BioFlo 75-063Recalls & reported problems
Navilyst Medical, Inc. · BioFlo PICC with ENDEXO and PASV Valve Technology (5F x 55cm, Single Lumen) - Maximal Barrier Dual Fenestrated Drape Nursing Kit. · Class 2 · Implant
All models of BioFlo (87)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI H965750630), received between Sep 20, 2021 and Dec 19, 2025 — classified as 2 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Material Puncture/Hole” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Material Puncture/Hole2
- Crack1
- Detachment of Device or Device Component1
- Fluid/Blood Leak1
- Infusion or Flow Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Peripheral Edema1
- Swelling/ Edema1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
InjuryDec 19, 2025Material Puncture/Hole; Material Puncture/Hole
AN END USER EXPERIENCED AN ISSUE WITH A CATHETER FROM BIOFLO PICC (V) 5FDL-55CM MAXIMAL BARRIER NURSING KIT W NIT AND DF DRAPE. IT WAS REPORTED THAT A HOLE DEVELOPED IN THE CATHETER, APPROXIMATELY 2 CM FROM THE INSERTION SITE. THE LEAK WAS NOTED 11 DAYS POST-INSERTION. THE PICC WAS REMOVED AFTER LEG SWELLING/EDEMA AND LEAKAGE OF FLUID WAS OBSERVED. THE PICC WAS REPLACED WITH ANOTHER OF THE SAME DEVICE.
Patient problems: Swelling/ Edema
MAUDE report 23850064 ↗InjuryDec 19, 2025Material Puncture/Hole; Material Puncture/Hole
AN END USER EXPERIENCED AN ISSUE WITH A CATHETER FROM BIOFLO PICC (V) 5FDL-55CM MAXIMAL BARRIER NURSING KIT W NIT AND DF DRAPE. IT WAS REPORTED THAT A HOLE DEVELOPED IN THE CATHETER, APPROXIMATELY 5-6 CM FROM THE HUB. THE LEAK WAS NOTED 6 DAYS POST-INSERTION. THE PICC WAS REMOVED AFTER THE SITE BECAME EDEMATOUS AND WAS LEAKING TPN (TOTAL PARENTERAL NUTRITION).
Patient problems: Peripheral Edema
MAUDE report 23850063 ↗MalfunctionApr 3, 2023Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak
IT WAS REPORTED FROM A TECH THAT THEY HAD A PICC FROM A TANDEM BIOFLO PICC (V) 5FDL-55CM MAXIMAL BARRIER NURSING KIT W NIT AND DF DRAPE KIT THAT WAS LEAKING FROM CRACK IN PURPLE LUMEN. THE VAT WAS NOTIFIED ON 03/08/2023 THAT ONE OF THE LUMENS WAS LEAKING. ON ASSESSMENT, THE PURPLE LUMEN WAS CRACKED JUST BELOW WHERE THE NEEDLESS CONNECTOR CONNECTS TO THE LUMEN. IT WAS DIFFICULT TO SEE, VISUALLY, UNLESS NEEDLESS CONNECTOR IS SCREWED IN JUST RIGHT, BUT WHEN SALINE IS HOOKED UP, IT SPRAYS OUT OF THE LUMEN, SHOWING INDEED THERE IS A CRACK. THE LAST TIME VAT ASSESSED THE LINE WAS ON 03/06/2023 AND THERE WERE NO ISSUES AT THAT TIME. SINCE 03/06/2023 THE PATIENT WAS ONLY ON 2 IV MEDS THAT WERE COMPATIBLE, SO THEY COULD HAVE BEEN JUST USING ONE LUMEN. WHEN ASKED TO ASSESS ON 03/08/2023, HER INFUSIONS WERE CONNECTED TO THE GREY LUMEN. AND IT WAS THE PATIENT WHO NOTED SHE FELT WET WHEN THE BEDSIDE RN ASSESSED THE DEVICE. THE RN DIDN'T NOTICE UNTIL PATIENT ASKED WHERE THE WETNESS WAS COMING FROM.THE PATIENT HAS BEEN INTERMITTENTLY CONFUSED AND LETHARGIC. THE REPORTING TECH WAS UNSURE EXACTLY WHEN THE CRACK HAPPENED (JUST THAT IT HAD TO BE BETWEEN 03/06/2023 AND 03/08/2023), AND WHAT/IF BEDSI
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16662510 ↗MalfunctionApr 18, 2022Infusion or Flow Problem; Infusion or Flow Problem
