BioFlo 45870Recalls & reported problems
Navilyst Medical, Inc. · BioFlo PICC with Endexo Technology - 5F x 55cm Dual Lumen Catheter Kit · Class 2 · Implant
All models of BioFlo (87)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI H965458700), received between Mar 13, 2020 and May 13, 2020 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Crack” (5 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Crack5
- Fluid/Blood Leak1
- Fracture1
Patient problems
- No Known Impact Or Consequence To Patient6
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 13, 2020Crack; Crack
IT WAS REPORTED THAT A BIOFLO NON-VALVED PICC LINE EXPERIENCED A HUB CRACK WHEN CHANGING OUT THE RYMED END CAPS. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED BUT PATIENT REQUIRED PICC REPLACEMENT. IT WAS REPORTED THE DEFECTIVE DISPOSABLE DEVICE IS NOT AVAILABLE FOR RETURN TO THE MANUFACTURER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10052316 ↗MalfunctionMay 13, 2020Crack; Crack
IT WAS REPORTED THAT A BIOFLO NON-VALVED PICC LINE EXPERIENCED A HUB CRACK WHEN CHANGING OUT THE RYMED END CAPS. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED BUT PATIENT REQUIRED PICC REPLACEMENT. IT WAS REPORTED THE DEFECTIVE DISPOSABLE DEVICE IS NOT AVAILABLE FOR RETURN TO THE MANUFACTURER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10052315 ↗MalfunctionMay 13, 2020Crack; Crack
IT WAS REPORTED THAT A BIOFLO NON-VALVED PICC LINE EXPERIENCED A HUB CRACK WHEN CHANGING OUT THE RYMED END CAPS. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED BUT PATIENT REQUIRED PICC REPLACEMENT. IT WAS REPORTED THE DEFECTIVE DISPOSABLE DEVICE IS NOT AVAILABLE FOR RETURN TO THE MANUFACTURER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10052311 ↗MalfunctionMay 13, 2020Crack; Crack
IT WAS REPORTED THAT A BIOFLO NON-VALVED PICC LINE EXPERIENCED A HUB CRACK WHEN CHANGING OUT THE RYMED END CAPS. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED BUT PATIENT REQUIRED PICC REPLACEMENT. IT WAS REPORTED THE DEFECTIVE DISPOSABLE DEVICE IS NOT AVAILABLE FOR RETURN TO THE MANUFACTURER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10052310 ↗MalfunctionMay 13, 2020Crack; Crack
IT WAS REPORTED THAT A BIOFLO NON-VALVED PICC LINE EXPERIENCED A HUB CRACK WHEN CHANGING OUT THE RYMED END CAPS. THERE WAS NO PATIENT HARM OR ADVERSE EVENT REPORTED BUT PATIENT REQUIRED PICC REPLACEMENT. IT WAS REPORTED THE DEFECTIVE DISPOSABLE DEVICE IS NOT AVAILABLE FOR RETURN TO THE MANUFACTURER.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10052309 ↗MalfunctionMar 13, 2020Fluid/Blood Leak; Fracture; Fluid/Blood Leak; Fracture
AS REPORTED: BIOSTABLE PICC LINE CRACKED. DEVICE WAS REMOVED AND REPLACED. NO PATIENT HARM OR INJURY DUE TO THIS EVENT. IT WAS REPORTED THE DEFECTIVE DISPOSABLE DEVICE IS AVAILABLE FOR RETURN TO THE MANUFACTURER FOR EVALUATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9829112 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC Catheter 3275118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days272 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page