Affinity PixieRecalls & reported problems across all models
Medtronic, Inc. · 2 models in the FDA device registry
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls
- 1
We found 2 models of Affinity Pixie by Medtronic, Inc. in the FDA's device registry (GUDID). Together they are named in 14 adverse event reports in the FDA's MAUDE database, received between Aug 2, 2019 and Jan 21, 2026 — classified as 3 injuries, 11 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (11 reports).
1 FDA recall record name at least one of these models by its device identifier.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Affinity Pixie recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Feb 4, 2021Class ITerminated
Products being recalled due to potentially elevated level of bacterial endotoxin. Patients exposed to elevated levels of bacterial endotoxins may develop an acute inflammatory response.
Z-1191-2021 · Root cause: Device Design · Recall details · FDA record ↗
Most reported Affinity Pixie problems
Device problems
- Fluid/Blood Leak11
- Crack4
- Adverse Event Without Identified Device or Use Problem1
- Delivered as Unsterile Product1
- Gas Output Problem1
- Increase in Pressure1
- Material Puncture/Hole1
- Obstruction of Flow1
Patient problems
- No Clinical Signs, Symptoms or Conditions10
- Hemorrhage/Blood Loss/Bleeding4
- No Consequences Or Impact To Patient1
Affinity Pixie models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Affinity Pixie™ BBP241RESERVOIR, BLOOD, CARDIOPULMONARY BYPASS1 recall2 reports
- Affinity Pixie™ BBP211OXYGENATOR, CARDIOPULMONARY BYPASS12 reports
Recent Affinity Pixie adverse event reports
The 14 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionJan 21, 2026Delivered as Unsterile Product
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A PIXIE OXYGENATOR AND PIXIE CARDIOTOMY VENOUS RESERVOIR, IT WAS REPORTED THAT THE PACKAGING OF THE DEVICE WAS IN POOR CONDITION AND THE INTERNAL PACKAGING WAS BROKEN. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THERE WAS DAMAGE TO THE PACKAGING WAS EXTERNAL AND INTERNAL AND THE STERILE POUCH IS ALSO COMPROMISED. THERE WERE NO MISSING OR WRONG DEVICES OR COMPONENTS IN THE PACKAGE. THE STERILE PACKAGING WAS DAMAGED BUT STILL SEALED. THE DEVICE WAS SEALED/LOCATED IN THE 'SEAL'.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24133008 ↗MalfunctionMay 7, 2024Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF AN AFFINITY PIXIE OXYGENATOR, IT WAS REPORTED THAT A LEAK WAS NOTED FROM THE TOP OF THE DEVICE. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THERE WAS NO DAMAGE TO THE DEVICE, THE PACKAGING OR OTHER CONTENTS WITHIN THE PACKAGE. THE CUSTOMER STATED THAT THE LEAK WAS FOUND DURING WATER PRIMING. THERE WAS NO BLOOD LOSS AND NO TRANSFUSION REQUIRED AS THE LEAK OCCURRED DURING PRIMING. THERE WAS NO AIR IN THE SYSTEM/TUBING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19257979 ↗MalfunctionJul 11, 2023Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A PIXIE OXYGENATOR, THE CUSTOMER REPORTED A BLOOD LEAK. THE DEVICE WAS REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NO ADVERSE PATIENT EFFECT ASSOCIATED WITH THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17297280 ↗MalfunctionJun 7, 2023Fluid/Blood Leak; Material Puncture/Hole; Fluid/Blood Leak; Material Puncture/Hole
