TiLock 1 (REF GLC124-S)Recalls & reported problems
Genesys Orthopedics Systems LLC · Blunt Nitinol K-Wire - Short · Class 2
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI M719GLC124S0), received between Oct 30, 2019 and Oct 30, 2019 — classified as 1 malfunction.
The most frequently reported specific device problem is “Break” (1 report).
1 other recall cover devices cleared under the same FDA submission (K100757); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Break1
Patient problems
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 30, 2019Break; Break
THE COMPLAINANT REACHED OUT TO GENESYS SPINE TO REPORT THAT TWO (2) NITINOL GUIDEWIRES HAD BROKEN DURING AN MIS PROCEDURE. THE COMPLAINANT REPORTED THAT THE PATIENT HAD "VERY, VERY HARD AND DENSE BONE" BUT THE SURGEON DID NOT NOTE ANY OTHER UNUSUAL OCCURENCE UNTIL HE TRIED TO REMOVE THE WIRES AFTER SCREW PLACEMENT. AT THIS POINT THE WIRES FELT "STUCK" AND WERE DIFFICULT TO REMOVE. THE SURGEON HAD TO ATTACH INSTRUMENTATION TO THE WIRES TO PULL HARD ENOUGH TO REMOVE THEM FROM THE SCREWS HOWEVER, IN DOING SO, BROKE OFF THE TIPS OF THE WIRES.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9256708 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K100757. One submission often covers many models, so check the product and lot details in each record.
- Z-2677-2014 · Apr 2, 2014 · TerminatedCircular Lock Screws, part # G826, lot # 11406. The Lock Screws are part of the Genesys Spine TiLock Pedicle Screw System. The TiLock Pedicle Screw System is intended to provide i
Other Orthosis, Spinal Pedicle Fixation, For Degenerative Disc Disease devices
- EXPEDIUM VERSE 199721000Medos International Sàrl · Orthosis, spinal pedicle fixation, for degenerative disc disease387 reports
- Sequoia® 3301-1Zimmer Spine, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease184 reports
- CD HORIZON® Spinal System 8590855Medtronic Sofamor Danek, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease179 reports
- Vitality® 07.02011.001Biomet Spine LLC · Orthosis, spinal pedicle fixation, for degenerative disc disease3 recalls0 reports
- EXPEDIUM VERSE 199721001SMedos International Sàrl · Thoracolumbosacral pedicle screw system124 reports
- LineSider LS-N6SA7550Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls1 report
- LineSider LS-N6SA6545Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls0 reports
- EXPEDIUM VERSE 199721000SMedos International Sàrl · Thoracolumbosacral pedicle screw system97 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.