Skip to content

Smart Port CT CT96STSARecalls & reported problems

Angiodynamics, Inc. · Smart Port CT Single Titanium Port System with Attached 9.6F x 66cm Silicone Catheter and 10F Introducer Kit · Class 2 · Implant

Adverse event reports
5
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI H787CT96STSA0), received between May 1, 2019 and Oct 28, 2024 — classified as 3 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Fracture” (3 reports).

2 other recalls cover devices cleared under the same FDA submission (K062414); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 2 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture3
  • Adverse Event Without Identified Device or Use Problem1
  • Difficult to Flush1
  • Material Fragmentation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Device Embedded In Tissue or Plaque1
  • No Known Impact Or Consequence To Patient1
  • Thrombosis/Thrombus1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • InjuryOct 28, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    ON OR ABOUT (B)(6) 2014, PLAINTIFF UNDERWENT PLACEMENT OF AN ANGIODYNAMICS SMARTPORT PRODUCT, REFERENCE NUMBER (B)(4), LOT NUMBER 4728535. THE DEVICE WAS IMPLANTED BY DR. (B))6), MD, AT (B)(6) HOSPITAL, FOR THE PURPOSE OF PROVIDING VASCULAR ACCESS. ON OR ABOUT (B)(6) 2015, PLAINTIFF WAS DIAGNOSED WITH THROMBUS APPEARING ROUND STRUCTURES AT THE BASE OF THE RIGHT ATRIUM AND SUPERIOR VENA CAVA AT (B)(6). PLAINTIFF'S MEDICAL TEAM DETERMINED THAT THE SMARTPORT WAS THE SOURCE OF THE THROMBOSIS AND THAT THE DEFECTIVE PORT HAD TO BE REMOVED. ON OR ABOUT (B)(6) 2015 PLAINTIFF'S DEFECTIVE PORT WAS REMOVED BY DR. (B)(6), MD AT (B)(6).

    Patient problems: Thrombosis/Thrombus

    MAUDE report 20547795 ↗
  • MalfunctionOct 28, 2024Fracture; Fracture

    ON OR ABOUT (B)(6) 2013 PLAINTIFF UNDERWENT PLACEMENT OF AN ANGIODYNAMICS SMARTPORT PRODUCT, REFERENCE NUMBER (B)(4), LOT NUMBER 582307. THE DEVICE WAS IMPLANTED BY DR. (B)(6), MD AT (B)(6), FOR THE PURPOSE OF PROVIDING VASCULAR ACCESS. ON OR ABOUT (B)(6) 2013, PLAINTIFFS PORT WAS FOUND TO HAVE A FRACTURED CATHETER AT THE CLAVICULAR THAT REQUIRED PORT REMOVAL AND REPLACEMENT. ON OR ABOUT (B)(6) 2013 PLAINTIFF'S DEFECTIVE PORT WAS REMOVED BY DR. (B)(6), MD AT (B)(6) HOSPITAL.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20547654 ↗
  • MalfunctionOct 24, 2020Fracture; Fracture

    A DISTRIBUTOR REPORTED RECEIVING INFORMATION, FROM A HEALTH PROFESSIONAL, REGARDING AN ISSUE WITH A SMARTPORT CT. DURING A PROCEDURE, THE VEIN DILATOR'S OUTER SHEATH SPLIT/FRACTURED AND CRUMPLED, AT THE DISTAL END TOWARDS PROXIMAL END, WHILE ATTEMPTING TO DILATE THE PATIENT'S VEIN. IT WAS INDICATED THAT THE REPORTED DEVICE IS NOT AVAILABLE TO BE RETURNED TO THE MANUFACTURER FOR EVALUATION. THERE WAS NO REPORT OF PERMANENT PATIENT HARM OR INJURY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10731851 ↗
  • InjuryJul 25, 2019Difficult to Flush; Difficult to Flush

    AS REPORTED BY ANGIODYNAMICS BY (B)(4) DISTRIBUTOR: ON (B)(6) 2019, CLINICAL TEAM LEADER OF HOSPITAL ADVISED: "YESTERDAY FOR THE INSERTION OF PORTACATH PROCEDURE USING THE SMART PORT CT, WE HAD A FAULTY PORT. PRIOR TO INSERTION TO THE PATIENT, THE PORT WAS PREPARED AND FLUSHED CORRECTLY WITH HEPARINISED SALINE. THE PORT WAS THEN INSERTED INTO THE PATIENT. SURGEON THEN WENT TO FLUSH THE PORT WITH HEPARINISED SALINE. HE ATTEMPTED TO ASPIRATE BLOOD PRIOR TO FLUSHING, HOWEVER COULD ONLY ASPIRATE AIR WITH NO BLOOD. THE CONNECTION TO THE NEEDLE AND PORT WAS DOUBLE CHECKED HOWEVER THE SAME THING WAS HAPPENING. SURGEON WAS CONCERNED ABOUT A POTENTIAL FAULTY CONNECTION OR HOLE IN THE TUBING CAUSING AIR. THE PORT HAD TO BE IMMEDIATELY REMOVED AT THE RISK OF AIR BEING INTRODUCED INTO THE PATIENT AND A NEW PORT OPENED AND INSERTED." ON 1 JULY 2019, CLINICAL TEAM LEADER PROVIDED FURTHER INFORMATION ON THE INCIDENT: IMPLANTATION: OPEN METHOD; RIGHT IJV PUNCTURED AND PORT PLACED ON RIGHT CHEST WALL; FORCEPS USED DURING IMPLANTATION: DEBAKEY FORCEPS; IMAGING DURING PROCEDURE: THE PROCEDURE WAS PERFORMED UNDER IMAGING; FREQUENT SCREENINGS WERE DONE AT THE STAGE OF NEEDLE PUNCTURING, WIRE INSERTING,

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8826486 ↗
  • InjuryMay 1, 2019Fracture; Material Fragmentation; Fracture; Material Fragmentation

    COMPLAINT CALLED IN BY PATIENT: " "HAD A PORT PUT IN AND THEN HAD IT TAKEN OUT, OR TRIED, WHEN THEY WERE TAKING IT OUT IT BROKE. THEY COULDN'T GET IT OUT AND HAD TO GO TO ANOTHER SURGEON, THEN THRU (THE) GROIN TO TRY AND GET IT OUT AND IT'S STILL THERE."

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 8568359 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K062414. One submission often covers many models, so check the product and lot details in each record.

  • Z-1466-2021 · Feb 22, 2021 · Terminated
    Smart Port CT-Implantable Port Catalogue #: CT96STSD-VI, CT96STSA-VI - Product Usage: indicated for any patient requiring repeated access of the vascular system, for delivery of me
  • Z-0056-2012 · Aug 1, 2011 · Terminated
    8F Smart Port CT, Catalog number: CT80STPD and CT80STPD-NF; AngioDynamics, INC. The Smart Port CT Implantable Port is a Single Titanium Port System with Attachable 8.0 Fr. Polyur

Other Set, administration, intravascular devices

All Set, administration, intravascular devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.