Penumbra Coil 400 4002C1135Recalls & reported problems
Penumbra, Inc. · PC400 Cplx Std, 11x35 · Class 2 · Implant
All models of Penumbra Coil 400 (120)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI M4244002C11350), received between Sep 23, 2016 and Apr 15, 2021 — classified as 2 injuries.
The most frequently reported specific device problem is “Break” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem1
- Break1
- Physical Resistance/Sticking1
- Unraveled Material1
Patient problems
- Dysphasia1
- No Clinical Signs, Symptoms or Conditions1
- Thrombus1
- Weakness1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryApr 15, 2021Break; Unraveled Material; Physical Resistance/Sticking; Break; Unraveled Material; Physical Resistance/Sticking
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE INTERNAL CAROTID ARTERY (ICA) USING PENUMBRA COIL 400S (PC400), PX SLIM DELIVERY MICROCATHETER (PX SLIM), AND A NON-PENUMBRA EMBOLIZATION DEVICE. DURING THE PROCEDURE, THE PHYSICIAN PLACED A PX SLIM BEHIND THE EMBOLIZATION DEVICE AND PLACED TWO PC400S IN THE ANEURYSM. WHILE ADVANCING ANOTHER PC400 AS THE THIRD COIL, THE PHYSICIAN EXPERIENCED RESISTANCE INSIDE THE ANEURYSM FROM THE PACKING DENSITY OF THE PREVIOUSLY PLACED PC400S. THE PHYSICIAN THEN PULLED THE PC400 IN AN ATTEMPT TO REMOVE IT; HOWEVER, THE PC400 HAD UNINTENTIONALLY DETACHED WITH APPROXIMATELY FIVE CENTIMETERS OF THE PC400 IN THE ANEURYSM AND THE REST OF THE COIL WAS WITHIN THE MICROCATHETER. SUBSEQUENTLY, WHILE PULLING THE PC400, THE COIL BECAME STRETCHED. THEREFORE, THE PHYSICIAN USED A SNARE DEVICE TO REMOVE THE REMAINING PC400 OUT OF THE MICROCATHETER AND AORTA; THEN USED ASPIRATION TO SUCCESSFULLY REMOVE THE REMAINING PC400. THE PHYSICIAN DECIDED TO END THE PROCEDURE AT THIS POINT DUE TO ACCESS WITH THE PX SLIM WAS LOST. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11675831 ↗InjurySep 23, 2016Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT UNDERWENT A COIL EMBOLIZATION PROCEDURE IN THE LEFT PERIOPHTHALMIC ARTERY USING PENUMBRA COIL 400 (PC400'S) WITH NO COMPLICATIONS DURING THE PROCEDURE. HOWEVER, SIX HOURS POST-PROCEDURE, THE PATIENT DEVELOPED APHASIA, WHICH PROGRESSED TO RIGHT-SIDED WEAKNESS. A FOLLOW-UP ANGIOGRAPHY WAS PERFORMED AND SHOWED DECREASE BLOOD FLOW INTO THE MIDDLE CEREBRAL ARTERY (MCA) AND A THROMBUS IN THE M3 BRANCH OF THE LEFT INTERNAL CAROTID ARTERY (LICA) WAS NOTICED. THEREFORE, THE PATIENT UNDERWENT A THROMBECTOMY PROCEDURE, WHICH PARTIALLY REMOVED THE THROMBUS; HOWEVER, A FRAGMENT OF THE THROMBUS WENT DOWNSTREAM AND COULD NOT BE REMOVED. THE PATIENT WAS STILL ADMITTED IN THE HOSPITAL. AS OF (B)(6) 2016, THE PATIENT WAS IN STABLE CONDITION. THIS SERIOUS ADVERSE EVENT WAS REPORTED AS AN ISCHEMIC STROKE AND WAS ADJUDICATED TO HAVE A PROBABLE RELATIONSHIP TO THE PC400'S, THE ANGIOGRAPHIC PROCEDURE AND THE ANEURYSM MALFORMATION.
Patient problems: Dysphasia; Thrombus; Weakness
MAUDE report 5975393 ↗
Other DEVICE, NEUROVASCULAR EMBOLIZATION devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page