Skip to content

Irix-C Cervical Integrated Fusion System T066-0059Recalls & reported problems

Xtant Medical Holdings, Inc. · Forceps Inserter · Class 1 · Implant

All models of Irix-C Cervical Integrated Fusion System (11)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI M697T06600591), received between May 26, 2020 and May 26, 2020 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Solder Joint Fracture” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Solder Joint Fracture2

Patient problems

  • No Known Impact Or Consequence To Patient2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 26, 2020Solder Joint Fracture; Solder Joint Fracture

    THIS MEDICAL DEVICE REPORT IS ASSOCIATED WITH MDR NUMBER 3005031160-2020-00013, WHERE IT WAS REPORTED THAT A FORCEPS IMPLANT INSERTER MALFUNCTIONED WHILE IN USE. THE COMPLAINANT ASSESSED A SECOND FORCEPS INSERTER (FROM THE SAME LOT) AND FOUND THAT THE SAME FAILURE WAS ACHIEVED WITH MINIMAL FORCE APPLIED BY HAND. THE SECOND MALFUNCTION WAS NOT REPORTEDLY INVOLVED IN A SURGICAL PROCEDURE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10089906 ↗
  • MalfunctionMay 26, 2020Solder Joint Fracture; Solder Joint Fracture

    THE COMPANY RECEIVED NOTIFICATION ON 5/05/2020 REGARDING A FORCEPS IMPLANT INSERTER THAT WAS REPORTED TO HAVE THE IMPACTION PLATE FRACTURE FROM THE INSTRUMENT WITH MINIMAL STRIKING FORCE DURING A SURGICAL PROCEDURE ON (B)(6) 2020. THE PROCEDURE WAS SUCCESSFULLY COMPLETED, THERE WERE NO KNOWN PATIENT COMPLICATIONS. THE COMPLAINANT ASSESSED A SECOND FORCEPS INSERTER (FROM THE SAME LOT) AND FOUND THAT THE SAME FAILURE WAS ACHIEVED WITH MINIMAL FORCE APPLIED BY HAND. REPORT 3005031160-2020-00014 HAS BEEN SUBMITTED FOR THE SECOND DEVICE.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10089904 ↗

Other Impactor devices

All Impactor devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page