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF THE AFFINITY PIXIE OXYGENATORS, IT WAS REPORTED BY A RESEARCHER THAT THE OXYGENATORS USED IN THEIR RESEARCH PROJECTS HAD MECHANICAL FAILURES WHEN USED WITHIN INLINE PERFUSION SYSTEMS FOR PIG LIVER AND PIG KIDNEY. THE FLOW RATES WERE BELOW 300 MMHG AND BETWEEN 30-60 MINUTES AFTER BEGINNING THE PROCESS, THE DEVICE WOULD FORM A CRACK AND CAUSE OUTWARD LEAKING. THE DEVICES WERE REPLACED TO COMPLETE THE PROCEDURES. THERE WERE NO ADVERSE PATIENT EFFECTS ASSOCIATED WITH THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE CUSTOMER IS PART OF A RESEARCH GROUP WORKING ON PIG AND HUMAN LIVER PERFUSIONS. THE BREAKAGE OF ALL THE DEVICES WERE NOTICED WITHIN THE SAME WEEK OF EXPERIMENTS, BUT THE EXACT DATES ARE NOT AVAILABLE. THE CUSTOMER STATED THAT NO BLOOD WAS USED BUT THE SOLUTION USED DID LEAK SIGNIFICANTLY THROUGHOUT THE PERFUSION. THE CUSTOMER STATED THAT THE DEFECT WAS NOT VISIBLE, ONLY THROUGH THE LEAKAGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17076769 ↗MalfunctionMay 30, 2023Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT PRIOR TO USE OF A PIXIE OXYGENATOR, IT WAS REPORTED THAT THERE WAS LEAK FROM THE TOP OF THE OXYGENATOR WHILE PRIMING. THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT, SO NO ADVERSE EFFECT OCCURRED. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THERE WAS NO DAMAGE TO THE DEVICE, THE PACKAGING OR OTHER CONTENTS WITHIN THE PACKAGE (EX. DEVICES IN A TRAY OR CUSTOM PACK). THE CIRCUIT WAS PRIMED WITH BLOOD.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17019228 ↗MalfunctionApr 20, 2023Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF AN AFFINITY PIXIE OXYGENATOR, IT WAS REPORTED BY A RESEARCHER THAT THE OXYGENATORS USED IN THEIR RESEARCH PROJECTS HAD MECHANICAL FAILURES WHEN USED WITHIN INLINE PERFUSION SYSTEMS FOR PIG LIVER AND PIG KIDNEY. THE FLOW RATES WERE BELOW 300 MMHG AND BETWEEN 30-60 MINUTES AFTER BEGINNING THE PROCESS, THE DEVICE WOULD FORM A CRACK AND CAUSE OUTWARD LEAKING. THE DEVICES WERE REPLACED TO COMPLETE THE PROCEDURES. THERE WERE NO ADVERSE PATIENT EFFECTS ASSOCIATED WITH THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE CUSTOMER IS PART OF A RESEARCH GROUP WORKING ON PIG AND HUMAN LIVER PERFUSIONS. THE BREAKAGE OF ALL THE DEVICES WERE NOTICED WITHIN THE SAME WEEK OF EXPERIMENTS, BUT THE EXACT DATES ARE NOT AVAILABLE. THE CUSTOMER STATED THAT NO BLOOD WAS USED BUT THE SOLUTION USED DID LEAK SIGNIFICANTLY THROUGHOUT THE PERFUSION. THE CUSTOMER STATED THAT THE DEFECT WAS NOT VISIBLE, ONLY THROUGH THE LEAKAGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16775174 ↗MalfunctionApr 20, 2023Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF AN AFFINITY PIXIE OXYGENATOR, IT WAS REPORTED BY A RESEARCHER THAT THE OXYGENATORS USED IN THEIR RESEARCH PROJECTS HAD MECHANICAL FAILURES WHEN USED WITHIN INLINE PERFUSION SYSTEMS FOR PIG LIVER AND PIG KIDNEY. THE FLOW RATES WERE BELOW 300 MMHG AND BETWEEN 30-60 MINUTES AFTER BEGINNING THE PROCESS, THE DEVICE WOULD FORM A CRACK AND CAUSE OUTWARD LEAKING. THE DEVICES WERE REPLACED TO COMPLETE