Skip to content

BioFlo 45802Recalls & reported problems

Navilyst Medical, Inc. · BioFlo PICC with Endexo and PASV Valve Technology - 3F x 55cm Catheter MST-70 Kit · Class 2 · Implant

All models of BioFlo (87)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI H965458020), received between Dec 27, 2024 and Jun 30, 2025 — classified as 7 malfunctions.

The most frequently reported specific device problem is “Crack” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 3 reportsJan 2025: 0 reportsFeb 2025: 1 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 2 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Crack7

Patient problems

  • No Clinical Signs, Symptoms or Conditions7

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 30, 2025Crack; Crack

    AN END USER EXPERIENCED AN ISSUE WHEN USING A BIOFLO PICC 3F SL-55CM MST-70 KIT VALVED PG. POST PLACEMENT, IT WAS NOTED THE HUB WAS CRACKED. THE PICC REQUIRED REMOVAL AND WAS REPLACED. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR HARM BECAUSE OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22366446 ↗
  • MalfunctionJun 30, 2025Crack; Crack

    AN END USER EXPERIENCED AN ISSUE WHEN USING A BIOFLO PICC 3F SL-55CM MST-70 KIT VALVED PG. POST PLACEMENT, IT WAS NOTED THE HUB WAS CRACKED. THE PICC REQUIRED REMOVAL. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR HARM BECAUSE OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22366445 ↗
  • MalfunctionMar 1, 2025Crack; Crack

    AN END USER EXPERIENCED AN ISSUE WITH A PICC FROM A BIO-STABLE 3F SL-55CM MST-70 KIT VALVED. IT WAS REPORTED POST PLACEMENT, THAT THE PICC HUB CRACKED, REQUIRING REMOVAL. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR HARM BECAUSE OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21499562 ↗
  • MalfunctionFeb 7, 2025Crack; Crack

    AN END USER REPORTED AN ISSUE WITH A PICC FROM A BIO-STABLE 3F SL-55CM MST-70 VALVED KIT. THE PATIENT HAD THE PICC PLACED ON (B)(6) 2024. IT WAS REPORTED POST PLACEMENT, THAT THE PICC HUB CRACKED, REQUIRING REMOVAL AND REPLACEMENT. THE PICC WAS REMOVED AND REPLACED ON JANUARY 4, 2025. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR HARM BECAUSE OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21338276 ↗
  • MalfunctionDec 27, 2024Crack; Crack

    A SPECTRUM SENIOR DISTRICT MANAGER REPORTED AN END USER EXPERIENCED AN ISSUE WITH A PICC FROM A BIO-STABLE 3F SL-55CM MST-70 KIT VALVED. IT WAS REPORTED POST PLACEMENT, THAT THE PICC HUB CRACKED, REQUIRING REMOVAL AND REPLACEMENT. PER STAFF, THEY HAVE NOT USED CLAMPS OR EXCESS TORQUE ON THE CONNECTION TO THE NEEDLELESS ACCESS VALVES. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR HARM BECAUSE OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21026653 ↗
  • MalfunctionDec 27, 2024Crack; Crack

    A SPECTRUM SENIOR DISTRICT MANAGER REPORTED AN END USER EXPERIENCED AN ISSUE WITH A PICC FROM A BIO-STABLE 3F SL-55CM MST-70 KIT VALVED. IT WAS REPORTED POST PLACEMENT, THAT THE PICC HUB CRACKED, REQUIRING REMOVAL AND REPLACEMENT. PER STAFF, THEY HAVE NOT USED CLAMPS OR EXCESS TORQUE ON THE CONNECTION TO THE NEEDLELESS ACCESS VALVES. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR HARM BECAUSE OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21026647 ↗
  • MalfunctionDec 27, 2024Crack; Crack

    A SPECTRUM SENIOR DISTRICT MANAGER REPORTED AN END USER EXPERIENCED AN ISSUE WITH A PICC FROM A BIO-STABLE 3F SL-55CM MST-70 KIT VALVED. IT WAS REPORTED POST PLACEMENT, THAT THE PICC HUB WAS CRACKED, REQUIRING REMOVAL AND REPLACEMENT. PER STAFF, THEY HAVE NOT USED CLAMPS OR EXCESS TORQUE ON THE CONNECTION TO THE NEEDLELESS ACCESS VALVES. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR HARM BECAUSE OF THIS INCIDENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions