Langston® dual lumen catheter 5540Recalls & reported problems
Vascular Solutions, Inc. · Each LANGSTON dual lumen catheter is indicated for delivery of contrast medium in angiographic studies and for simultaneous pressure measurement from two sites. This type of pressure measurement is useful in determining transvalvular, intravascular and intraventricular pressure gradients. · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI M20655400), received between Sep 28, 2017 and Dec 7, 2017 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Detachment Of Device Component” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Detachment Of Device Component2
- Detachment of Device or Device Component2
Patient problems
- No Consequences Or Impact To Patient3
- Vessel Or Plaque, Device Embedded In1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjuryDec 7, 2017Detachment Of Device Component; Detachment Of Device Component
LANGSTON CATHETER WAS USED THROUGH A 6 FR SHEATH IN THE RADIAL ARTERY. PATIENT DEVELOPED SEVERE ARTERIAL SPASM DURING REMOVAL OF THE LANGSTON. LANGSTON BECAME LOCKED IN THE RADIAL ARTERY IN THE FOREARM AREA DUE TO THE SEVERE SPASM. SEVERAL ATTEMPTS TO REMOVE IT WERE MADE. PATIENT WAS ALSO GIVEN NITRO, VERSED AND FENTANYL TO TRY TO REDUCE THE SPASM. UPON FINAL ATTEMPT TO REMOVE THE LANGSTON, THE CATHETER BROKE AND THE TIP REMAINED LODGED IN THE RADIAL ARTERY. PATIENT WAS SUBSEQUENTLY TRANSFERRED TO SURGERY FOR FOREIGN BODY REMOVAL.
Patient problems: Vessel Or Plaque, Device Embedded In
MAUDE report 7095640 ↗MalfunctionNov 10, 2017Detachment of Device or Device Component; Detachment of Device or Device Component
THERE WAS NOT A CRACK IN THE CATHETER. THERE WAS A BLUE PIECE PROTRUDING FROM THE CATHETER AND RIGHT WHERE THEY APPEAR TO BE SEPERATED THERE WAS BLOOD COMING OUT FROM IN -BETWEEN THE TWO CATHETER PIECES. NO ADVERSE PROBLEMS OCCURED. THE CATHETER WAS NOT SAVED. MDR 2134812-2017-00091 IS ALSO RELATED TO THIS REPORT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7018394 ↗MalfunctionNov 10, 2017Detachment of Device or Device Component; Detachment of Device or Device Component
DURING THE PROCEDURE EXPERIENCED AIR DRAWING BACK FROM THE ENGAGED CATHETER WHILE FLUSHING AND PREPPING TO ZERO THE LINE. AN UNUSUAL BLUE SEGMENT OF CATHETER WAS NOTED TO BE EXTENDING FROM THE YELLOW SEGMENT OF THE CATHETER. THE CATHETER WAS REMOVED AND ANOTHER LANGSTON WAS USED SUCCESSFULLY TO COMPLETE THE CASE WITHOUT INCIDENT. NO PATIENT IMPACT OR INJURY REPORTED. MDR 2134812-2017-00090 IS ASSOCIATED TO THIS REPORT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7018110 ↗MalfunctionSep 28, 2017Detachment Of Device Component; Detachment Of Device Component
DURING THE LV POWER INJECTION, THE YELLOW TUBING AT THE HUB OF THE CATHETER FRACTURED. 900 PSI WAS USED FOR THE INJECTION. ANOTHER LANGSTON CATHETER WAS USED TO COMPLETE THE PROCEDURE WITHOUT FURTHER INCIDENT. REFERENCE MDR 2134812-2017-00074 FOR ASSOCIATED REPORT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 6900588 ↗
Other Catheter, Intravascular, Diagnostic devices
- TrapEase Permanent Vena Cava Filter with the VisEase Angiographic Vessel Dilator 466P306AUCordis Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC6 deaths186 reports
- Dragonfly™ OPTIS™ C408646Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC4 deaths150 reports
- Dragonfly OpStar™ 1014651Lightlab Imaging, Inc. · CATHETER, INTRAVASCULAR, DIAGNOSTIC1 recall87 reports
- LANGSTON IPN913559Teleflex Incorporated · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls14 reports
- Trevo 90238Stryker Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls5 reports
- Expo™ H749085263011Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls3 reports
- Expo™ H74908526501Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls1 report
- Expo™ H749085263001Boston Scientific Corporation · CATHETER, INTRAVASCULAR, DIAGNOSTIC2 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page