WITH AN AV 5 100 035 OF 1CM 21RO SIZING CATHETER. DURING THE PROCEDURE, THE MARKER FELL OFF, INTO THE PATIENT. THE DECISION WAS MADE TO LEAVE THE MARKER IN SITU. THE PROCEDURE WAS COMPLETED WITH THIS CATHETER. THE TREATING PHYSICIAN ATTRIBUTED THE MARKERBAND DETACHMENT POSSIBLY TO ITS USE WITH THE COOK RUPS SET. DURING THE PROCEDURE, THE CATHETER IS WITHDRAWN THROUGH THE METAL CANNULA SEVERAL TIMES DURING A CASE AND IN DOING SO, IT MAY HAVE CAUSED THE MARKERBAND TO DETACH. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR HARM AS A RESULT OF THIS INCIDENT.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 13840066 ↗InjurySep 20, 2017Crack; Crack
AS REPORTED TO ANGIODYNAMICS ON SEPTEMBER 01, 2017: DURING A TIPSS PROCEDURE, THE CATHETER TIP BROKE OFF INSIDE OF THE PATIENT. THE FRAGMENT WAS SUCCESSFULLY REMOVED VIA A SNARE. THE DEVICE WAS REMOVED AND SET ASIDE. THE PROCEDURE WAS COMPLETED WITH A NEW OF THE SAME DEVICE. THE PATIENT DID NOT SUFFER ANY HARM OR INJURY DUE TO THIS EVENT. THE REPORTED DEFECTIVE DISPOSABLE DEVICE HAS BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION.
Patient problems: Device Embedded In Tissue or Plaque
MAUDE report 6878973 ↗MalfunctionSep 20, 2017Fracture; Fracture
AS REPORTED TO ANGIODYNAMICS ON (B)(6) 2017 VIA A (B)(4); THE OMNI FLUSH CATHETER WAS PLACED OVER A GLIDE WIRE. THE GLIDE WIRE WAS REMOVED FOR INJECTION OF CONTRAST VIA POWER INJECTOR FOR VISUALIZATION. THE INJECTION WAS NOT COMPLETE AFTER THREE ATTEMPTS. THE OMNI FLUSH CATHETER FOUND WITH A TEAR/ LARGE HOLE IN CATHETER, THUS MAKING IT UNABLE TO DELIVER THE CONTRACT TO THE END OF THE CATHETER. THE REPORTED DEFECTIVE DISPOSABLE DEVICE HAS BEEN RETURNED TO THE MANUFACTURER FOR EVALUATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6878968 ↗MalfunctionApr 28, 2016Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information; Adverse Event Without Identified Device or Use Problem; Insufficient Device Problem Information
AS REPORTED APRIL 05, 2016, A (B)(6), FEMALE PATIENT PRESENTED FOR AN ANGIOGRAPHIC PROCEDURE FOR A FILTER REMOVAL. DURING THE CASE, UPON THREADING THE CATHETER OVER THE WIRE, THE PHYSICIAN NOTED THE CATHETER TO BE SPLIT. THE CATHETER WAS REMOVED FROM THE WIRE WITH NO REPORTED COMPLICATIONS. THE DEVICE WAS SET ASIDE AND A NEW OF THE SAME WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WAS NO HARM OR INJURY TO THE PATIENT DUE TO THIS EVENT. IT WAS REPORTED THE DEFECTIVE DISPOSABLE DEVICE IS AVAILABLE FOR RETURN TO THE MANUFACTURER FOR EVALUATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 5615760 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K112452, K161596. One submission often covers many models, so check the product and lot details in each record.
