PROFEMUR® PRTL0025Recalls & reported problems
Microport Orthopedics Inc. · PROFEMUR® TL STEM SZ 5 · Class 2 · Implant
All models of PROFEMUR (116)
Data updated Nov 27, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI M684PRTL00251), received between Mar 10, 2020 and Mar 7, 2025 — classified as 6 injuries.
The most frequently reported specific device problem is “Corroded” (1 report).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Appropriate Device Problem Term/Code Not Available1
- Corroded1
- Device Dislodged or Dislocated1
- Insufficient Device Problem Information1
- Loss of Osseointegration1
- Material Discolored1
- Mechanical Problem1
- Use of Device Problem1
Patient problems
- Pain5
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Loss of Range of Motion1
- Metal Related Pathology1
- Unspecified Infection1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryMar 7, 2025Corroded; Corroded
ALLEGEDLY, DOCTOR RECOMMENDED THE REVISION SURGERY AFTER PLAINTIFF PRESENTED WITH PROGRESSIVE RIGHT HIP PAIN, ELEVATED ION LEVELS, METALLOSIS CHANGES AROUND THE FEMORAL NECK AND SURROUNDING SOFT TISSUES AROUND THE PROXIMAL FEMUR, ADVERSE LOCAL TISSUE REACTION AND LACK OF MOBILITY. THE REVISION WAS NECESSARY BECAUSE THE DEVICE FAILED DUE TO CORROSION AT THE NECK-STEM JUNCTION OF THE DEVICE. PRODUCTS NOT SPECIFIED THEY WERE REVISED: PRODUCT ID: DBFPGD52 DYNASTY® BF SHELL PRIMARY 52MM GROUP D, LOT: 1684715, QTY:1.
Patient problems: Loss of Range of Motion; Pain
MAUDE report 21548583 ↗InjuryMay 12, 2022Material Discolored
ALLEGEDLY, THE PATIENT HAS BEEN SUFFERING MODERATE PAIN, AND THE IMPLANTS WERE REVISED FOR METALLOSIS, ACCORDING TO THE SURGEON. THE MODULAR NECK TRUNION (FITTING IN TO THE STEM POCKET) WAS NOTICEABLY "BLACKENED". ALL IMPLANTS WERE REMOVED, AND REPLACED WITH A THR CEMENT SPACER. UNFORTUNATELY IT WASN'T POSSIBLE TO REMOVE THE LINER FROM THE PROCOTYL L SHELL, SO I CANT PROVIDE THE LOT NUMBER FOR THE BIOLOX LINER.
Patient problems: Metal Related Pathology; Pain; Unspecified Infection
MAUDE report 14380060 ↗InjuryFeb 15, 2022Mechanical Problem; Loss of Osseointegration; Mechanical Problem; Loss of Osseointegration
ALLEGEDLY, THE PATIENT WAS COMPLAINING OF PAIN AND CRP LEVELS WERE HIGH, SO WE SUSPECTED INFECTION AND PLANNED TO REMOVE THE TL, BUT AS IT TURNED OUT, THERE WAS NO INFECTION IN THE TEST RESULTS. DURING THE OP, THE NECK COULD NOT BE REMOVED FROM THE STEM NO MATTER WHAT I DID, AND AS A RESULT, THE ENTIRE STEM WAS REMOVED. THE AFFECTED AREA WAS CLEAN AND THERE WAS NO LOOSENING ON THE CUP SIDE. THERE WAS NO EVIDENCE OF BONE ON GROWTH AT THE PROXIMAL PLASMAPHERESIS COATING. (B)(4).
Patient problems: Pain
MAUDE report 13534768 ↗InjuryNov 8, 2021Appropriate Device Problem Term/Code Not Available
SAE# (B)(4). ALLEGEDLY, THE PATIENT STARTED HAVING LT HIP PAIN 3 WEEKS AFTER SURGERY. MULTIPLE ATTEMPTS TO RELIEF THE PAIN WITH CORTICOSTEROID INJECTIONS INTRA-ARTICULAR WITH NO SUCCESS. THE PATIENT UNDERWENT A SURGICAL INTERVENTION FOR RELEASE OF THE PSOAS TENDON ON (B)(6) 2018. THE PAIN AND WEAKNESS IN THE LEFT THIGH IS STILL PERSISTENT. ADDITIONAL INFORMATION FROM CLINICAL RECEIVED ON (B)(6) 2021 CONFIRMING THAT COMPONENTS WERE NOT REVISED.
