BioFlo 45895Recalls & reported problems

Navilyst Medical, Inc. · BioFlo PICC with Endexo Technology - 5F x 55cm Dual Lumen Catheter MST-70 Kit with Nitinol Wire · Class 2 · Implant

All models of BioFlo (87)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
6
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI H965458950), received between Mar 20, 2018 and Mar 4, 2026 — classified as 2 injuries, 4 malfunctions.

The most frequently reported specific device problem is “Fracture” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture3
  • Physical Property Issue2
  • Infusion or Flow Problem1

Patient problems

  • No Known Impact Or Consequence To Patient3
  • Fever2
  • Unspecified Infection2
  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 4, 2026Infusion or Flow Problem

    A SPECTRUM DISTRIBUTOR REPORTED AN END USER EXPERIENCED AN ISSUE WHEN USING A BIOFLO 5F DL-55CM MST-70 KIT NON-VALVED WITH NITINOL GUIDEWIRE BRAZIL. IT WAS REPORTED THAT THERE WAS DIFFICULTY EXTRAVASATING CONTRAST FROM THE CATHETER POSSIBLE DUE TO IMPROPER POSITIONING. DUE TO THIS ISSUE, THE PICC WAS REMOVED AND REPLACED. THERE IS NO REPORT OF PATIENT HARM, OR ADVERSE EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24505714 ↗
  • InjuryJun 12, 2018Physical Property Issue; Physical Property Issue

    ANGIODYNAMICS WAS NOTIFIED BY (B)(4), THE SUPPLIER OF INTRODUCER SHEATHS, THAT THEY HAVE INITIATED A RECALL (1649395-05-02-2018-001-R) ON SPECIFIC DEVICE LOTS DUE TO "UNSAFE LEVELS OF BACTERIAL ENDOTOXINS (PYROGENS) THAT WERE INTRODUCED DURING A MANUFACTURING STEP." (B)(4) HAS INFORMED ANGIODYNAMICS THAT USE OF THE INTRODUCERS MAY EXPOSE PATIENTS/END USERS TO "BACTERIAL ENDOTOXINS, ALSO CALLED PYROGENIC BACTERIA THAT CAN ACTIVATE THE INFLAMMATORY PROCESS AND PRODUCE FEVER, CHILLS, AND HYPOTENSION IN A PATIENT." AS A RESULT OF THE (B)(4) RECALL, ANGIODYNAMICS IS CONDUCTING A VOLUNTARY RECALL (REMOVAL) OF THE AFFECTED ANGIODYNAMICS PRODUCTS. THE RECALL NOTIFICATION LETTERS WERE DELIVERED TO AFFECTED CONSIGNEES ON FRIDAY MAY 11, 2018. AFTER RECEIVING NOTIFICATION, THE END USER HOSPITAL HAS REPORTED: "PATIENT EXHIBITED POTENTIAL INFECTIOUS SYMPTOMS THE DAY AFTER PROCEDURE UTILIZING A BIOFLO PICC. INCREASED TEMP AND HR OVER THE FOLLOWING 24 HOURS TO 38.5 (101.3) DESPITE ANTIBIOTICS. BLOOD CULTURES, CHEST X-RAY ID CONSULT ON (B)(6). NO TREATMENT REQUIRED. NO ADDITIONAL MEDICATIONS INITIATED - HAD BEEN ON MORPHINE Q4 IV AND OXYCODONE 4-6 HOURS FOR SEVERAL DAYS. NO CHANGE IN FREQUENCY OF T

    Patient problems: Fever; Unspecified Infection

    MAUDE report 7591595 ↗
  • InjuryJun 12, 2018Physical Property Issue; Physical Property Issue

    ANGIODYNAMICS WAS NOTIFIED BY (B)(4), THE SUPPLIER OF INTRODUCER SHEATHS, THAT THEY HAVE INITIATED A RECALL (1649395-05-02-2018-001-R) ON SPECIFIC DEVICE LOTS DUE TO "UNSAFE LEVELS OF BACTERIAL ENDOTOXINS (PYROGENS) THAT WERE INTRODUCED DURING A MANUFACTURING STEP." (B)(4) HAS INFORMED ANGIODYNAMICS THAT USE OF THE INTRODUCERS MAY EXPOSE PATIENTS/END USERS TO "BACTERIAL ENDOTOXINS, ALSO CALLED PYROGENIC BACTERIA THAT CAN ACTIVATE THE INFLAMMATORY PROCESS AND PRODUCE FEVER, CHILLS, AND HYPOTENSION IN A PATIENT." AS A RESULT OF THE (B)(4) RECALL, ANGIODYNAMICS IS CONDUCTING A VOLUNTARY RECALL (REMOVAL) OF THE AFFECTED ANGIODYNAMICS PRODUCTS. THE RECALL NOTIFICATION LETTERS WERE DELIVERED TO AFFECTED CONSIGNEES ON FRIDAY MAY 11, 2018. AFTER RECEIVING NOTIFICATION, THE END USER HOSPITAL HAS REPORTED: "PATIENT EXHIBITED POTENTIAL INFECTIOUS SYMPTOMS THE NEXT DAY. NO TREATMENT REQUIRED. (B)(6) - INCREASED TEMP ONE TIME TO 38.6 (101.5) DESPITE ANTIBIOTICS WITH C/O CHILLS, SWEATS, GENERALIZED ACHES AND HA PATIENT HAD NOW RECALLED PICC LINE DEVICE PLACED (B)(6) 2018.

    Patient problems: Fever; Unspecified Infection

    MAUDE report 7591592 ↗
  • MalfunctionMar 20, 2018Fracture; Fracture

    AS REPORTED, AN INDWELLING PICC WAS REMOVED AND REPLACED DUE TO A CRACKED HUB. NO PATIENT INJURY OR COMPLICATION WAS REPORTED. THE USED PICC WILL BE RETURNED TO ANGIODYNAMICS FOR EVALUATION.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7352057 ↗
  • MalfunctionMar 20, 2018Fracture; Fracture

    AS REPORTED, AN INDWELLING PICC WAS REMOVED AND REPLACED DUE TO A CRACKED HUB. NO PATIENT INJURY OR COMPLICATION WAS REPORTED. THE USED PICC WILL BE RETURNED TO ANGIODYNAMICS FOR EVALUATION.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7352055 ↗
  • MalfunctionMar 20, 2018Fracture; Fracture

    AS REPORTED, AN INDWELLING PICC WAS REMOVED AND REPLACED DUE TO A CRACKED HUB. NO PATIENT INJURY OR COMPLICATION WAS REPORTED. THE USED PICC WILL BE RETURNED TO ANGIODYNAMICS FOR EVALUATION.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7352054 ↗

Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices

All Catheter,intravascular,therapeutic,long-term greater than 30 days devices →