- Z-1866-2026 · Mar 3, 2026 · Open, ClassifiedSoft-Vu Angiographic Catheter, Kumpe, Non-Braided; SOFT-VU KMP 4F X 100CM 038 NB 0SH; Catalog No.: 10714021; Product/UPN No.: H787107140215 (Box), H787107140210 (Pouch); Box Quanti
- Z-1871-2026 · Mar 3, 2026 · Open, ClassifiedSoft-Vu Angiographic Catheter, Kumpe, Non-Braided; SOFT-VU KMP 4F X 40CM 035 NB 0SH; Catalog No.: 10734301; Product/UPN No.: H787107343015 (Box), H787107343010 (Pouch); Box Quantit
- Z-1864-2026 · Mar 3, 2026 · Open, ClassifiedSoft-Vu Angiographic Catheter, Cobra (1), Non-Braided; SOFT-VU CB1 4F X 100CM 038 NB 0SH; Catalog No.: 10714017; Product/UPN No.: H787107140175 (Box), H787107140170 (Pouch); Box Qu
- Z-1860-2026 · Mar 3, 2026 · Open, ClassifiedSoft-Vu Angiographic Catheter, Berenstein, Non-Braided; SOFT-VU BER 4F X 65CM 038 NB 0SH; Catalog No.: 10714009; Product/UPN No.: H787107140095 (Box), H787107140090 (Pouch); Box Qu
- Z-1875-2026 · Mar 3, 2026 · Open, ClassifiedAccuVu Angiographic Catheter, Pigtail, Non-Braided, With 21 R/O Markers; ACCU-VU PT 4F X 70CM 035 NB 10SH 21 R/O 1CM; Catalog No.: 13709804; Product/UPN No.: H787137098045 (Box), H
- Z-1862-2026 · Mar 3, 2026 · Open, ClassifiedSoft-Vu Angiographic Catheter, Kumpe, Non-Braided; SOFT-VU KMP 4F X 65CM 038 NB 0SH; Catalog No.: 10714014; Product/UPN No.: H787107140145 (Box), H787107140140 (Pouch); Box Quantit
- Z-1863-2026 · Mar 3, 2026 · Open, ClassifiedSoft-Vu Angiographic Catheter, Berenstein, Non-Braided; SOFT-VU BER 4F X 100CM 038 NB 0SH; Catalog No.: 10714016; Product/UPN No.: H787107140165 (Box), H787107140160 (Pouch); Box Q
- Z-1859-2026 · Mar 3, 2026 · Open, ClassifiedSoft-Vu Angiographic Catheter, Pigtail, Non-Braided; SOFT-VU PT 4F X 65CM 038 NB 10SH; Catalog No.: 10714002; Product/UPN No.: H787107140025 (Box), H787107140020 (Pouch); Box Quant
- Z-1865-2026 · Mar 3, 2026 · Open, ClassifiedSoft-Vu Angiographic Catheter, Cobra (2), Non-Braided; SOFT-VU CB2 4F X 100CM 038 NB 0SH; Catalog No.: 10714018; Product/UPN No.: H787107140185 (Box), H787107140180 (Pouch); Box Qu
- Z-1873-2026 · Mar 3, 2026 · Open, ClassifiedAccuVu Angiographic Catheter, Straight Flush, Non-Braided, With 10 R/O Markers; ACCU-VU STR FL 4F X 100CM 035 NB 6SH 10 R/O 6@1CM 4@5CM; Catalog No.: 13709604; Product/UPN No.: H78
- Z-1872-2026 · Mar 3, 2026 · Open, ClassifiedAccuVu Angiographic Catheter, Omni Flush, Non-Braided, With 2 R/O Markers; ACCU-VU OF 4F X 70CM 035 NB 6SH 2 R/O 2CM; Catalog No.: 13709005; Product/UPN No.: H787137090055 (Box), H
- Z-1856-2026 · Mar 3, 2026 · Open, ClassifiedSoft-Vu Angiographic Catheter, Pigtail, Non-Braided; SOFT-VU PT 4F X 100CM 035 NB 10SH; Catalog No.: 10707103; Product/UPN No.: H787107071035 (Box), H787107071030 (Pouch); Box Quan
- Z-1870-2026 · Mar 3, 2026 · Open, ClassifiedSoft-Vu Angiographic Catheter, Omni Flush, Non-Braided; SOFT-VU OF 4F X 65CM 035 NB 6SH; Catalog No.: 10732301; Product/UPN No.: H787107323015 (Box), H787107323010 (Pouch); Box Qua
- Z-1861-2026 · Mar 3, 2026 · Open, ClassifiedSoft-Vu Angiographic Catheter, Cobra (2), Non-Braided; SOFT-VU CB2 4F X 65CM 038 NB 0SH; Catalog No.: 10714011; Product/UPN No.: H787107140115 (Box), H787107140110 (Pouch); Box Qua
- Z-1874-2026 · Mar 3, 2026 · Open, ClassifiedAccuVu Angiographic Catheter, Omni Flush, Non-Braided, With 21 R/O Markers; ACCU-VU OF 4F X 70CM 035 NB 6SH 21 R/O 1CM; Catalog No.: 13709704; Product/UPN No.: H787137097045 (Box),
…and 6 more.
Other CATHETER, INTRAVASCULAR, DIAGNOSTIC devices
- TrapEase Permanent Vena Cava Filter with the VisEase Angiographic Vessel Dilator 466P306AUCordis Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC6 deaths186 reports
- Dragonfly™ OPTIS™ C408646Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC4 deaths150 reports
- Dragonfly OpStar™ 1014651Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall87 reports
- LANGSTON IPN913559Teleflex Incorporated · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls14 reports
- Trevo 90238Stryker Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls5 reports
- Expo™ H749085263011Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls3 reports
- Expo™ H74908526501Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls1 report
- Expo™ H74908526031Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page