Skip to content

CONSERVE® 38CH5410Recalls & reported problems

Microport Orthopedics Inc. · CONSERVE® TOTAL 54MM BCH® FEMORAL HEAD · Class 3 · Implant

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI M68438CH54101), received between May 12, 2022 and May 12, 2022 — classified as 1 injury.

The most frequently reported specific device problem is “Corroded” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Corroded1

Patient problems

  • Metal Related Pathology1
  • Neuropathy1
  • Pain1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryMay 12, 2022Corroded

    ALLEGEDLY PT. NOW HAS ELEVATED METAL IONS. ORIGINALLY PART OF IDE / CERAMIC ON METAL ARTICULATION SYSTEM CLINICAL STUDY BUT NOT ANY LONGER. ADDITIONAL INFORMATION RECEIVED ON APRIL 19, 2022: PATIENT DEVELOPED ADVERSE SYMPTOMS INCLUDING, BUT NOT LIMITED TO, PROGRESSIVELY WORSENING PAIN IN AND AROUND HIS RIGHT HIP JOINT, AS WELL AS NEUROLOGIC SYMPTOMS INCLUDING PERIPHERAL NEUROPATHY, TINNITUS, SOME BALANCE/COORDINATION ISSUES, A HEAL FRACTURE AS A DIRECT RESULT OF THOSE BALANCE/COORDINATION ISSUES, AND B 12 DEFICIENCY. LABORATORY TEST RESULTS REPORTED ON (B)(6) 2020, REVEALED ELEVATED SERUM CHROMIUM AND BLOOD COBALT LEVELS IN PATIENT. ON (B)(6) 2021, PATIENT HAD A REVISION SURGERY PERFORMED ON HIS RIGHT HIP BY (B)(6), M.D., AT (B)(6). EARLY IN HIS (B)(6) 2021, OPERATIVE REPORT, DR. (B)(6) NOTED, "METALLOSIS DEBRIS AT THE EXTERNAL ROTATORS." AFTER REMOVAL OF THE CONSERVE® FEMORAL HEAD FROM THE PROFEMUR® PLUS MODULAR NECK, DR. (B)(6) NOTED "A FAIR AMOUNT OF BLACK CORROSION DISCOLORATION AT THE TAPER ON BOTH COMPONENTS." IN (B)(6) 2021, REVISION SURGERY, DR. (B)(6), REMOVED THE WRIGHT MEDICAL PROFEMUR® HIP SYSTEM, INCLUDING THE PROFEMUR® PLUS COCR MODULAR NECK AND PROFEMUR® PLASMA Z STE

    Patient problems: Metal Related Pathology; Neuropathy; Pain

    MAUDE report 14378221 ↗

Other Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular Component) devices

All Prosthesis, Hip, Semi-Constrained (Metal Cemented Acetabular Component) devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.