Penumbra Coil 400 4004C1440Recalls & reported problems
Penumbra, Inc. · PC400 Cplx XSft, 14x40 · Class 2 · Implant
All models of Penumbra Coil 400 (120)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI M4244004C14400), received between Nov 10, 2017 and Dec 29, 2020 — classified as 3 malfunctions.
The most frequently reported specific device problem is “Bent” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Bent1
- Break1
- Device Handling Problem1
- Difficult or Delayed Separation1
- Kinked1
- Physical Resistance1
- Physical Resistance/Sticking1
Patient problems
- No Consequences Or Impact To Patient3
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 29, 2020Physical Resistance/Sticking; Difficult or Delayed Separation; Physical Resistance/Sticking; Difficult or Delayed Separation
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE INTERNAL CAROTID ARTERY (ICA) USING PENUMBRA COILS 400 (PC400S), A PENUMBRA COIL 400 DETACHMENT HANDLE (HANDLE), AND PX SLIM DELIVERY MICROCATHETER (PX SLIM). DURING THE PROCEDURE, THE PHYSICIAN PLACED SEVEN COILS IN THE TARGET VESSEL. THE EIGHTH COIL, A PC400, WAS THEN BEING PLACED. WHILE ATTEMPTING TO PUSH THE PUSHER ASSEMBLY OF THE PC400, THE PHYSICIAN FELT RESISTANCE IN THE MICROCATHETER DESPITE THE PC400 WAS PROPERLY SHAPED IN THE ANEURYSM. WHEN THE PHYSICIAN ATTEMPTED TO DETACH THE PC400 USING THE HANDLE, THE PC400 DID NOT DETACH. SUBSEQUENTLY, THE PHYSICIAN RECAPTURED THE PC400 INTO THE PX SLIM. IT WAS OBSERVED THAT THE PC400 UNINTENTIONALLY DETACHED IN THE PX SLIM. THEREFORE, THE PX SLIM CONTAINING THE DETACHED PC400 WAS REMOVED. THE PROCEDURE WAS COMPLETED USING TWO NEW COILS AND THE SAME HANDLE AND PX SLIM. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 11085867 ↗MalfunctionApr 25, 2018Kinked; Device Handling Problem; Kinked; Device Handling Problem
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE FOR A VESSEL SACRIFICE USING PENUMBRA COILS 400 (PC400¿S) AND A PENUMBRA COIL 400 DETACHMENT HANDLE (HANDLE). DURING THE PROCEDURE, WHILE ADVANCING A PC400 THROUGH A PX SLIM DELIVERY MICROCATHETER (PX SLIM), THE HOSPITAL FELLOW WAS TOO AGGRESSIVE AND INADVERTENTLY BENT THE PC400 PUSHER ASSEMBLY; THEREFORE, IT WAS REMOVED. IT WAS REPORTED THAT ANOTHER PC400 PUSHER ASSEMBLY WAS ALSO BENT WHILE ADVANCING THE PC400 THROUGH THE PX SLIM; THEREFORE, IT WAS REMOVED. THE PROCEDURE CONTINUED USING ADDITIONAL COILS. AROUND THE 30TH COIL, IT WAS REPORTED THAT BLOOD MIGHT HAVE GOTTEN INSIDE THE HANDLE AND ITS THE TRIGGER BECAME A BIT STICKY; THEREFORE, THE HANDLE WAS SET ASIDE. THE PROCEDURE WAS COMPLETED USING THE SAME PX SLIM, A NEW HANDLE AND ADDITIONAL PC400'S. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7461671 ↗MalfunctionNov 10, 2017Bent; Break; Physical Resistance; Bent; Break; Physical Resistance
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE TO TREAT A RUPTURED ANEURYSM IN THE ANTERIOR COMMUNICATING ARTERY (ACOM) USING PENUMBRA COILS 400 (PC400S). DURING THE PROCEDURE, WHILE ATTEMPTING TO INTRODUCE A PC400 INTO THE PX SLIM DELIVERY MICROCATHETER (PX SLIM), THE PHYSICIAN EXPERIENCED RESISTANCE AND SUBSEQUENTLY, THE PC400 PUSHER ASSEMBLY BECAME BENT THEN BROKE; THEREFORE, IT WAS REMOVED. THE PROCEDURE WAS COMPLETED USING 11 PC400S AND THE SAME PX SLIM. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7022356 ↗
Other DEVICE, NEUROVASCULAR EMBOLIZATION devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page