CONSERVE® 38AC3600Recalls & reported problems
Microport Orthopedics Inc. · CONSERVE® TOTAL A-CLASS® 36MM FEMORAL HEAD · Class 3 · Implant
All models of CONSERVE (74)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI M68438AC36001), received between Feb 21, 2020 and Apr 5, 2023 — classified as 2 injuries.
The most frequently reported specific device problem is “Corroded” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Corroded1
- Detachment of Device or Device Component1
- Material Disintegration1
- Naturally Worn1
Patient problems
- Pain2
- Metal Related Pathology1
- Reaction1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryApr 5, 2023Corroded; Detachment of Device or Device Component; Naturally Worn; Corroded; Detachment of Device or Device Component; Naturally Worn
ALLEGEDLY, PLAINTIFF UNDERWENT A RIGHT-SIDED REVISION OF THE ACETABULAR LINER/FEMORAL HEAD AND NECK. THE PROFEMUR TOTAL HIP SYSTEM HAS FAILED DUE TO CORROSION RELATED ADVERSE TISSUE REACTIONS AND THE DEVICE HAS BEEN SURGICALLY REMOVED, PLAINTIFF PRESENTED PAIN AND DISASSOCIATION OF THE ACETABULAR POLUETHYLENE COMPONENT WITHOUT DISLOCATION, MODULAR HEAD (PHAC1202) AND COBALT CHROME HEAD (26000025) WERE IMPLANTED. PRODUCTS NOT REVISED: PRODUCT ID: 18080303, CANCELLOUS SELF-TAPPING 6.5MM BONE SCREW 3.0CM LENGTH, LOT: 1558083, QTY: 1. PRODUCT ID: PPW38372, PROFEMUR® 115MM STRAIGHT FULLY COATED CYL/PS SZ 12, LOT: 1571390, QTY: 1. PRODUCT ID: PPW38354, "PROFEMUR® R" PROX BODY PLASMA SPRAY X-SMALL MODULAR LOT: 1562181, QTY: 1. PRODUCT ID: DSBFGE54, DYNASTY® BF SHELL 54MM GROUP E LOT: 1565510, QTY: 1 PRODUCT ID: 18080303, CANCELLOUS SELF-TAPPING 6.5MM BONE SCREW 3.0CM LENGTH LOT: 1558084, QTY: 1.
Patient problems: Metal Related Pathology; Pain
MAUDE report 16685749 ↗InjuryFeb 21, 2020Material Disintegration; Material Disintegration
ALLEGEDLY, PATIENT PRESENTED ELEVATED COBALT ION LEVELS AND PAIN. THE DEVICE FAILED DUE TO CORROSION AT THE NECK-STEM JUNCTION OF THE DEVICE, CORROSION ON THE OBLONG TAPER WHERE THE NECK WAS SEATED IN THE POCKET OF THE PROFEMUR TITANIUM STEM, WHICH CAUSED ADDUCTOR DAMAGE, SOFT TISSUE REACTION AND PSEUDOTUMOR REACTION. DURING THE REVISION THE SURGEON REMOVED THE FAILED COMPONENTS OF THE PROFEMUR TOTAL HIP SYSTEM.
Patient problems: Pain; Reaction
MAUDE report 9739206 ↗
FDA clearance and classification
The FDA regulates product code JDL (prosthesis, hip, semi-constrained (metal cemented acetabular component)) under 21 CFR 888.3320 as a Class III device. 21 CFR 888.3320 ↗
Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the CONSERVE® 38AC3600
Has the CONSERVE® 38AC3600 been recalled?
No FDA recall record in our data names this exact device identifier (DI M68438AC36001), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the CONSERVE® 38AC3600?
The FDA's MAUDE database lists 2 adverse event reports for this device, received between Feb 21, 2020 and Apr 5, 2023. In those reports, the most common specific device problems are “Corroded” (1), “Detachment of Device or Device Component” (1) and “Material Disintegration” (1); the most common patient problems are “Pain” (2), “Metal Related Pathology” (1) and “Reaction” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the CONSERVE® 38AC3600?
None of the reports are classified as deaths, 2 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the CONSERVE® 38AC3600 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the CONSERVE® 38AC3600, and is it still sold?
The FDA's device database lists Microport Orthopedics Inc. as the company responsible for it (brand CONSERVE®, model 38AC3600). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular Component) devices
- Omnifit 2001-2246Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained (metal cemented acetabular component)1 recall0 reports
- Taperloc Complete Micro Primary Femoral 51-107150Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained (metal cemented acetabular component)40 reports
- Taperloc Complete Micro Primary Femoral 51-107140Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained (metal cemented acetabular component)36 reports
- Dynasty DSBFGD52Microport Orthopedics Inc. · Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular Component)12 reports
- CONSERVE® 38AM4600Microport Orthopedics Inc. · Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular Component)8 reports
- CONSERVE® 38AM4804Microport Orthopedics Inc. · Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular Component)7 reports
- CONSERVE® 38AM4404Microport Orthopedics Inc. · Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular Component)6 reports
- TRANSCEND® 26000025Microport Orthopedics Inc. · Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular Component)5 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page