Skip to content

AngioDynamics 45748Recalls & reported problems

Navilyst Medical, Inc. · Mini Stick MAX Coaxial Microintroducer Kit · Class 2

All models of AngioDynamics (16)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI H965457480), received between Oct 23, 2023 and May 7, 2024 — classified as 2 injuries, 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (4 reports).

15 other recalls cover devices cleared under the same FDA submission (K132713); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 1 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Detachment of Device or Device Component4
  • Difficult to Remove2
  • Fracture1
  • Material Twisted/Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • Foreign Body In Patient1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 7, 2024Detachment of Device or Device Component; Detachment of Device or Device Component

    AN END USER REPORTED AND ISSUE WITH A MINI STICK MAX 4F X 10 CM STD .018 SS/SS ECHO 2.75" PG. DURING A PROCEDURE, THE GUIDEWIRE SHEERED WHEN REMOVING AND PART OF THE GUIDEWIRE REMAINED IN THE ACCESS VEIN OF THE PATIENT. THE PATIENT REQUIRED SURGERY TO REMOVE THE FRAGMENT. DESPITE MULTIPLE REQUESTS, NO FURTHER DETAILS HAVE BEEN OBTAINED.

    Patient problems: Foreign Body In Patient

    MAUDE report 19258760 ↗
  • InjuryMar 4, 2024Fracture; Detachment of Device or Device Component; Material Twisted/Bent; Fracture; Detachment of Device or Device Component; Material Twisted/Bent

    A RISK MANAGER REPORTED AN ISSUE WITH A MINI STICK MAX 4F X 10 CM STD .018 SS/SS ECHO 2.75" PG. A PATIENT WAS SCHEDULED FOR A MEDIPORT PLACEMENT. INTERVENTIONAL RADIOLOGY (IR) PHYSICIAN NOTICED A KINK IN THE SHEATH OF THE ANGIODYNAMICS MICROPUNCTURE SET, EXCHANGED IT TO A NEW SET AND COMPLETED THE PROCEDURE. FLUOROSCOPY OF THE PATIENT'S CHEST WAS PERFORMED AND POST PROCEDURE IMAGES SHOWED A FOREIGN OBJECT ON PATIENT'S RIGHT VENTRICLE. A PIECE OF MICROPUNCTURE CATHETER HAD FRAGMENTED AND EMBOLIZED TO THE RIGHT ATRIUM (RA) AND RIGHT VENTRICLE (RV). ENDOVASCULAR RETRIEVAL OF THE CATHETER WAS PERFORMED AND THE OUTER PORTION OF THE MICROPUNCTURE CATHETER WAS SUCCESSFULLY RETRIEVED; HOWEVER, THE INNER PORTION OF THE CATHETER IS STILL RETAINED BETWEEN THE RA AND RV. THE PATIENT WILL BE MONITORED, AS OUTPATIENT, FOR ANY SYMPTOMATIC ARRHYTHMIAS AND FOR ANY WORSENING MITRAL REGURGITATION (MR) WITH ECHOCARDIOGRAM IN 1 MONTH, 3 MONTHS AND 6 MONTHS, PER CARDIOLOGY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18827173 ↗
  • MalfunctionOct 23, 2023Difficult to Remove; Detachment of Device or Device Component; Difficult to Remove; Detachment of Device or Device Component

    AN END USER REPORTED AN ISSUE WITH A MINI STICK MAX 4F X 10CM STD .018 SS/SS ECHO 2.75" PG. DURING A PROCEDURE, THE MICROPUNCTURE WIRE BECAME STUCK AND WAS REMOVED FROM PATIENT. TOWARD THE END OF CASE, IT WAS DISCOVERED THAT PART OF THE MICROPUNCTURE WIRE WAS STILL INSIDE PATIENT. VASCULAR WAS CONSULTED AND THE PATIENT STAYED OVERNIGHT IN THE HOSPITAL FOR OBSERVATION AND WAS THEN DISCHARGED. THE PHYSICIAN IS AN EXPERIENCED USER OF THIS DEVICE AND THE PROCEDURE WAS PERFORMED PER THE IFU; HOWEVER, THE NURSE MANAGER REPORTED BELIEVING THAT THIS WAS NOT A PRODUCT ISSUE BUT RATHER POSSIBLY RELATED TO EITHER CALCIFIED AREAS OF ACCESS OR TECHNIQUE USED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17987340 ↗
  • MalfunctionOct 23, 2023Difficult to Remove; Detachment of Device or Device Component; Difficult to Remove; Detachment of Device or Device Component

