CONSERVE® 38AC3600Recalls & reported problems
Microport Orthopedics Inc. · CONSERVE® TOTAL A-CLASS® 36MM FEMORAL HEAD · Class 3 · Implant
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI M68438AC36001), received between Feb 21, 2020 and Apr 5, 2023 — classified as 2 injuries.
The most frequently reported specific device problem is “Corroded” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Corroded1
- Detachment of Device or Device Component1
- Material Disintegration1
- Naturally Worn1
Patient problems
- Pain2
- Metal Related Pathology1
- Reaction1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryApr 5, 2023Corroded; Detachment of Device or Device Component; Naturally Worn; Corroded; Detachment of Device or Device Component; Naturally Worn
ALLEGEDLY, PLAINTIFF UNDERWENT A RIGHT-SIDED REVISION OF THE ACETABULAR LINER/FEMORAL HEAD AND NECK. THE PROFEMUR TOTAL HIP SYSTEM HAS FAILED DUE TO CORROSION RELATED ADVERSE TISSUE REACTIONS AND THE DEVICE HAS BEEN SURGICALLY REMOVED, PLAINTIFF PRESENTED PAIN AND DISASSOCIATION OF THE ACETABULAR POLUETHYLENE COMPONENT WITHOUT DISLOCATION, MODULAR HEAD (PHAC1202) AND COBALT CHROME HEAD (26000025) WERE IMPLANTED. PRODUCTS NOT REVISED: PRODUCT ID: 18080303, CANCELLOUS SELF-TAPPING 6.5MM BONE SCREW 3.0CM LENGTH, LOT: 1558083, QTY: 1. PRODUCT ID: PPW38372, PROFEMUR® 115MM STRAIGHT FULLY COATED CYL/PS SZ 12, LOT: 1571390, QTY: 1. PRODUCT ID: PPW38354, "PROFEMUR® R" PROX BODY PLASMA SPRAY X-SMALL MODULAR LOT: 1562181, QTY: 1. PRODUCT ID: DSBFGE54, DYNASTY® BF SHELL 54MM GROUP E LOT: 1565510, QTY: 1 PRODUCT ID: 18080303, CANCELLOUS SELF-TAPPING 6.5MM BONE SCREW 3.0CM LENGTH LOT: 1558084, QTY: 1.
Patient problems: Metal Related Pathology; Pain
MAUDE report 16685749 ↗InjuryFeb 21, 2020Material Disintegration; Material Disintegration
ALLEGEDLY, PATIENT PRESENTED ELEVATED COBALT ION LEVELS AND PAIN. THE DEVICE FAILED DUE TO CORROSION AT THE NECK-STEM JUNCTION OF THE DEVICE, CORROSION ON THE OBLONG TAPER WHERE THE NECK WAS SEATED IN THE POCKET OF THE PROFEMUR TITANIUM STEM, WHICH CAUSED ADDUCTOR DAMAGE, SOFT TISSUE REACTION AND PSEUDOTUMOR REACTION. DURING THE REVISION THE SURGEON REMOVED THE FAILED COMPONENTS OF THE PROFEMUR TOTAL HIP SYSTEM.
Patient problems: Pain; Reaction
MAUDE report 9739206 ↗
Other Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular Component) devices
- Omnifit 2001-2246Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained (metal cemented acetabular component)1 recall0 reports
- Taperloc Complete Micro Primary Femoral 51-107150Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained (metal cemented acetabular component)41 reports
- Taperloc Complete Micro Primary Femoral 51-107140Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained (metal cemented acetabular component)36 reports
- Dynasty DSBFGD52Microport Orthopedics Inc. · Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular Component)12 reports
- CONSERVE® 38AM4600Microport Orthopedics Inc. · Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular Component)8 reports
- CONSERVE® 38AM4804Microport Orthopedics Inc. · Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular Component)7 reports
- CONSERVE® 38AM4404Microport Orthopedics Inc. · Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular Component)6 reports
- Dynasty DSBFGF56Microport Orthopedics Inc. · Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular Component)5 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.