CONSERVE® 38CH5410Recalls & reported problems
Microport Orthopedics Inc. · CONSERVE® TOTAL 54MM BCH® FEMORAL HEAD · Class 3 · Implant
All models of CONSERVE (74)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI M68438CH54101), received between May 12, 2022 and May 12, 2022 — classified as 1 injury.
The most frequently reported specific device problem is “Corroded” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Most reported problems
Device problems
- Corroded1
Patient problems
- Metal Related Pathology1
- Neuropathy1
- Pain1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryMay 12, 2022Corroded
ALLEGEDLY PT. NOW HAS ELEVATED METAL IONS. ORIGINALLY PART OF IDE / CERAMIC ON METAL ARTICULATION SYSTEM CLINICAL STUDY BUT NOT ANY LONGER. ADDITIONAL INFORMATION RECEIVED ON APRIL 19, 2022: PATIENT DEVELOPED ADVERSE SYMPTOMS INCLUDING, BUT NOT LIMITED TO, PROGRESSIVELY WORSENING PAIN IN AND AROUND HIS RIGHT HIP JOINT, AS WELL AS NEUROLOGIC SYMPTOMS INCLUDING PERIPHERAL NEUROPATHY, TINNITUS, SOME BALANCE/COORDINATION ISSUES, A HEAL FRACTURE AS A DIRECT RESULT OF THOSE BALANCE/COORDINATION ISSUES, AND B 12 DEFICIENCY. LABORATORY TEST RESULTS REPORTED ON (B)(6) 2020, REVEALED ELEVATED SERUM CHROMIUM AND BLOOD COBALT LEVELS IN PATIENT. ON (B)(6) 2021, PATIENT HAD A REVISION SURGERY PERFORMED ON HIS RIGHT HIP BY (B)(6), M.D., AT (B)(6). EARLY IN HIS (B)(6) 2021, OPERATIVE REPORT, DR. (B)(6) NOTED, "METALLOSIS DEBRIS AT THE EXTERNAL ROTATORS." AFTER REMOVAL OF THE CONSERVE® FEMORAL HEAD FROM THE PROFEMUR® PLUS MODULAR NECK, DR. (B)(6) NOTED "A FAIR AMOUNT OF BLACK CORROSION DISCOLORATION AT THE TAPER ON BOTH COMPONENTS." IN (B)(6) 2021, REVISION SURGERY, DR. (B)(6), REMOVED THE WRIGHT MEDICAL PROFEMUR® HIP SYSTEM, INCLUDING THE PROFEMUR® PLUS COCR MODULAR NECK AND PROFEMUR® PLASMA Z STE
Patient problems: Metal Related Pathology; Neuropathy; Pain
MAUDE report 14378221 ↗
FDA clearance and classification
The FDA regulates product code JDL (prosthesis, hip, semi-constrained (metal cemented acetabular component)) under 21 CFR 888.3320 as a Class III device. 21 CFR 888.3320 ↗
Devices of this type need premarket approval (PMA), the FDA's most demanding review, usually with clinical data.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the CONSERVE® 38CH5410
Has the CONSERVE® 38CH5410 been recalled?
No FDA recall record in our data names this exact device identifier (DI M68438CH54101), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the CONSERVE® 38CH5410?
The FDA's MAUDE database lists 1 adverse event report for this device, received on May 12, 2022. That report lists device problem “Corroded” (1); patient problems “Metal Related Pathology” (1), “Neuropathy” (1) and “Pain” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the CONSERVE® 38CH5410?
None of the reports are classified as deaths, 1 as an injury and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the CONSERVE® 38CH5410 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the CONSERVE® 38CH5410, and is it still sold?
The FDA's device database lists Microport Orthopedics Inc. as the company responsible for it (brand CONSERVE®, model 38CH5410). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular Component) devices
- Omnifit 2001-2246Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained (metal cemented acetabular component)1 recall0 reports
- Taperloc Complete Micro Primary Femoral 51-107150Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained (metal cemented acetabular component)40 reports
- Taperloc Complete Micro Primary Femoral 51-107140Biomet Orthopedics, LLC · Prosthesis, hip, semi-constrained (metal cemented acetabular component)36 reports
- Dynasty DSBFGD52Microport Orthopedics Inc. · Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular Component)12 reports
- CONSERVE® 38AM4600Microport Orthopedics Inc. · Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular Component)8 reports
- CONSERVE® 38AM4804Microport Orthopedics Inc. · Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular Component)7 reports
- CONSERVE® 38AM4404Microport Orthopedics Inc. · Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular Component)6 reports
- TRANSCEND® 26000025Microport Orthopedics Inc. · Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular Component)5 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.