Calix Spinal Implant System X034-0015Recalls & reported problems

Xtant Medical Holdings, Inc. · ASSEMBLY, PLIF/TLIF INSERTER · Class 1 · Implant

Data updated Nov 27, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI M697X03400151), received between Oct 5, 2018 and Jul 14, 2026 — classified as 8 malfunctions.

The most frequently reported specific device problem is “Break” (4 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.

Reports received per month

Dec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 2 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reportsNov 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break4
  • Solder Joint Fracture2
  • Fracture1
  • Material Fragmentation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions7
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 14, 2026Break

    THE MANUFACTURER WAS MADE AWARE OF A PRODUCT COMPLAINT ON 06/16/2026. IT WAS REPORTED THAT THE DISTAL THREADED TIP OF THE INNER SHAFT OF A SYSTEM INSERTER WAS FRACTURED WHILE TAPPING IN THE IMPLANT DURING A TWO LEVEL TRANSFORAMINAL LUMBAR INTERBODY FUSION PROCEDURE PERFORMED ON (B)(6)2026. THE COMPLAINANT CONFIRMED THERE WERE NO KNOWN PATIENT COMPLICATIONS OR DELAYS IN PROCEDURE. THE SURGEON TWISTED THE BROKEN TIP OUT OF THE PEEK WITH A NEEDLE DRIVER, AND THE PROCEDURE WAS SUCCESSFULLY COMPLETED. A RETURN AUTHORIZATION NUMBER WAS ISSUED FOR THE RETURN OF THE COMPLAINT INSTRUMENTS, WHICH WERE RECEIVED AT THE MANUFACTURER FOR ASSESSMENT ON 06/22/2026.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25816951 ↗
  • MalfunctionMar 26, 2026Break

    THE MANUFACTURER WAS CONTACTED ON 03/02/2026 TO REPORT AN ISSUE WITH TWO SYSTEM INSERTERS. IT WAS REPORTED THAT THE DISTAL THREADED TIPS OF THE INNER SHAFTS OF THE INSERTERS WERE FRACTURED FROM THE INSTRUMENTS WHILE THEY WERE BEING USED TO PLACE IMPLANTS DURING A L3-S1 TLIF. THIS INCIDENT OCCURRED WITH TWO DIFFERENT INSERTERS ON TWO SEPARATE IMPLANTS. THE FRACTURED TIPS OF THE INSERTERS WERE NOT RECOVERABLE AS THEY REMAINED EMBEDDED IN THE IMPLANTS, WHICH WERE SUCCESSFULLY IMPLANTED WITH THE INSERTERS DESPITE THE REPORTED ISSUES. A RETURN AUTHORIZATION NUMBER WAS ISSUED FOR THE RETURN OF THE COMPLAINT INSTRUMENTS, WHICH WERE RECEIVED AT THE MANUFACTURER FOR ASSESSMENT ON 03/12/2026. REPORT 2 OF 2

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24698820 ↗
  • MalfunctionMar 26, 2026Break

    THE MANUFACTURER WAS CONTACTED ON 03/02/2026 TO REPORT AN ISSUE WITH TWO SYSTEM INSERTERS. IT WAS REPORTED THAT THE DISTAL THREADED TIPS OF THE INNER SHAFTS OF THE INSERTERS WERE FRACTURED FROM THE INSTRUMENTS WHILE THEY WERE BEING USED TO PLACE IMPLANTS DURING A L3-S1 TLIF. THIS INCIDENT OCCURRED WITH TWO DIFFERENT INSERTERS ON TWO SEPARATE IMPLANTS. THE FRACTURED TIPS OF THE INSERTERS WERE NOT RECOVERABLE AS THEY REMAINED EMBEDDED IN THE IMPLANTS, WHICH WERE SUCCESSFULLY IMPLANTED WITH THE INSERTERS DESPITE THE REPORTED ISSUES. A RETURN AUTHORIZATION NUMBER WAS ISSUED FOR THE RETURN OF THE COMPLAINT INSTRUMENTS, WHICH WERE RECEIVED AT THE MANUFACTURER FOR ASSESSMENT ON 03/12/2026. REPORT 1 OF 2.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24698161 ↗
  • MalfunctionFeb 3, 2026Break

    THE MANUFACTURER WAS MADE AWARE OF A SYSTEM COMPLAINT ON 01/09/2026 IN WHICH IT WAS REPORTED THAT THE DISTAL TIP OF AN INSERTER SHAFT BROKE FROM THE DEVICE AS IT WAS BEING USED TO PREP AN IMPLANT FOR A LUMBAR FUSION. THE COMPLAINANT STATED THAT, "THE SURGICAL TECH WENT TO SCREW THE INSERTER INTO THE IMPLANT ON THE BACK TABLE AND THE TIP BROKE OFF IN THE IMPLANT, THE TIP WAS UNSCREWED OUT OF THE IMPLANT AND A NEW INSERTER INSTALLED AND IMPLANT WAS LOADED ONTO NEW INSERTER AND IMPLANTED WITH NO ISSUE. WE REMOVED THE OLD INSERTER SHAFT AND TIP THAT BROKE OFF AND ITS BEING SENT BACK". THERE WERE NO REPORTED PATIENT COMPLICATIONS OR DELAYS IN SURGERY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24251804 ↗
  • MalfunctionAug 1, 2022Solder Joint Fracture; Solder Joint Fracture