Patient problems: Pain
MAUDE report 12771850 ↗InjuryOct 12, 2021Use of Device Problem; Device Dislodged or Dislocated; Use of Device Problem; Device Dislodged or Dislocated
ALLEGEDLY, THE PATIENT DISLOCATED A FEW DAYS AFTER THE ORIGINAL SURGERY, THIS WAS REDUCED AND THE PATIENT WAS OK FOR ABOUT 10 DAYS. BEFORE THERE WAS A FURTHER DISLOCATION. THE SURGEON FELT THAT THE REASON FOR THE DISLOCATION WAS A LACK OF TENSION, OR OFFSET. THIS WAS RESOLVED WITH A LONGER NECK, CREATING MORE OFFSET. THE OPERATION WAS COMPLETED SUCCESSFULLY, THE PATIENT HAS SINCE BEEN DISCHARGED FROM HOSPITAL. THE SURGEON PUTS THIS DOWN TO SURGEON ERROR.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available
MAUDE report 12615224 ↗InjuryMar 10, 2020Insufficient Device Problem Information; Insufficient Device Problem Information
ALLEGEDLY, SUBJECT WAS SEEN IN EMERGENCY ROOM WITH RIGHT HIP PAIN FOR 2 WEEKS, CRUNCHING WITH MOTION AND DIFFICULTY AMBULATING. LINER FRACTURE WAS CONFIRMED BY X RAY AND CT OF THE HIP. SUBJECT HAD A REVISION AND REMOVAL OF THE ORIGINAL ACETABULAR, LINER, MODULAR NECK AND FEMORAL HEAD AND WAS DISCHARGED ON (B)(6) 2015.
Patient problems: Pain
MAUDE report 9813778 ↗
FDA clearance and classification
The FDA regulates product code LWJ (prosthesis, hip, semi-constrained, metal/polymer, uncemented) under 21 CFR 888.3360 as a Class II device. 21 CFR 888.3360 ↗
Devices of this type normally need 510(k) clearance before sale.
Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.
Common questions about the PROFEMUR® PRTL0025
Has the PROFEMUR® PRTL0025 been recalled?
No FDA recall record in our data names this exact device identifier (DI M684PRTL00251), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the PROFEMUR® PRTL0025?
The FDA's MAUDE database lists 6 adverse event reports for this device, received between Mar 10, 2020 and Mar 7, 2025. In those reports, the most common specific device problems are “Corroded” (1), “Device Dislodged or Dislocated” (1) and “Loss of Osseointegration” (1); the most common patient problems are “Pain” (5), “Loss of Range of Motion” (1) and “Metal Related Pathology” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the PROFEMUR® PRTL0025?
None of the reports are classified as deaths, 6 as injuries and 0 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Is the PROFEMUR® PRTL0025 MRI safe?
According to the device's FDA GUDID record, its labeling lists it as MR Conditional: it can be scanned only under the specific conditions given in the labeling. The radiology team should check those conditions before an MRI.
Who makes the PROFEMUR® PRTL0025, and is it still sold?
The FDA's device database lists Microport Orthopedics Inc. as the company responsible for it (brand PROFEMUR®, model PRTL0025). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Uncemented devices
- LFIT 6260-9-136Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented166 reports
- Trident 623-00-36EHowmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented112 reports
- Accolade 6021-0435Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented88 reports
- EXETER 0580-1-044Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented1 recall23 reports
- Endo-Model SL Knee Prosthesis System 16-2840/02WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Uncemented1 recall16 reports
- Endo-Model SL Knee Prosthesis System 16-2840/05WALDEMAR LINK GmbH & Co. KG · Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Uncemented1 recall13 reports
- Trident 623-10-36EHowmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented57 reports
- LFIT 6260-9-336Howmedica Osteonics Corp. · Prosthesis, hip, semi-constrained, metal/polymer, uncemented56 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page