    AN END USER REPORTED AN ISSUE WITH A MINI STICK MAX 4F X 10CM STD .018 SS/SS ECHO 2.75" PG. DURING A PROCEDURE, WHEN THE GUIDEWIRE WAS BEING REMOVED, THE WIRE BECAME STUCK. THE GUIDEWIRE WAS PULLED OUT AND A PIECE OF THE GUIDEWIRE WAS FOUND TO HAVE BROKEN OFF, INSIDE THE PATIENT. VASCULAR AND INTERVENTIONAL CARDIOLOGY WAS CONSULTED TO ASSESS SITE AND DETERMINED THAT NO INTERVENTION WAS NEEDED. THE RIGHT FEMORAL ARTERIAL SHEATH WAS REMOVED AND CLOSED WITH ANGIO-SEAL. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR HARM AS A RESULT OF THIS EVENT. THE PHYSICIAN IS AN EXPERIENCED USER OF THIS DEVICE AND THE PROCEDURE WAS PERFORMED PER THE IFU; HOWEVER, THE NURSE MANAGER REPORTED BELIEVING THAT THIS WAS NOT A PRODUCT ISSUE BUT RATHER POSSIBLY RELATED TO EITHER CALCIFIED AREAS OF ACCESS OR TECHNIQUE USED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17987294 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K132713. One submission often covers many models, so check the product and lot details in each record.

  • Z-1338-2024 · Jan 30, 2024 · Open, Classified
    MINI STICK MAX 5F X 10 CM STD .018 SS/SS ECHO 2.75" PG -Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-754
  • Z-1348-2024 · Jan 30, 2024 · Open, Classified
    MINI STICK MAX 4F X 10CM STIFF .018 NI/TU ECHO 1.6" PG- Used for the percutaneous introduction of a guidewire into the vascular system. - Catalog Number: 45-781
  • Z-1340-2024 · Jan 30, 2024 · Open, Classified
    MINI STICK MAX 5F X 10 CM STIFF .018 SS/PD ECHO 2.75" PG-Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-758
  • Z-1347-2024 · Jan 30, 2024 · Open, Classified
    MINI STICK MAX 5F X 15 CM STIFF .018 NI/TU NON-ECHO 2.75" PG-Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-776
  • Z-1344-2024 · Jan 30, 2024 · Open, Classified
    MINI STICK MAX 5F X 10 CM STD .018 SS/SS NON-ECHO 2.75" PG-Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-767
  • Z-1346-2024 · Jan 30, 2024 · Open, Classified
    MINI STICK MAX 5F X 10 CM STD .018 NI/TU ECHO 1.6" PG- Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-775
  • Z-1339-2024 · Jan 30, 2024 · Open, Classified
    MINI STICK MAX 5F X 10 CM STD .018 NI/TU ECHO 2.75" PG Catalog Number: 45-756
  • Z-1342-2024 · Jan 30, 2024 · Open, Classified
    MINI STICK MAX 4F X 10 CM STIFF .018 NI/TU NON-ECHO 2.75" PG- Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-762
  • Z-1345-2024 · Jan 30, 2024 · Open, Classified
    MINI STICK MAX 5F X 10 CM STD .018 SS/SS ECHO 1.6" PG- Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-774
  • Z-1334-2024 · Jan 30, 2024 · Open, Classified
    MINI STICK MAX 4F X 10 CM STD .018 SS/PD ECHO 2.75" PG-Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-749
  • Z-1335-2024 · Jan 30, 2024 · Open, Classified
    MINI STICK MAX 4F X 10 CM STD .018 NI/TU ECHO 2.75" PG-Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-750
  • Z-1343-2024 · Jan 30, 2024 · Open, Classified
    MINI STICK MAX 5F X 10 CM STD .018 SS/SS NON-ECHO 2.75" PG- Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-765
  • Z-1336-2024 · Jan 30, 2024 · Open, Classified
    MINI STICK MAX 4F X 10 CM STIFF .018 SS/SS ECHO 2.75" PG-Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-751
  • Z-1337-2024 · Jan 30, 2024 · Open, Classified
    MINI STICK MAX 4F X 10 CM STIFF .018 SS/PD ECHO 2.75" PG-Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-752
  • Z-1341-2024 · Jan 30, 2024 · Open, Classified
    MINI STICK MAX 5F X 10 CM STIFF .018 NI/TU ECHO 2.75" PG-Used for the percutaneous introduction of a guidewire into the vascular system. Catalog Number: 45-759

Other DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION devices

All DILATOR, VESSEL, FOR PERCUTANEOUS CATHETERIZATION devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page