Skip to content

Penumbra Coil 400 4006U0305Recalls & reported problems

Penumbra, Inc. · PC400 Curve Sft, 3x5 · Class 2 · Implant

All models of Penumbra Coil 400 (120)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
3
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI M4244006U03050), received between Dec 21, 2015 and Apr 10, 2017 — classified as 1 injury, 2 malfunctions.

The most frequently reported specific device problem is “Defective Device” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Defective Device1
  • Deformation Due to Compressive Stress1
  • Detachment Of Device Component1
  • Out-Of-Box Failure1

Patient problems

  • Neurological Deficit/Dysfunction1
  • No Consequences Or Impact To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 10, 2017Detachment Of Device Component; Out-Of-Box Failure; Defective Device; Detachment Of Device Component; Out-Of-Box Failure; Defective Device

    DURING PREPARATION FOR A COIL EMBOLIZATION PROCEDURE, THE HOSPITAL STAFF NOTICED THAT A PENUMBRA COIL 400 (PC400) HAD UNINTENTIONALLY DETACHED FROM ITS PUSHER ASSEMBLY UPON REMOVAL FROM THE DISPENSER HOOP. THE DETACHED PC400 WAS FOUND PRIOR TO USE AND THEREFORE, WAS NOT USED FOR THE PROCEDURE. THE PROCEDURE WAS COMPLETED USING A NEW PC400.

    Patient problems: No Patient Involvement

    MAUDE report 6478489 ↗
  • MalfunctionJun 30, 2016Deformation Due to Compressive Stress; Deformation Due to Compressive Stress

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE USING PENUMBRA COILS 400 (PC400). DURING THE PROCEDURE, WHILE ADVANCING A PC400 THROUGH A PX SLIM DELIVERY MICROCATHETER (PX SLIM), THE PC400 UNINTENTIONALLY DETACHED WITHIN THE PX SLIM; THEREFORE, IT WAS REMOVED. IT WAS ALSO REPORTED THAT A PC400 PUSHER ASSEMBLY WAS FOUND BENT WHILE REMOVING IT FROM THE PACKAGING. THERE WAS NO REPORT OF ADVERSE EFFECT TO THE PATIENT.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 5761660 ↗
  • InjuryDec 21, 2015Adverse Event Without Identified Device or Use Problem

    THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION IN THE RIGHT POSTERIOR INTERNAL CAROTID ARTERY (ICA) USING PENUMBRA COIL 400 COILS (PC400 COILS) AND A PX SLIM DELIVERY MICROCATHETER (PX SLIM). DURING THE PROCEDURE, THE PHYSICIAN DEPLOYED AND DETACHED SIX PC400 COILS INTO THE ANEURYSM WITH NO IMMEDIATE COMPLICATIONS USING THE PX SLIM. THE FOLLOWING DAY, THE PATIENT EXPERIENCED A SEVERE RIGHT POSTERIOR LIMBIC STROKE. THE EVENT REMAINED UNRESOLVED AND NO ACTIONS WERE TAKEN. THE REPORTED RIGHT POSTERIOR LIMBIC STROKE WAS ADJUDICATED TO HAVE AN UNCERTAIN RELATIONSHIP TO THE PC400 COILS AND A POSSIBLE RELATIONSHIP TO THE PX SLIM, ANGIOGRAPHIC PROCEDURE AND THE ANEURYSM STATE.

    Patient problems: Neurological Deficit/Dysfunction

    MAUDE report 5317240 ↗

Other DEVICE, NEUROVASCULAR EMBOLIZATION devices

All DEVICE, NEUROVASCULAR EMBOLIZATION devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page