Irix-C Cervical Integrated Fusion System T066-0045Recalls & reported problems
Xtant Medical Holdings, Inc. · Irix-C Forceps Implant Inserter · Class 1 · Implant
All models of Irix-C Cervical Integrated Fusion System (11)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI M697T06600451), received between Sep 19, 2019 and Apr 4, 2023 — classified as 4 malfunctions.
The most frequently reported specific device problem is “Fracture” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Fracture2
- Break1
- Material Fragmentation1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Known Impact Or Consequence To Patient2
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 4, 2023Break; Break
THE MANUFACTURER WAS MADE AWARE OF A PRODUCT COMPLAINT ON (B)(6) 2023. IT WAS REPORTED THAT A FORCEPS IMPLANT INSERTER WAS BROKEN WHEN IMPACTED WITH A MALLET FOR INSERTION OF A SYSTEM IMPLANT. THERE WERE NO KNOWN PATIENT COMPLICATIONS OR DELAY IN TREATMENT. AN ALTERNATE AVAILABLE INSTRUMENT WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. AN RA# WAS ISSUED FOR RETURN OF THE COMPLAINT INSTRUMENT, WHICH WAS RECEIVED AT THE MANUFACTURER FOR ASSESSMENT ON 3/13/2023.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16677650 ↗MalfunctionNov 17, 2022Material Fragmentation; Material Fragmentation
THE MANUFACTURER RECEIVED NOTIFICATION ON 10/24/2022 OF A SYSTEM IMPLANT INSERTER THAT MALFUNCTIONED DURING A SURGICAL PROCEDURE. IT WAS REPORTED THAT A DISTAL ARM OF THE IMPLANT INSERTER FRACTURED WHILE PLACING A SYSTEM IMPLANT. THERE WERE NO KNOWN PATIENT COMPLICATIONS ASSOCIATED WITH THIS COMPLAINT. THE SURGEON USED HOSPITAL INSTRUMENTATION TO SUCCESSFULLY COMPLETE THE SURGICAL PROCEDURE. A RETURN AUTHORIZATION WAS ISSUED FOR RETURN OF THE COMPLAINT INSTRUMENT, WHICH WAS RECEIVED AT THE MANUFACTURER ON 10/28/2022 FOR COMPLAINT ASSESSMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15819719 ↗MalfunctionApr 3, 2020Fracture; Fracture
ON (B)(6) 2020 THE COMPANY RECEIVED A COMMUNICATION FROM A DISTRIBUTOR THAT A CRACK WAS IDENTIFIED ON AN INSTRUMENT. IT WAS REPORTED THAT DURING A SURGICAL PROCEDURE ON (B)(6) 2020 THE COMPLAINANT NOTICED THAT THE INSTRUMENT WAS NOT HOLDING THE IMPLANT WELL AND IDENTIFIED THE CRACK. IT IS UNKNOWN HOW THE CRACK PROPAGATED ON THE INSTRUMENT. THERE WERE NO KNOWN PATIENT COMPLICATIONS ASSOCIATED WITH THIS COMPLAINT. THE PROCEDURE WAS SUCCESSFULLY COMPLETED USING A GENERAL HOSPITAL INSTRUMENT.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9924463 ↗MalfunctionSep 19, 2019Fracture; Fracture
THE COMPLAINANT REPORTED THAT DURING AN ACDF REVISION DUE TO NONUNION, A FORCEPS IMPLANT INSERTER AND A GUIDED ANGLED AWL WERE BROKEN. THE INITIAL SURGERY WAS REPORTEDLY A NON-X-SPINE PRODUCT. THERE WERE NO REPORTED PATIENT COMPLICATIONS ASSOCIATED WITH THE INSTRUMENT MALFUNCTIONS. THE FOLLOWING MDRS CORRESPOND TO THIS EVENT: 3005031160-2019-00043.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9088042 ↗
Other IMPACTOR devices
- PINNACLE 2217-50-041Depuy (Ireland) · IMPACTOR4,295 reports
- ATTUNE 2544-01-006Depuy (Ireland) · IMPACTOR1 recall2,539 reports
- ATTUNE 2544-01-003Depuy (Ireland) · IMPACTOR1 recall1,498 reports
- ATTUNE 2544-01-017Depuy (Ireland) · IMPACTOR1 recall1,456 reports
- ATTUNE 2544-01-004Depuy (Ireland) · IMPACTOR1 recall875 reports
- DURALOC 2360-71-000Depuy Orthopaedics, Inc. · IMPACTOR716 reports
- SPECIALIST 2581-11-000Depuy (Ireland) · IMPACTOR704 reports
- INHANCE 6205-10-103Depuy (Ireland) · IMPACTOR337 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page