Skip to content

TiLock 1 (REF G711-65-45)Recalls & reported problems

Genesys Orthopedics Systems LLC · Break-Off Pedicle Screw · Class 2

Adverse event reports
1
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI M719G71165450), received between Aug 8, 2024 and Aug 8, 2024 — classified as 1 injury.

The most frequently reported specific device problem is “Use of Device Problem” (1 report).

1 other recall cover devices cleared under the same FDA submission (K100757); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Use of Device Problem1

Patient problems

  • Failure of Implant1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 8, 2024Use of Device Problem; Use of Device Problem

    A DISTRIBUTOR INFORMED GENESYS SPINE THAT DR. (B)(6) DISCOVERED ON X-RAY A BROKEN 6.5MM DIAMETER TILOCK2 PEDICLE SCREW (ORIGINALLY IMPLANTED IN 2020 OR 2021, L4 TO S1 FUSION) WHEN DOING A PLANNED IMPLANT OF ADDITIONAL ADJACENT SCREWS. REVISION OF THE BROKEN SCREW HEAD AND SCREW TOOK PLACE DURING THE SAME DAY. THE SCREW BODY WAS MISPLACED BUT THE SCREW HEAD WAS RETURNED THE WEEK OF (B)(6) 2024. NO PATIENT SYMPTOMS OR COMPLAINTS WERE NOTED THAT INDICATED A BROKEN SCREW PRIOR TO THE X-RAY IDENTIFICATION. DISTRIBUTOR INDICATED THAT THE BROKEN SCREW WAS ONLY REMOVED BECAUSE OF INTERFERENCE WITH THE PLANNED IMPLANTS. BONY FUSION HAD BEEN ACHIEVED AT L4-S1 PRIOR TO THE SCREW FRACTURE. ACCORDING TO GENESYS SPINE VP OF FIELD SERVICES, THE SCREW CHOSEN AT IMPLANT MAY BE UNDERSIZED FOR THE INTENDED LOCATION AND APPLICATION, AND IDEALLY WOULD HAVE BEEN EITHER A 7.5MM OR 8.5MM SCREW.

    Patient problems: Failure of Implant

    MAUDE report 19939325 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K100757. One submission often covers many models, so check the product and lot details in each record.

  • Z-2677-2014 · Apr 2, 2014 · Terminated
    Circular Lock Screws, part # G826, lot # 11406. The Lock Screws are part of the Genesys Spine TiLock Pedicle Screw System. The TiLock Pedicle Screw System is intended to provide i