A NURSE REPORTED AN ISSUE WITH A BIOFLO PICC (V) 5FDL-55CM MAXIMAL BARRIER NURSING KIT. THE PICC LINE WAS PLACED IN THE PATIENT'S LEFT, MID-THIGH, FEMORAL ARTERY, DUE TO SVC COMPRESSION. DURING INITIAL PLACEMENT, ADEQUATE BLOOD RETURN WAS ACHIEVED; HOWEVER, APPROXIMATELY 3 HOURS AFTER PLACEMENT, THE PICC LINE WAS NOTED TO BE DIFFICULT TO FLUSH AND WOULD NOT GIVE BLOOD RETURN. BOTH THE PURPLE AND GREY LUMENS WERE AFFECTED. A KUB WAS PERFORMED AND CONFIRMED THAT THE TIP HAD NOT MIGRATED SO ALTEPASE WAS ORDERED FOR INSTILLING IN BOTH LUMENS. AFTER ALTEPASE, BOTH LINES HAD POSITIVE BLOOD RETURN; HOWEVER, THE GRAY LUMEN WAS STILL SLUGGISH, PROMPTING A SECOND DOSE OF ALTEPASE. BLOOD RETURN COULD NOT BE ESTABLISHED AFTER THE SECOND DOSE; THEREFORE, IT WAS ASSUMED THAT THERE WAS AN ISSUE WITH THE DEVICE, OTHER THAN, OR IN ADDITION TO, POSSIBLE CLOTTING. DIFFICULTY FLUSHING THE PORT AND ESTABLISHING BLOOD RETURN CONTINUED TO BE AN ISSUE. BLOOD RETURN COULD BE ESTABLISHED WITH PATIENT LYING IN RIGHT, RECUMBENT POSITION ; HOWEVER, IT WAS NOT POSSIBLE IF PATIENT WAS IN A LOW FOWLER'S POSITION. THE RN BELIEVED THAT THE LINE MAY BE FLIPPING OR TWISTING IN ON ITSELF, WHEN THE PATIENT STRAIGHTENS
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14135565 ↗MalfunctionSep 20, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
A VASCULAR ACCESS (VA) NURSE REPORTED AN ISSUE WITH A BIOFLO PICC (V) 5FDL-55CM MAXIMAL BARRIER NURSING KIT W NIT AND DF DRAPE PG. THE NURSE RECEIVED A CONSULT REQUEST TO EVALUATE THE PICC THAT WAS PLACED A FEW DAYS PRIOR. UPON ENTERING THE PATIENT'S ROOM, IT WAS NOTED THAT THE HUB WAS MISSING FROM THE CLEAR EXTENSION TUBING. THE NURSE FELT IT LOOKED LIKE IT WAS CUT OFF. THE CLEAR EXTENSION TUBING WAS COVERED BY A TRANSPORT MEDICAL TAPE. THE ANGIODYNAMICS CLINICAL SPECIALIST, WHO WAS PRESENT, ADVISED THE VA NURSE TO USE A KELLY CLAMP TO CLAMP THE TUBING TO PREVENT AIR EMBOLI AND/OR BLEEDING. THE PATIENT'S NURSE REPORTED FINDING THE LUMEN DETACHED AND WAS ABLE TO SHOW THE RESPONDING NURSE AND CS THE COMPONENT IN THE TRASH CAN, STILL CONNECTED TO THE IV TUBING AND IV FLUID BAG. IT WAS ALSO REPORTED THAT THE PATIENT'S SITTER STATED THAT THE PATIENT FREQUENTLY "TWISTS AND TURNS" IN BED. THE VA NURSE RECEIVED ORDERS TO EXCHANGE THE BIOFLO PICC 5FR DL PASV WITH A BIOFLO MIDLINE 4FR SL. THE REPLACEMENT PROCEDURE WAS COMPLETED WITH OUT ANY INCIDENT. BOTH THE DETACHED PICC HUB AND BIOFLO PICC 5FR DL PASV WERE SAVED. THE PATIENT DID NOT EXPERIENCE ANY ADVERSE EFFECTS OR HARM AS A RESULT OF T
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12496171 ↗
Other Catheter, intravascular, therapeutic, long-term greater than 30 days devices
- Groshong NXT ClearVue Catheter 7617405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,708 reports
- Groshong NXT ClearVue Catheter 7655405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 deaths1,536 reports
- PowerPICC Catheter 3174118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall402 reports
- L-Cath™ S/L PICC S/L 26GA (1.9F) x 30cmArgon Medical Devices, Inc. · Catheter,Intravascular,Therapeutic,Long-Term Greater Than 30 Days2 recalls245 reports
- PowerPICC SOLO2 Catheter 1194108DBard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days2 recalls216 reports
- Broviac 0600520Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days1 recall265 reports
- Groshong® Catheter 7717405Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days279 reports
- PowerPICC Catheter 3275118Bard Access Systems, Inc. · Catheter,intravascular,therapeutic,long-term greater than 30 days272 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page