AngioDynamics 18100201Recalls & reported problems
Angiodynamics, Inc. · Percutaneous Introducer Set · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI H787181002010), received between Oct 31, 2022 and Oct 31, 2022 — classified as 1 injury.
The most frequently reported specific device problem is “Material Split, Cut or Torn” (1 report).
5 other recalls cover devices cleared under the same FDA submission (K780126); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Material Split, Cut or Torn1
Patient problems
- Cardiac Arrest1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryOct 31, 2022Material Split, Cut or Torn; Material Split, Cut or Torn
PER THE CATH LAB TEAM, THE STEMI AND TEMP PACER INSERTION PROCEDURE WAS COMPLETED ANDAN ARNOLD SHEATH FROM AN ARNOLD INTRO SET 6F WAS SUTURED IN PLACE BY A SCRUB TECHNICIAN. THE PATIENT BEGAN COUGHING AND THE SHEATH'S SUTURE EYELET SLIT, CAUSING THE SHEATH AND PACING WIRE TO BE PULLED BACK. ONLY ONE SUTURE HAD BEEN USED IN THE EYELET. THE PATIENT"S COUGHING AND A LARGE HERNIA EXERTING PRESSURE ON THE EYELET WERE BELIEVED TO CONTRIBUTE TO THE EVENT. THE PATIENT REQUIRED BRIEF 20 SECONDS OF CPR AND NEW SHEATH AND PACING WIRE INSERTED." IT WAS REPORTED THE PATIENT HAS SINCE BEEN DISCHARGED.
Patient problems: Cardiac Arrest
MAUDE report 15700266 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K780126. One submission often covers many models, so check the product and lot details in each record.
- Z-2705-2026 · Jun 1, 2026 · Open, ClassifiedOne-Piece Introducer Tray, REF: 497455, 497434 cardiovascular
- Z-2806-2020 · Jun 19, 2020 · TerminatedClosureFast Micro Introducer Sheath Set, REF MIS-7F11; 7Fr/Ch (2.3mm)x11 cm. STERILE|EO; Rx ONLY - Product Usage: Introducer Products are stand-alone components, standard kits, or
- Z-1938-2020 · Apr 8, 2020 · TerminatedPAC Tray/5 Catheter Introducer Tray 8F, Model 497424, Rx Only, STERILE EO, UDI: 00886333214440 - Product Usage: The introducers are indicated for use in arterial and venous procedu
- Z-1029-2016 · Feb 17, 2016 · TerminatedClosureFast Micro Introducer Sheath Set 7Fr/Ch(2.3mm) x 7 cm REF MIS-F707 and ClosureFast Introducer Sheath Set 7Fr/Ch(2.3mm) x 11 cm REF IS-7F11. Indicated for use in arterial and
- Z-0764-2010 · Jan 6, 2009 · TerminatedPercutaneous Catheter Introducer Set, VNUS Medical Technologies Cath Shth Intro 7F x 7cm, catalog number IS-7F07, Disposable, Single Use, Do Not Re-sterilize, Manufactured in USA f
Other DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION devices
- VersaCross Connect LAAC Access Solution VXAK0007Baylis Médicale Cie Inc · Dilator, Vessel, For Percutaneous Catheterization3 deaths108 reports
- TandemLife 5740-2417Cardiacassist, Inc. · Dilator, Vessel, For Percutaneous Catheterization2 recalls0 reports
- TightRail Rotating Dilator Sheath 545-513Spectranetics Corporation · Dilator, Vessel, For Percutaneous Catheterization20 deaths74 reports
- VersaCross Connect LAAC Access Solution VXAK0003Baylis Médicale Cie Inc · Dilator, Vessel, For Percutaneous Catheterization2 deaths70 reports
- VersaCross Connect LAAC Access Solution VXAK0007Baylis Médicale Cie Inc · Dilator, Vessel, For Percutaneous Catheterization2 deaths54 reports
- AVANTI 504656XCordis Corporation · DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION1 recall2 reports
- Abiomed® 00884450732311Merit Medical Systems, Inc. · Dilator, Vessel, For Percutaneous Catheterization1 recall1 report
- Mini Stick MAX H965457510Angiodynamics, Inc. · DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION1 recall1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page