BioFlo 60M780126Recalls & reported problems
Navilyst Medical, Inc. · PICC Convenience Kit · Class 2 · Implant
All models of BioFlo (87)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI H96560M7801260), received between May 24, 2018 and May 24, 2018 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Obstruction of Flow” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Obstruction of Flow3
- Malposition of Device2
- Kinked1
Patient problems
- No Known Impact Or Consequence To Patient3
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 24, 2018Obstruction of Flow; Malposition of Device; Obstruction of Flow; Malposition of Device
AS REPORTED BY HOSPITAL IN (B)(6): PICC PLACED (B)(6) - HD ETOPOSIDE GIVE (B)(6). XRAY ORDERED BY PHYSICIAN FOR IMPROPERLY FUNCTIONING PICC (B)(6) - SEEN COILED IN SVC - COULD NOT BE FLUSHED. WAS REMOVED AND REPLACED (B)(6) - LINE NOTED TO BE DISCOLORED. HOSPITAL HAS STATED THAT THE DFU FOR ETOPOSIDE LISTS POLYURETHANE AS CONTRAINDICATED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7540994 ↗MalfunctionMay 24, 2018Obstruction of Flow; Malposition of Device; Obstruction of Flow; Malposition of Device
AS REPORTED BY HOSPITAL IN (B)(6): PICC PLACED (B)(6) 2018 - SEEN FOR OCCLUSIONS (B)(6) (TPA). INCIDENTAL XRAY (B)(6) SHOWS GOOD POSITIONING. HD ETOPOSIDE GIVEN (B)(6). INCIDENTAL XRAY DONE (B)(6) SHOWS CURL. PICC REMOVED AND REPLACED (B)(6). HOSPITAL HAS STATED THAT THE DFU FOR ETOPOSIDE LISTS POLYURETHANE AS CONTRAINDICATED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7540993 ↗MalfunctionMay 24, 2018Kinked; Obstruction of Flow; Kinked; Obstruction of Flow
AS REPORTED BY HOSPITAL IN (B)(6): PICC PLACED (B)(6) 2018. HD ETOPOSIDE GIVEN (B)(6) AND (B)(6). INCIDENTAL XRAYS DONE (B)(6) SHOW PROPER LINE PLACEMENT. XRAY ORDERED BY PHYSICIAN (B)(6) FOR "POOR FLOW". LINE NOTED TO BE "KINKED IN SVC". PICC WAS REMOVED FOR NEW INSERTION AND LINE DISCOLORATION NOTED. HOSPITAL HAS STATED THAT THE DFU FOR ETOPOSIDE LISTS POLYURETHANE AS CONTRAINDICATED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7540992 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC® Catheters 1174108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall222 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.