BioFlo 45894Recalls & reported problems
Navilyst Medical, Inc. · BioFlo PICC with Endexo and PASV Valve Technology - 3F x 55cm Catheter MST-70 Kit with Nitinol Wire · Class 2 · Implant
All models of BioFlo (87)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI H965458940), received between Apr 20, 2018 and Jun 19, 2025 — classified as 6 malfunctions.
The most frequently reported specific device problem is “Fracture” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture3
- Crack2
- Fluid/Blood Leak1
- Material Puncture/Hole1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- No Known Impact Or Consequence To Patient3
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
MalfunctionJun 19, 2025Fluid/Blood Leak; Material Puncture/Hole; Fluid/Blood Leak; Material Puncture/Hole
A DISTRIBUTOR REPORTED A CUSTOMER COMPLAINT REGARDING A BIO-STABLE 3F SL-55CM MST-70 KIT VALVED WITH NITINOL GUIDEWIRE PG. ONE DAY FOLLOWING PICC PLACEMENT, WHEN NURSES INJECTED WATER THROUGH THE SYRINGE CONNECTOR, WATER WAS FOUND TO BE LEAKING FROM THE CATHETER LINE. THERE WAS NO RESISTANCE WHEN PUSHING WATER OR MEDICATION THROUGH THE DEVICE. ULTIMATELY, THE PATIENT REQUIRED REMOVAL AND REPLACEMENT OF THE DEVICE 6 DAYS LATER. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR HARM BECAUSE OF THIS INCIDENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22292870 ↗MalfunctionJul 18, 2024Crack; Crack
IT WAS REPORTED THAT "A 3 FR BIOFLO PICC WAS RECENTLY INSERTED ON A 2-WEEK INFANT. THE PURPLE HUB HAS CRACKED AND THE DEVICE NEEDS TO REPLACED." THE CRACK WAS FOUND TO HAVE OCCURRED AT THE AREA OF THE RIDGES THE WHERE THE NEEDLELESS ACCESS DEVICE SCREWS ON. THIS INCIDENT WAS NOTICED ON THE (B)(6) 2024 WHEN THE "BABY WAS CRYING, WHEN CHECKED PATIENTS' BLANKETS AND CLOTHING WERE SOAKED WITH PN & SMOF" CUSTOMER HAS ALSO ADVISED THAT A NEEDLESS CONNECTOR (LUER-LOCK) WAS CONNECTED TO THE HUB AT THE TIME THE ISSUE WAS NOTICED. THE BRAND NAME IS REPORTED TO BE: NANOCLAVE. THE PATIENT HAS HAD THEIR PICC REMOVED TODAY ((B)(6) 2024) UNDER GENERAL ANAESTHESIA AND HAS HAD A REPLACEMENT DEVICE INSERTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19772640 ↗MalfunctionNov 6, 2023Crack; Crack
A DISTRIBUTOR REPORTED AN END USER EXPERIENCED AN ISSUE WITH A PICC FROM A BIO-STABLE 3F SL-55CM MST-70 KIT, VALVED WITH NITINOL GUIDEWIRE, THE CUSTOMER REPORTED THAT THE PICC WAS REMOVED DUE TO A [?]FRACTURE' (I.E. CRACK) IN THE HUB. THIS INCIDENT WAS NOTICED ON THE (B)(6) WITH A PATIENT RECEIVING HOME PN. THE PICC WAS FIRST IMPLANTED IN THE PATIENT ON THE (B)(6)2023. THE EXACT CIRCUMSTANCES OF HOW THE ISSUE WAS FIRST NOTED WERE NOT ADVISED BY THE CUSTOMER IN THEIR REPORT. THE PICC WAS THEN REMOVED, AND WAS REPLACED WITH ANOTHER PICC. THERE WAS NO REPORTED INJURY/HARM TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18073651 ↗MalfunctionSep 19, 2018Fracture; Fracture
AS REPORTED BY ANGIODYNAMICS' DISTRIBUTOR IN (B)(4), INDWELLING BIOFLO PICC WAS REMOVED DUE TO A CRACKED HUB. WHILE THE EXACT IMPLANT AND EXPLANT DATES ARE NOT KNOWN, EXPLANTATION WAS ON OR EARLIER THAN (B)(6) 2018. NO PATIENT INJURY OR COMPLICATIONS WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7888884 ↗MalfunctionSep 19, 2018Fracture; Fracture
AS REPORTED BY ANGIODYNAMICS' DISTRIBUTOR IN (B)(4), INDWELLING BIOFLO PICC WAS REMOVED DUE TO A CRACKED HUB. WHILE THE EXACT IMPLANT AND EXPLANT DATES ARE NOT KNOWN, EXPLANTATION WAS ON OR EARLIER THAN (B)(6) 2018. NO PATIENT INJURY OR COMPLICATIONS WERE REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7888883 ↗MalfunctionApr 20, 2018Fracture; Fracture
AS REPORTED BY ANGIODYNAMICS' DISTRIBUTOR IN (B)(4), AN INDWELLING PICC WAS REMOVED AND REPLACE DUE TO A HOLE LOCATED DISTAL TO THE SUTURE WING. NO PATIENT INJURY WAS REPORTED. THE USED DEVICE HAS BEEN RETURNED TO ANGIODYNAMICS FOR EVALUATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7446238 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC® Catheters 1174108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall222 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page