Skip to content

TMINI™ Navigation Console 108757Recalls & reported problems

Think Surgical, Inc. · Class 2

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI M9461087570), received between Aug 28, 2024 and Feb 6, 2026 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Communication or Transmission Problem” (1 report).

1 other recall cover devices cleared under the same FDA submission (K230202, K232802, K241031); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem1
  • Communication or Transmission Problem1
  • Data Problem1
  • Insufficient Device Problem Information1
  • Malposition of Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Bone Notching1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionFeb 6, 2026Insufficient Device Problem Information; Data Problem

    THE SURGEON REQUESTED A POSTERIOR REFERENCING PLAN; HOWEVER, AN ANTERIOR REFERENCING PLAN WAS UPLOADED FOR THE PROCEDURE. THE DISCREPANCY WAS IDENTIFIED PRIOR TO THE PROCEDURE DURING THE CONFIRMATION OF INSTRUMENTS SCREEN ON THE TMINI NAVIGATION CONSOLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24280655 ↗
  • MalfunctionMay 30, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    THE SURGEON REPORTED FEMORAL NOTCHING WAS OBSERVED ON THE POST-OP X-RAY. THE SURGEON STATED THAT A NOTCH WARNING APPEARED ON SCREEN DURING GAP ASSESSMENT BUT WAS RESOLVED AFTER MAKING NECESSARY FEMORAL IMPLANT POSITION ADJUSTMENTS. HE NOTED THAT AFTER SAVING THE ADJUSTMENTS, THE EXPECTED CONFIRMATION SCREEN DID NOT APPEAR. THE CASE WAS COMPLETED ROBOTICALLY, AND A VISIBLE FEMORAL NOTCH WAS OBSERVED AFTER COMPLETING THE 4-IN-1 FEMORAL CUTS. THE NOTCH WAS CONSEQUENTLY CONFIRMED WITH THE POST-OP X-RAY.

    Patient problems: Bone Notching

    MAUDE report 22119531 ↗
  • MalfunctionMar 26, 2025Communication or Transmission Problem; Communication or Transmission Problem

    THE SURGEON REVIEWED THE SURGICAL PLAN DURING THE PROCEDURE AND MODIFIED THE POSITIONING OF THE FEMORAL COMPONENT. AFTER COMPLETING THE FEMORAL CUTS, THE SURGEON USED A CALIPER TO VERIFY THE CUTS BEFORE PROCEEDING WITH THE TIBIAL CUTS. ONCE THE FLEXION AND EXTENSION GAPS WERE CAPTURED, THE SURGEON ADJUSTED THE POSITIONING OF THE TIBIAL COMPONENT TO BALANCE THE GAPS. THE TIBIAL PINS WERE THEN PLACED CLOSER TOGETHER. AFTER FINISHING THE TIBIAL CUTS AND VERIFYING THEM WITH A CALIPER, IT WAS OBSERVED THAT THE ALIGNMENT APPEARED VALGUS RATHER THAN THE PLANNED VARUS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21702595 ↗
  • MalfunctionAug 28, 2024Malposition of Device; Malposition of Device

    IT WAS REPORTED THAT DURING SURGERY,THAT THE DISTAL MEDIAL RESECTION WAS LARGER THAN PLANNED. AS A RESULT IT APPERED THAT IMPLANT PLACEMENT WAS NOT CONSISTENT WITH THE APPROVED SURGICAL PLAN.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available

    MAUDE report 20092780 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K230202, K232802, K241031. One submission often covers many models, so check the product and lot details in each record.

  • Z-1318-2024 · Feb 12, 2024 · Open, Classified
    Assembly, Array Guide TMINI, REF: 109101, part of the TMINI Knee Instrument Set, used with the TMINI Miniature Robotic System

Other Orthopedic stereotaxic instrument devices

All Orthopedic stereotaxic instrument devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.