Skip to content

CONSERVE® 38AC4200Recalls & reported problems

Microport Orthopedics Inc. · CONSERVE® TOTAL A-CLASS® 42MM FEMORAL HEAD · Class 3 · Implant

Adverse event reports
4
Deaths reported
0
Injuries reported
4
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI M68438AC42001), received between Jan 29, 2020 and Apr 6, 2022 — classified as 4 injuries.

The most frequently reported specific device problem is “Corroded” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Corroded1
  • Device Dislodged or Dislocated1
  • Insufficient Device Problem Information1
  • Loss of Osseointegration1
  • Material Fragmentation1
  • Patient Device Interaction Problem1

Patient problems

  • Metal Related Pathology2
  • No Information2
  • Pain1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 6, 2022Material Fragmentation; Patient Device Interaction Problem

    ALLEGEDLY, THE PATIENT HAD METAL-ON-METAL HIP ARTHROPLASTY THAT HAS BEEN PAINFUL FOR 4 YEARS. AFTER REVIEWING HIS RADIOGRAPHS AND OPERATIVE REPORT THE CHROMIC COBALT LEVELS WERE ALL SIGNIFICANTLY ELEVATED. THERE WAS MILD EVIDENCE OF CORROSION AT THE TRUNNION OF THE BASE OF THE MODULAR NECK HOWEVER IT WAS VERY MINIMAL AND ON THE INSIDE OF THE FEMORAL COMPONENT IT WAS ALMOST COMPLETELY ABSENT. THE FEMORAL COMPONENTS HAD METALLIC DEBRIS.

    Patient problems: Metal Related Pathology; Pain

    MAUDE report 14025956 ↗
  • InjuryMar 29, 2021Corroded; Loss of Osseointegration; Corroded; Loss of Osseointegration

    ALLEGEDLY, ON OR ABOUT (B)(6) 2020 IT WAS DISCOVERED THAT THE DEVICE FAILED DUE TO METAL DEBRIS, CORROSION AND RESULTANT METAL IONS, DUE TO THE METAL ON METAL DESIGN BETWEEN THE ARTICULATING SURFACES. THIS PATIENT HAD METAL ON POLYETHYLENE ARTICULATING COMPONENTS. IT IS ALSO ALLEGED THAT THE FEMORAL STEM WAS LOOSE AND HAD AREAS OF CORROSION BETWEEN THE MODULAR NECK AND FEMORAL STEM.

    Patient problems: Metal Related Pathology

    MAUDE report 11585522 ↗
  • InjuryOct 30, 2020Device Dislodged or Dislocated; Device Dislodged or Dislocated

    ALLEGEDLY, PATIENT WAS REVISED DUE TO HEAD DISLOCATED. ADDITIONAL INFORMATION RECEIVED ON 10/08/2020 FROM (B)(6): ADDING PRODUCT INFORMATION. SECOND REVISION CAPTURED UNDER INCIDENT NUMBER: (B)(4).

    Patient problems: No Information

    MAUDE report 10766610 ↗
  • InjuryJan 29, 2020Insufficient Device Problem Information; Insufficient Device Problem Information

    ALLEGEDLY, PATIENT WAS REVISED DUE TO UNKNOWN REASONS.

    Patient problems: No Information

    MAUDE report 9641504 ↗

Other Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component) devices

All Prosthesis, Hip, Semi-Constrained (Metal Uncemented Acetabular Component) devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.