RUBY Coil RBY2C0612-ARecalls & reported problems
Penumbra, Inc. · PC400 Cplx Std, 6x12 · Class 2 · Implant
All models of RUBY Coil (90)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI M424RBY2C06120), received between Dec 16, 2015 and May 8, 2018 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Detachment Of Device Component” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Detachment Of Device Component3
- Bent1
- Device Handling Problem1
- Failure to Advance1
- Mechanical Jam1
- Out-Of-Box Failure1
Patient problems
- No Consequences Or Impact To Patient4
- No Patient Involvement1
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 8, 2018Detachment Of Device Component; Detachment Of Device Component
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE IN THE GONADAL ARTERY USING RUBY COILS. DURING THE PROCEDURE, THE PHYSICIAN LOADED A RUBY COIL INTO A NON-PENUMBRA MICROCATHETER AND THE RUBY COIL UNINTENTIONALLY DETACHED; THEREFORE, THE PHYSICIAN REMOVED THE MICROCATHETER CONTAINING THE DETACHED COIL AND REMOVED THE COIL FROM THE MICROCATHETER. THE PROCEDURE WAS COMPLETED USING THE SAME MICROCATHETER AND A NEW RUBY COIL. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7495002 ↗MalfunctionFeb 22, 2018Detachment Of Device Component; Mechanical Jam; Detachment Of Device Component; Mechanical Jam
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE TO TREAT PORTAL HYPERTENSION AND ESOPHAGEAL VARICES USING A RUBY COIL. DURING THE PROCEDURE, FOLLOWING THE PLACEMENT OF THE STENT, A 5F GLIDE CATHETER WAS DELIVERED OVER A GUIDEWIRE AND A NON-PENUMBRA COIL WAS DEPLOYED IN AN ATTEMPT TO EMBOLIZE THE VARICES. THE PHYSICIAN THEN DELIVERED A NON-PENUMBRA MICROCATHETER THROUGH THE GLIDE CATHETER AND ATTEMPTED TO USE A RUBY COIL. WHILE ADVANCING THE COIL THROUGH THE MICROCATHETER, THE RUBY COIL UNINTENTIONALLY DETACHED INSIDE THE MICROCATHETER. IN AN ATTEMPT TO REMOVE THE DETACHED COIL, THE PHYSICIAN PULLED THE MICROCATHETER THROUGH THE GLIDE CATHETER AND THE COIL BECAME LODGED IN THE GLIDE CATHETER. NUMEROUS ATTEMPTS WERE MADE TO FLUSH THE COIL OUT THE GLIDE CATHETER AND INTO THE VARICES; HOWEVER, THE COIL WOULD NOT ADVANCE. THE PHYSICIAN THEN DELIVERED A GUIDE WIRE THROUGH THE GLIDE CATHETER AND ATTEMPTED TO PUSH THE COIL OUT OF THE GLIDE CATHETER; HOWEVER, THE COIL WOULD STILL NOT ADVANCE. THEREFORE, THE DIAGNOSTIC CATHETER WAS REMOVED WITH THE COIL AND THE PROCEDURE WAS COMPLETED. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 7290432 ↗MalfunctionAug 19, 2016Bent; Device Handling Problem; Bent; Device Handling Problem
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE USING RUBY COILS. DURING THE PROCEDURE, WHILE ADVANCING A RUBY COIL THROUGH ANOTHER MANUFACTURER¿S MICROCATHETER, RESISTANCE WAS NOT ENCOUNTERED; HOWEVER, THE TECHNOLOGIST INADVERTENTLY BENT THE RUBY COIL PUSHER ASSEMBLY. THEREFORE, THE RUBY COIL WAS REMOVED AND THE PROCEDURE WAS COMPLETED USING A NEW RUBY COIL AND THE SAME MICROCATHETER. IT SHOULD BE NOTED THAT A ROTATING HEMOSTASIS VALVE (RHV) AND A FLUSH BAG WERE NOT USED DURING THE PROCEDURE. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 5889824 ↗MalfunctionFeb 25, 2016Failure to Advance; Failure to Advance
THE PATIENT WAS UNDERGOING A COIL EMBOLIZATION PROCEDURE USING RUBY COILS. DURING THE PROCEDURE, THE PHYSICIAN DEPLOYED AND DETACHED SEVERAL RUBY COILS INTO THE ANEURYSM USING A PX SLIM DELIVERY MICROCATHETER (PX SLIM). WHILE THE PHYSICIAN WAS ADVANCING A NEW RUBY COIL THROUGH THE PX SLIM, THE RUBY COIL ADVANCED ABOUT 20 CENTIMETERS THROUGH THE PX SLIM AND THEN STOPPED ADVANCING. THE RUBY COIL WAS REMOVED WITHOUT ANY ISSUES AND THE PROCEDURE WAS COMPLETED USING NEW RUBY COILS AND THE SAME PX SLIM. THERE WAS NO REPORT OF AN ADVERSE EFFECT TO THE PATIENT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 5462020 ↗MalfunctionDec 16, 2015Detachment Of Device Component; Out-Of-Box Failure
DURING PREPARATION FOR A COIL EMBOLIZATION PROCEDURE, THE STAFF NOTICED THAT THE RUBY COIL WAS DETACHED UPON REMOVAL FROM ITS HOOP. THE DETACHED RUBY COIL WAS FOUND PRIOR TO USE AND WAS NOT USED FOR THE PROCEDURE. THE PROCEDURE WAS COMPLETED USING A NEW RUBY COIL.
Patient problems: No Patient Involvement
MAUDE report 5304644 ↗
Other DEVICE, NEUROVASCULAR EMBOLIZATION devices
- POD RBYPODJ60-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION355 reports
- Penumbra System SCH1Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION295 reports
- RUBY Coil RBY4C0860-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION276 reports
- Smart Coil 400SMTHXSFT0208Penumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION165 reports
- Penumbra System RH1-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION154 reports
- Optima Coil System OPTI0208CSS10Balt USA LLC · Device, Neurovascular Embolization3 recalls3 reports
- POD RBYPOD6-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION148 reports
- POD RBYPODJ30-APenumbra, Inc. · DEVICE, NEUROVASCULAR EMBOLIZATION132 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page