The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI M20674440), received between Oct 25, 2018 and Mar 11, 2020 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Most reported problems

Device problems

  • Detachment of Device or Device Component1
  • Material Split, Cut or Torn1
  • Peeled/Delaminated1

Patient problems

  • No Known Impact Or Consequence To Patient2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 11, 2020Peeled/Delaminated; Material Split, Cut or Torn; Peeled/Delaminated; Material Split, Cut or Torn

    IT WAS REPORTED THAT A CTO OF THE RCA WITH TORTUOSITY AND HEAVY CALCIUM BURDEN WAS PERFORMED. THE BANDIT GUIDEWIRE NAVIGATED THE MICROCHANNEL WELL, HOWEVER, THE TURNPIKE SPIRAL WAS UNABLE TO GET THROUGH THE TIGHT CALCIFIED BEND AND CONTINUOUSLY BUCKLED AND THEY WERE UNABLE TO ADVANCE A BALLOON OVER THE BANDIT WIRE AS IT WAS GETTING STUCK ON THE WIRE. UPON REMOVAL OF THE WIRE IT WAS SEEN THAT THE POLYMER HAD BUNCHED UP. THEY THEN TOOK A DIFFERENT (SECOND) WIRE AND WERE UNABLE TO ADVANCE THIS WIRE THROUGH THE TIGHT LESION. FOR THEIR FINAL ATTEMPT THEY USED ANOTHER WIRE FOR A "HOLE-IN-ONE" TECHNIQUE WHERE THEY SPIN THE WIRE AS FAST AS THEY CAN AND HOPE THAT IT WILL GO THROUGH THE MICROCHANNEL. THEY HAD ISSUES ADVANCING THIS WIRE THROUGH THE SAME TIGHT CALCIFIED BEND AND THE TURNPIKE SPIRAL WAS STILL UNABLE TO MOVE BEYOND THIS BEND. DURING THE SPINNING OF THIS WIRE, IT SNAPPED. DUE TO THE CTO OF THE RCA AND THE HEAVY CALCIUM. SITE DECIDED TO LEAVE THIS PIECE OF WIRE IN THE VESSEL AND ABORTED THE CASE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 9819764 ↗
  • MalfunctionOct 25, 2018Detachment of Device or Device Component; Detachment of Device or Device Component

    WHILE PHYSICIAN WAS USING THE BANDIT GUIDEWIRE, IT KNUCKLED IN A STENT. THE WIRE BECAME KNOTTED WHEN HE KNUCKLED IT, AND THE COILS UNCOILED/SEPARATED FROM THE CORE WIRE. THE DEVICE REMAINED INTACT AND NO PIECES OF THE WIRE WERE LEFT IN THE PATIENT.

    Patient problems: No Known Impact Or Consequence To Patient