Axle Interspinous Fusion System C060-0010Recalls & reported problems
Xtant Medical Holdings, Inc. · SPECIAL PLIERS UNIVERSAL AIO ASSEMBLY AXLE · Class 1
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI M697C06000101), received between Jul 15, 2024 and Jul 15, 2024 — classified as 2 malfunctions.
The most frequently reported specific device problem is “Fracture” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture2
Patient problems
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
MalfunctionJul 15, 2024Fracture; Fracture
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL POSTERIOR LUMBER INTERSPINOUS PROCESS FUSION PROCEDURE AT AN UNKNOWN LEVEL. DURING THE PROCEDURE, WHILE TIGHTENING THE RETENTION SCREW OF THE ALL-IN-ONE (AIO) INSERTER TO ENGAGE THE SYSTEM IMPLANT, THE DISTAL IMPLANT GRASPING PORTION ON THE LOCKING PLATE SIDE OF THE INSTRUMENT FRACTURED. A SECOND INSERTER WAS USED TO PROCEED TO FORWARD WITH THE PROCEDURE. SUBSEQUENTLY, DURING COMPRESSION OF THE IMPLANT ONTO THE SPINOUS PROCESSES, THE DISTAL FUNCTIONAL FEATURE OF THE CROSS BAR PLATE SIDE OF THE SECOND INSTRUMENT FRACTURED OFF. A THIRD INSTRUMENT WAS USED TO COMPLETE THE PROCEDURE WITH NO FURTHER ISSUES. IT WAS ALSO REPORTED THAT THE DISTAL TIP OF THE SCREWDRIVER, USED TO TIGHTEN THE RETENTION SCREW OF THE INSERTER FRACTURED OFF DURING THE CASE; HOWEVER, IT IS UNKNOWN IF THIS OCCURRED DURING THE USE OF THE FIRST OR SECOND INSERTER DEVICE. THERE WERE NO KNOWN PATIENT COMPLICATIONS OR DELAY IN TREATMENT ASSOCIATED WITH THIS COMPLAINT. NO ADDITIONAL INFORMATION IS AVAILABLE. REPORT 2 OF 3.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19744265 ↗MalfunctionJul 15, 2024Fracture; Fracture
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL POSTERIOR LUMBER INTERSPINOUS PROCESS FUSION PROCEDURE AT AN UNKNOWN LEVEL. DURING THE PROCEDURE, WHILE TIGHTENING THE RETENTION SCREW OF THE ALL-IN-ONE (AIO) INSERTER TO ENGAGE THE SYSTEM IMPLANT, THE DISTAL IMPLANT GRASPING PORTION ON THE LOCKING PLATE SIDE OF THE INSTRUMENT FRACTURED. A SECOND INSERTER WAS USED TO PROCEED TO FORWARD WITH THE PROCEDURE. SUBSEQUENTLY, DURING COMPRESSION OF THE IMPLANT ONTO THE SPINOUS PROCESSES, THE DISTAL FUNCTIONAL FEATURE OF THE CROSS BAR PLATE SIDE OF THE SECOND INSTRUMENT FRACTURED OFF. A THIRD INSTRUMENT WAS USED TO COMPLETE THE PROCEDURE WITH NO FURTHER ISSUES. IT WAS ALSO REPORTED THAT THE DISTAL TIP OF THE SCREWDRIVER, USED TO TIGHTEN THE RETENTION SCREW OF THE INSERTER FRACTURED OFF DURING THE CASE; HOWEVER, IT IS UNKNOWN IF THIS OCCURRED DURING THE USE OF THE FIRST OR SECOND INSERTER DEVICE. THERE WERE NO KNOWN PATIENT COMPLICATIONS OR DELAY IN TREATMENT ASSOCIATED WITH THIS COMPLAINT. NO ADDITIONAL INFORMATION IS AVAILABLE. REPORT 1 OF 3.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19744250 ↗
Other Pliers, Surgical devices
- NA 391.962Synthes GmbH · PLIERS, SURGICAL36 reports
- NA 03.130.270Synthes GmbH · PLIERS, SURGICAL34 reports
- NA 388.50Synthes GmbH · PLIERS, SURGICAL23 reports
- Axle Interspinous Fusion System X060-1520Xtant Medical Holdings, Inc. · Pliers, Surgical19 reports
- NA 388.509Synthes GmbH · PLIERS, SURGICAL14 reports
- VA-LCP 03.211.005Synthes GmbH · PLIERS, SURGICAL6 reports
- NA 391.82Synthes GmbH · PLIERS, SURGICAL5 reports
- NA 391.88Synthes GmbH · PLIERS, SURGICAL4 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.