THE PROCEDURES. THERE WERE NO ADVERSE PATIENT EFFECTS ASSOCIATED WITH THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE CUSTOMER IS PART OF A RESEARCH GROUP WORKING ON PIG AND HUMAN LIVER PERFUSIONS. THE BREAKAGE OF ALL THE DEVICES WERE NOTICED WITHIN THE SAME WEEK OF EXPERIMENTS, BUT THE EXACT DATES ARE NOT AVAILABLE. THE CUSTOMER STATED THAT NO BLOOD WAS USED BUT THE SOLUTION USED DID LEAK SIGNIFICANTLY THROUGHOUT THE PERFUSION. THE CUSTOMER STATED THAT THE DEFECT WAS NOT VISIBLE, ONLY THROUGH THE LEAKAGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16775162 ↗MalfunctionApr 20, 2023Crack; Fluid/Blood Leak; Crack; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF AN AFFINITY PIXIE OXYGENATOR, IT WAS REPORTED BY A RESEARCHER THAT THE OXYGENATORS USED IN THEIR RESEARCH PROJECTS HAD MECHANICAL FAILURES WHEN USED WITHIN INLINE PERFUSION SYSTEMS FOR PIG LIVER AND PIG KIDNEY. THE FLOW RATES WERE BELOW 300 MMHG AND BETWEEN 30-60 MINUTES AFTER BEGINNING THE PROCESS, THE DEVICE WOULD FORM A CRACK AND CAUSE OUTWARD LEAKING. THE DEVICES WERE REPLACED TO COMPLETE THE PROCEDURES. THERE WERE NO ADVERSE PATIENT EFFECTS ASSOCIATED WITH THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE CUSTOMER IS PART OF A RESEARCH GROUP WORKING ON PIG AND HUMAN LIVER PERFUSIONS. THE BREAKAGE OF ALL THE DEVICES WERE NOTICED WITHIN THE SAME WEEK OF EXPERIMENTS, BUT THE EXACT DATES ARE NOT AVAILABLE. THE CUSTOMER STATED THAT NO BLOOD WAS USED BUT THE SOLUTION USED DID LEAK SIGNIFICANTLY THROUGHOUT THE PERFUSION. THE CUSTOMER STATED THAT THE DEFECT WAS NOT VISIBLE, ONLY THROUGH THE LEAKAGE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16775140 ↗MalfunctionDec 8, 2022Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A PIXIE OXYGENATOR, IT WAS REPORTED THAT A LEAK OCCURRED. THE DEVICE WAS REPLACED, BUT THE REST OF THE CUSTOM TUBING PACK WAS USED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT IMPACT ASSOCIATED WITH THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT EVENT OCCURRED AT THE BEGINNING OF THE PROCEDURE, WHILE ON BYPASS, BUT BEFORE X-CLAMP. THE AMOUNT OF PATIENT BLOOD LOSS AS A RESULT OF THIS LEAK WAS BETWEEN 5 TO 10ML. A TRANSFUSION WAS NOT REQUIRED AS A RESULT OF THIS LEAK. THERE WAS NO DAMAGE TO THE DEVICE, THE PACKAGING OR OTHER CONTENTS WITHIN THE PACKAGE. THERE WAS NO AIR IN THE SYSTEM/TUBING AND NO ADVERSE PATIENT EFFECTS.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 15938213 ↗InjuryOct 21, 2022Gas Output Problem; Increase in Pressure; Obstruction of Flow; Gas Output Problem; Increase in Pressure; Obstruction of Flow
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A PIXIE OXYGENATOR, IT WAS REPORTED THAT THE CLINICIAN NOTED THERE WAS A DIFFERENCE BETWEEN THE 'DIALED IN' PUMP FLOW ON THE QUANTUM HEART AND LUNG MACHINE (HLM) CONTROLLER AND THAT ON THE ARTERIAL FLOW PROBE WHICH WAS LOCATED BETWEEN THE OXYGENATOR AND PATIENT. THE CLINICIAN WAS CONCERNED THERE WAS SOMETHING WRONG WITH BLOOD FLOW DELIVERY AND INCREASED THE OCCLUSION ON THE ROLLER PUMP. THEY FELT THE TUBING EXITING THE ROLLER PUMP (BETWEEN THE PUMP AND OXYGENATOR) AND DESCRIBED IT AS 'ROCK SOLID' TO THE TOUCH INDICATING HIGH PRESSURES. PRE-OXYGENATOR PRESSURES ARE NOT MEASURED. ON