Skip to content

NAMIC 90703062Recalls & reported problems

Navilyst Medical, Inc. · ARTERIAL EXTENSION SET 6" · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI H965907030620), received between Aug 17, 2018 and Aug 22, 2018 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Component Misassembled” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Component Misassembled2

Patient problems

  • Not Applicable2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 22, 2018Component Misassembled; Component Misassembled

    AS REPORTED BY ANGIODYNAMICS' DISTRIBUTOR IN (B)(4), BLUE NON-VENTED PORT PROTECTOR CAP CONNECTED TO STOPCOCK END PORT FITTING SHOULD HAVE BEEN A VENTED WHITE PORT PROTECTOR CAP. THE NON-VENTED CAP POSSIBLY COMPROMISES A STERILE PATHWAY FOR THIS DEVICE DURING THE STERILIZATION PROCESS. THIS ONE UNIT WAS FOUND DURING 100% INCOMING INSPECTION AT THE DISTRIBUTOR AND HAD NOT BEEN DISTRIBUTED TO A HOSPITAL.

    Patient problems: Not Applicable

    MAUDE report 7804936 ↗
  • MalfunctionAug 17, 2018Component Misassembled; Component Misassembled

    AS REPORTED BY ANGIODYNAMICS' DISTRIBUTOR IN (B)(4), BLUE NON-VENTED PORT PROTECTOR CAP CONNECTED TO STOPCOCK END PORT FITTING SHOULD HAVE BEEN A VENTED WHITE PORT PROTECTOR CAP. THE NON-VENTED CAP POSSIBLY COMPROMISES A STERILE PATHWAY FOR THIS DEVICE DURING THE STERILIZATION PROCESS. THIS ONE UNIT WAS FOUND DURING 100% INCOMING INSPECTION AT THE DISTRIBUTOR AND HAD NOT BEEN DISTRIBUTED TO A HOSPITAL.

    Patient problems: Not Applicable

    MAUDE report 7794319 ↗