The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI H965750010), received between Oct 9, 2024 and Oct 9, 2024 — classified as 1 malfunction.

The most frequently reported specific device problem is “Fracture” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture1

Patient problems

  • Device Embedded In Tissue or Plaque1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 9, 2024Fracture; Fracture

    A SPECTRUM MEDICAL TERRITORY MANAGER REPORTED AN END USER EXPERIENCED AN ISSUE WITH AN XCELA 4FSL-55CM MAXIMAL BARRIER NURSING KIT . THE FOLLOWING WAS REPORTED: "SINGLE LUMEN PICC INSERTED 8/27/24 FOR VANCOMYCIN ADMINISTRATION. RIGHT UPPER ARM BASILIC VEIN USED, VESSEL DIAM 3.1MM, CVR APPROX 42%, INSERTION UNCOMPLICATED AND 1 ATTEMPT. CATHETER PERFORMED WITHOUT ISSUE FOR DURATION OF THERAPY. UPON REMOVAL OF CATHETER, RN EXPERIENCED SIGNIFICANT RESISTANCE. SURGEON WAS CONTACTED. APPROX AN HOUR WAS WAITED UNTIL NEXT REMOVAL ATTEMPT. NO WARM PACKS WERE TRIED. NO ANTI-ANXIETY MEDS WERE GIVEN. THE SURGEON TRIED AND EXPERIENCED THE SAME FINDING: SEVERE RESISTANCE. TOOK THE PATIENT TO A PROCEDURE ROOM FOR REMOVAL USING LIVE ULTRASOUND, FLUOROSCOPY, AND A 0.018 GUIDEWIRE. CATHETER WAS EVENTUALLY REMOVED (SNAPPED IN TWO DUE TO ALL THE PULLING AND RESISTANCE). ENTIRE CATHETER WAS RETRIEVED/REMOVED. CATHETER REMOVED WITH SIGNIFICANT RESISTANCE THE ENTIRE LENGTH OF THE VEIN. NO CLOT OBSERVED, CATHETER LUMEN OPEN/PATENT, NO KINKS OBSERVED ON ULTRASOUND OR FLURO. PT HAD A VENOGRAM POST PROCEDURE AND IT WAS NORMAL. IT WAS EITHER COMPLETED RIGHT AFTER THE PROCEDURE AND/OR APPROX 1 WEEK POST REMOVA

    Patient problems: Device Embedded In Tissue or Plaque

    MAUDE report 20407382 ↗