BioFlo 45876Recalls & reported problems
Navilyst Medical, Inc. · BioFlo PICC with Endexo Technology - 4F x 55cm Catheter MST-70 Kit · Class 2 · Implant
All models of BioFlo (87)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI H965458760), received between Apr 18, 2022 and Jan 20, 2026 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Detachment of Device or Device Component1
- Fluid/Blood Leak1
- Fracture1
- Material Puncture/Hole1
Patient problems
- Device Embedded In Tissue or Plaque1
- Foreign Body In Patient1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 20, 2026Fracture; Detachment of Device or Device Component
A DISTRIBUTOR REPORTED AN END USER EXPERIENCED AN ISSUE WHEN USING A BIO-STABLE 4F SL 55CM MST-70 KIT NON-VALVED. DURING PICC PLACEMENT, THE STAINLESS STEEL GUIDEWIRE FRACTURED INTO TWO SEPARATE PIECES INSIDE THE PATIENT. THE WIRE HAD UNRAVELED PRIOR TO FRACTURING DUE TO THE EXTENDED TIME OF MANIPULATING THE WIRE AND TRYING TO FREE IT FROM THE COLLATERAL. THEY WERE UNABLE TO REMOVE THE RETAINED FRAGMENT FROM THE PATIENT'S RIGHT ARM WITHIN THE COLLATERAL VEIN CONNECTED TO THE BRACHIAL VEIN. THE COLLATERAL WAS SMALL AND TORTUOUS, AND HAD SPASMED. THEY BELIEVE THE FRAGMENT WILL NOT MIGRATE FROM THAT POSITION. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH ANOTHER OF THE SAME DEVICE. THE PATIENT WAS INFORMED OF THE EVENT AND REPORTED TO HAVE NO CONCERNS AT THAT TIME. IT HAD BEEN REPORTED THE PATIENT HAD A NICKEL ALLERGY. THE PATIENT DID NOT EXPERIENCE ANY HARM AS A RESULT OF THIS INCIDENT.
Patient problems: Device Embedded In Tissue or Plaque; Foreign Body In Patient
MAUDE report 24116888 ↗MalfunctionApr 18, 2022Fluid/Blood Leak; Material Puncture/Hole; Fluid/Blood Leak; Material Puncture/Hole
A SENIOR CLINICAL RISK ADVISOR REPORTED AN ISSUE WITH A BIO-STABLE 4F SL 55CM MST-70 KIT. SOME TIME AFTER PLACEMENT OF THE DEVICE, THE PICC LINE TUBING DEVELOPED A HOLE IN THE STERILE, DRESSED PORTION OF THE TUBING; THUS RENDERING THE LINE INEFFECTIVE. IT IS UNKNOWN HOW OR WHEN THE HOLE OCCURRED; HOWEVER, THE PICC WAS REMOVED AND A PERIPHERAL INTRAVENOUS LINE (PIV) HAD TO BE INSERTED. THIS INHIBITED THE PATIENT FROM RECEIVING TPN. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR AS A RESULT OF THIS INCIDENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14135570 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC Catheter 3275118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days272 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page