BioFlo 75014Recalls & reported problems
Navilyst Medical, Inc. · BioFlo PICC with Endexo Technology - 5F x 55cm Dual Lumen Catheter Maximal Barrier Nursing Kit · Class 2 · Implant
All models of BioFlo (87)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI H965750140), received between Jun 21, 2016 and Nov 8, 2023 — classified as 1 injury, 10 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (6 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak6
- Material Puncture/Hole3
- Material Rupture3
- Fracture2
- Crack1
- Detachment of Device or Device Component1
Patient problems
- No Known Impact Or Consequence To Patient9
- Bacterial Infection1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
MalfunctionNov 8, 2023Crack; Crack
A TERRITORY MANAGER REPORTED AN END USER EXPERIENCED AN ISSUE WITH A PICC FROM A BIOFLO PICC (NON-VALVED) 5FDL-55CM MAXIMAL BARRIER NURSING KIT. IT WAS REPORTED THE HUB WAS FOUND TO BE CRACKED AT THE CLOSURE OF THE TPN INFUSION PROCEDURE. THE DEVICE WAS REMOVED AND REPLACED. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS, HARM, OR REQUIRE MEDICAL INTERVENTION BECAUSE OF THIS INCIDENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18094824 ↗MalfunctionJul 26, 2017Detachment of Device or Device Component; Detachment of Device or Device Component
AS REPORTED, PICC PLACED IN THE ICU. AS SOON AS THE NURSE LEFT THE ICU ONE OF THE HUBS FELL OFF THE DUAL LUMEN PICC. THE PICC NURSE DID AN OVER-THE-WIRE EXCHANGE WITH A NEW PICC. SHE CHECKED THE HUBS ON THE 2ND PICC AND THEY WERE SECURE. THE PATIENT CONDITION WAS REPORTED AS "STABLE."
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6744558 ↗MalfunctionFeb 22, 2017Fluid/Blood Leak; Material Puncture/Hole; Fluid/Blood Leak; Material Puncture/Hole
AS REPORTED BY NURSE PRO PLUS, AN IV SERVICES COMPANY, PICC PLACED AT (B)(6) MEDICAL CENTER DEVELOPED A "PIN HOLE IN THE EXTENSION (TUBING) CAUSING LEAKAGE." THE LINE WAS REMOVED AND REPLACED. THE PATIENT CONDITION WAS REPORTED AS "UNAFFECTED" AND THE DEVICE HAS BEEN RETURNED TO ANGIODYNAMICS FOR EVALUATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6349707 ↗MalfunctionFeb 22, 2017Fluid/Blood Leak; Material Puncture/Hole; Fluid/Blood Leak; Material Puncture/Hole
AS REPORTED BY NURSE PRO PLUS, AN IV SERVICES COMPANY, PICC PLACED AT ROANE MEDICAL CENTER DEVELOPED A "PIN HOLE IN THE EXTENSION (TUBING) CAUSING LEAKAGE." THE LINE WAS REMOVED AND REPLACED. THE PATIENT CONDITION WAS REPORTED AS "UNAFFECTED" AND THE DEVICE HAS BEEN RETURNED TO ANGIODYNAMICS FOR EVALUATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6349680 ↗InjuryFeb 22, 2017Fluid/Blood Leak; Material Puncture/Hole; Fluid/Blood Leak; Material Puncture/Hole
AS REPORTED BY NURSE PRO PLUS, AN IV SERVICES COMPANY, PICC PLACED AT THE (B)(6) MEDICAL CENTER DEVELOPED A "PIN HOLE IN THE EXTENSION (TUBING) CAUSING LEAKAGE." THE LINE WAS REMOVED AND ANOTHER PICC WAS PLACED THE DAY AFTER. THE PATIENT HAD AN INFECTION WHICH WAS SUSPECTED TO BE THE RESULT OF THE HOLE IN THE EXTENSION TUBING. THE PATIENT HAD THE PICC LINE IN PLACE FOR MORE THAN TWO MONTHS PRIOR TO THE HOLE BEING DISCOVERED IN THE LINE. THE PATIENT WAS DISCHARGED AND RE-ADMITTED TWICE PRIOR TO THE HOLE IN THE LINE BEING NOTICED. THE PATIENT HAD ORIGINALLY BEEN ADMITTED FOR A GUNSHOT WOUND AND RECEIVED NUMEROUS MEDICATIONS THROUGH THE PICC LINE BOTH AS AN INPATIENT AND AS AN OUTPATIENT. THE USED PICC WAS DISCARDED AT THE HOSPITAL.
Patient problems: Bacterial Infection
MAUDE report 6349664 ↗MalfunctionSep 21, 2016Material Rupture; Material Rupture
AS REPORTED, DURING POWER INJECTION OF 5ML/SEC., THE PICC CATHETER TUBING BURST AT APPROXIMATELY THE 0CM MARK. THE PATIENT CONDITION WAS REPORTED AS "STABLE."
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 5966419 ↗MalfunctionSep 21, 2016Material Rupture; Material Rupture
AS REPORTED, DURING POWER INJECTION OF 5ML/SEC., THE PICC CATHETER TUBING BURST AT APPROXIMATELY THE 0CM MARK. THE PATIENT CONDITION WAS REPORTED AS "STABLE."
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 5966405 ↗MalfunctionSep 21, 2016Material Rupture; Material Rupture
AS REPORTED, DURING POWER INJECTION OF 5ML/SEC., THE PICC CATHETER TUBING BURST AT APPROXIMATELY THE 0CM MARK. THE PATIENT CONDITION WAS REPORTED AS "STABLE."
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 5966384 ↗MalfunctionAug 24, 2016Fluid/Blood Leak; Fracture; Fluid/Blood Leak; Fracture
PICC PLACED INTO A PATIENT PRIOR TO THE PATIENT GOING TO A CAT SCAN. WHILE GETTING CONTRAST INFUSED, THE PICC STARTED TO LEAK BETWEEN THE "0" MARK ON THE CATHETER AND THE WHITE SUTURE WING, RESULTING IN THE CATHETER BEING REMOVED. IT HAD ONLY BEEN IN PLACE FOR A FEW HOURS. THE PATIENT CONDITION WAS REPORTED AS "STABLE." THE USED CATHETER WAS DISCARDED AT THE HOSPITAL.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 5900558 ↗MalfunctionAug 24, 2016Fluid/Blood Leak; Fracture; Fluid/Blood Leak; Fracture
PICC PLACED INTO AN ICU PATIENT WITH SEVERAL DRIPS INFUSING. WITHIN A FEW HOURS THE PICC STARTED LEAKING ("PIN HOLE" LEAKAGE) BETWEEN THE "0" MARK ON THE CATHETER AND THE WHITE SUTURE WING, RESULTING IN THE CATHETER BEING REMOVED. THE PATIENT CONDITION WAS REPORTED AS "STABLE." THE USED CATHETER WAS DISCARDED AT THE HOSPITAL.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 5900551 ↗MalfunctionJun 21, 2016Fluid/Blood Leak; Fluid/Blood Leak
AS REPORTED, THE PATIENT EXPERIENCED LEAKING AT THE PICC INSERTION SITE, RESULTING IN THE PICC BEING REMOVED. THE USED DEVICE WAS RETURNED TO ANGIODYNAMICS FOR EVALUATION.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 5739720 ↗
Other Catheter,intravascular,therapeutic,long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC Catheter 3275118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days272 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