Penumbra Coil 400 4004C0509Recalls & reported problems
Penumbra, Inc. · PC400 Cplx XSft, 5x9 · Class 2 · Implant
All models of Penumbra Coil 400 (120)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI M4244004C05090), received between Sep 18, 2015 and Apr 1, 2020 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Premature Separation” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem2
- Premature Separation2
Patient problems
- No Consequences Or Impact To Patient2
- Ischemia1
- Thrombus1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 1, 2020Premature Separation; Premature Separation
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE LEFT INTERNAL CAROTID ARTERY (ICA) USING A PENUMBRA COIL 400 (PC400) AND A PX SLIM DELIVERY MICROCATHETER (PX SLIM). DURING THE PROCEDURE, WHILE ADVANCING THE PC400 THROUGH THE PX SLIM, THE PC400 UNINTENTIONALLY DETACHED INSIDE THE PX SLIM. THEREFORE, THE PHYSICIAN REMOVED THE PX SLIM CONTAINING THE DETACHED PC400 AND THEN REMOVED THE PC400. THE PROCEDURE WAS COMPLETED USING A NEW PC400 AND THE SAME PX SLIM. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 9911055 ↗MalfunctionNov 24, 2018Premature Separation; Premature Separation
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE M1 SEGMENT OF THE MIDDLE CEREBRAL ARTERY (MCA) USING PENUMBRA COILS 400 (PC400). DURING THE PROCEDURE, THE PHYSICIAN FELT RESISTANCE WHILE ADVANCING A PC400 WITHIN A PX SLIM DELIVERY MICROCATHETER (PX SLIM) AND, CONSEQUENTLY, THE PC400 UNINTENTIONALLY DETACHED WITHIN THE PX SLIM. THEREFORE, THE PHYSICIAN REMOVED THE PX SLIM WITH THE PC400 INSIDE. THE PROCEDURE WAS THEN COMPLETED USING A NEW MICROCATHETER AND NEW COILS. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8100122 ↗InjuryJul 26, 2016Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE RIGHT INTERNAL CAROTID ARTERY-POSTERIOR COMMUNICATING ARTERY (ICA-PCOM) USING PENUMBRA COIL 400 COILS (PC400 COILS). DURING THE PROCEDURE, AFTER EIGHT PC400 COILS WERE DEPLOYED AND DETACHED IN THE ANEURYSM, MINIMAL THROMBUS FORMATION WAS NOTED AT THE ANEURYSM NECK. THEREFORE, THE PATIENT WAS ADMINISTERED FOUR THOUSAND UNITS OF HEPARIN AND SUBSEQUENTLY, THE THROMBUS RESOLVED. THE REPORTED MINIMAL THROMBUS FORMATION WAS ADJUDICATED TO HAVE A DEFINITE RELATIONSHIP TO THE PC400 COIL SYSTEM, AN UNRELATED RELATIONSHIP TO THE PROCEDURE AND A POSSIBLE RELATIONSHIP TO THE DISEASE STATE.
Patient problems: Thrombus
MAUDE report 5824132 ↗InjurySep 18, 2015Adverse Event Without Identified Device or Use Problem
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE INTERNAL CAROTID ARTERY USING PENUMBRA COIL 400 (PC400) COILS. THE PROCEDURE WAS SUCCESSFUL WITH NO COMPLICATIONS. HOWEVER, THE NEXT DAY, MAGNETIC RESONANCE IMAGING REVEALED SMALL FOCI OF SILENT ISCHEMIA IN THE LEFT CEREBRAL HEMISPHERE OF THE PATIENT. THE RELATIONSHIP OF THE PC400 COILS TO THE ISCHEMIA IS UNCERTAIN; HOWEVER THE RELATIONSHIP OF THE PROCEDURE TO THE ISCHEMIA IS DEFINITE.
Patient problems: Ischemia
MAUDE report 5089326 ↗
Other DEVICE, NEUROVASCULAR EMBOLIZATION devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page