    THE MANUFACTURER RECEIVED NOTIFICATION ON 7/06/2022 OF AN IMPLANT INSERTER THAT MALFUNCTIONED DURING A PROCEDURE. IT WAS REPORTED THAT WHILE IMPACTING THE INSERTER WITH A MALLET, A HANDLE T-PIN FELL OFF THE IMPLANT INSERTER. THERE WERE NO KNOWN PATIENT COMPLICATIONS ASSOCIATED WITH THIS COMPLAINT. THE SURGICAL PROCEDURE WAS SUCCESSFULLY COMPLETED WITH AN ALTERNATE AVAILABLE INSTRUMENT. THE COMPLAINT INSTRUMENT WAS RECEIVED ON 7/07/2022 AT THE MANUFACTURER FOR COMPLAINT ASSESSMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15138474 ↗
  • MalfunctionSep 27, 2021Solder Joint Fracture; Solder Joint Fracture

    THE COMPANY RECEIVED NOTIFICATION ON 8/31/2021 OF AN IMPLANT INSERTER THAT MALFUNCTIONED DURING A PROCEDURE ON AN UNKNOWN DATE. IT WAS REPORTED THAT WHILE IMPACTING THE INSERTER WITH A MALLET, THE PHYSICIAN INADVERTENTLY IMPACTED THE INSERTER T-HANDLE. THE IMPACT TO THE INSERTER T-HANDLE RESULTED IN AN INSTRUMENT MALFUNCTION. THERE WERE NO KNOWN PATIENT COMPLICATIONS ASSOCIATED WITH THIS COMPLAINT. THE SURGICAL PROCEDURE WAS SUCCESSFULLY COMPLETED WITH THE COMPLAINT INSTRUMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12537470 ↗
  • MalfunctionNov 12, 2020Material Fragmentation; Material Fragmentation

    THE COMPANY RECEIVED NOTIFICATION ON (B)(6) 2020 REGARDING AN INSTRUMENT MALFUNCTION THAT OCCURRED DURING A SURGICAL PROCEDURE. IT WAS REPORTED THAT THE THREADED DISTAL TIP OF A SYSTEM INSERTER FRACTURED INSIDE OF AN IMPLANT WHEN BEING PLACED. THE IMPLANT WITH THE FRACTURED DISTAL TIP OF THE INSERTER WAS REMOVED AND REPLACED WITH ANOTHER IMPLANT. THERE WAS AN ALTERNATE INSERTER AVAILABLE, WHICH WAS USED TO SUCCESSFULLY COMPLETE THE PROCEDURE. THERE WERE NO KNOWN PATIENT COMPLICATIONS ASSOCIATED WITH THIS INSTRUMENT MALFUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 10829704 ↗
  • MalfunctionOct 5, 2018Fracture; Fracture

    THE COMPLAINANT REPORTED THAT TWO INSERTERS FAILED IN THE FIELD. THE MALFUNCTION WAS DISCOVERED PRIOR TO IMPLANTATION DURING A SURGERY. AN ALTERNATE INSTRUMENT WAS AVAILABLE TO COMPLETE THE PROCEDURE. THERE WERE NO KNOWN PATIENT COMPLICATIONS.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 7940494 ↗

FDA clearance and classification

The FDA regulates product code HWA (impactor) under 21 CFR 888.4540 as a Class I device. 21 CFR 888.4540 ↗

Devices of this type are generally exempt from premarket notification (510(k)), so a model can be sold without an FDA clearance of its own; it still has to meet FDA registration, labeling and reporting rules.

Manufacturers may send malfunction reports for this device type to the FDA in quarterly summaries, where one report can stand for several events.

Common questions about the Calix Spinal Implant System X034-0015

Has the Calix Spinal Implant System X034-0015 been recalled?

No FDA recall record in our data names this exact device identifier (DI M697X03400151), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.

What problems have been reported with the Calix Spinal Implant System X034-0015?

The FDA's MAUDE database lists 8 adverse event reports for this device, received between Oct 5, 2018 and Jul 14, 2026. In those reports, the most common specific device problems are “Break” (4), “Solder Joint Fracture” (2) and “Fracture” (1). Reports are not verified by the FDA and do not prove the device caused the problem.

Have deaths or injuries been reported with the Calix Spinal Implant System X034-0015?

None of the reports are classified as deaths, 0 as injuries and 8 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.

Is the Calix Spinal Implant System X034-0015 FDA approved?

Devices of this type (product code HWA) are generally exempt from FDA premarket notification, so this model has no FDA clearance number of its own. It is listed in the FDA's device database (GUDID) by Xtant Medical Holdings, Inc.

Is the Calix Spinal Implant System X034-0015 MRI safe?

According to the device's FDA GUDID record, its labeling does not contain MRI safety information. Ask the manufacturer or the radiology team before an MRI.

Who makes the Calix Spinal Implant System X034-0015, and is it still sold?

The FDA's device database lists Xtant Medical Holdings, Inc. as the company responsible for it (brand Calix Spinal Implant System, model X034-0015). Its record shows it as in commercial distribution.

Answers are built automatically from the FDA records on this page. They are not medical advice.

Other IMPACTOR devices

All IMPACTOR devices →