INCREASING FIO2 AS PER PROTOCOL FOR CYANOTIC CHILDREN, THE OXYGENATION DID NOT INCREASE AS EXPECTED AND THE CEREBRAL OXIMETRY STARTED TO DECREASE FROM 80 TO 40% (MASSIMO NIRS). INCREASING THE PUMP FLOW DID NOT RESULT IN INCREASED FLOW ON THE FLOW PROBE CLOSEST TO THE PATIENT AND MANUALLY PALPATED PRESSURE REMAINED GREAT IN THE TUBING. THE PATIENT WAS VENTILATED BY THE ANESTHETIST, BYPASS WAS TERMINATED AND THE OXYGENATOR WAS CHANGED OUT. BYPASS RECOMMENCED, THE TWO FLOWS MATCHED, PRESSURE IN THE TUBING WAS NORMAL AND OXYGENATION WAS WITHIN THE EXPECTED RAN
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15648383 ↗MalfunctionOct 20, 2022Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A PIXIE OXYGENATOR, IT WAS REPORTED THAT THERE WAS A BLOOD LEAK NOTICED FROM THE TOP OF THE OXYGENATOR DURING CARDIOPULMONARY BYPASS FOR A PATIENT WITH CONGENITAL HEART DISEASE. THE DEVICE WAS REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NO ADDITIONAL PATIENT IMPACT ASSOCIATED WITH THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THERE WAS NO DAMAGE TO THE DEVICE, THE PACKAGING OR OTHER CONTENTS WITHIN THE PACKAGE (E.G. DEVICES IN A TRAY OR CUSTOM PACK). THE AMOUNT OF PATIENT BLOOD LOSS AND WHETHER A TRANSFUSION WAS REQUIRED WERE REQUESTED BUT THIS INFORMATION COULD NOT BE PROVIDED.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 15638713 ↗MalfunctionJul 21, 2022Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A PIXIE OXYGENATOR, IT WAS REPORTED THAT DURING EXTRACORPOREAL CIRCULATION, THE OXYGENATOR OOZED BLOOD. THE DEVICE WAS REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NO PATIENT IMPACT ASSOCIATED WITH THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT THE LOCATION OF THE LEAK WAS FROM ONE OF OXYGENATOR'S PORTS. THE CUSTOMER STATED THAT THERE WAS A SMALL AMOUNT OF BLOOD LOSS BUT NO SPECIFIC STATISTICS AND NO TRANSFUSION WAS REQUIRED AS A RESULT OF THIS LEAK.
Patient problems: Hemorrhage/Blood Loss/Bleeding
MAUDE report 15075307 ↗InjuryApr 21, 2022Fluid/Blood Leak; Fluid/Blood Leak
MEDTRONIC RECEIVED INFORMATION THAT DURING USE OF A PIXIE OXYGENATOR, THE CUSTOMER REPORTED THAT AFTER FOR 20 MINUTES ON CARDIOPULMONARY BYPASS, THE OXYGENATOR APPEARED TO SEEP BLOOD, AND THE DEVICE COULD NO LONGER BE USED. THE DEVICE WAS REPLACED TO COMPLETE THE PROCEDURE. THERE WAS NO REPORTED PATIENT IMPACT ASSOCIATED WITH THIS EVENT. MEDTRONIC RECEIVED ADDITIONAL INFORMATION THAT IT IS STILL INCONCLUSIVE WHETHER THE ISSUE WILL AFFECT THE PATIENT, AND THE POSSIBILITIES SUCH AS INFECTION AND WATER INFLOW TO THE OXYGENATOR DURING THE REPLACEMENT PROCESS CANNOT BE RULED OUT. THE AMOUNT OF PATIENT BLOOD LOSS AS A RESULT OF THIS LEAK WAS ASKED BUT UNKNOWN. A TRANSFUSION WAS REQUIRED AS A RESULT OF THIS LEAK. THE BLOOD LEAK OCCURRED FROM THE BOTTOM OF THE OXYGENATOR. THERE WAS NO DAMAGE TO THE PACKAGING OR THE DEVICE IDENTIFIED.
Patient problems: Hemorrhage/Blood Loss/Bleeding; No Clinical Signs, Symptoms or Conditions
MAUDE report 14165666 ↗InjuryAug 2, 2019Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
MEDTRONIC RECEIVED INFORMATION THAT DURING USE, THE CUSTOMER REPORTED DECREASED OXYGEN WHILE USING THE PIXIE OXYGENATOR. THE OXY WAS USED TO COMPLETE THE PROCEDURE WITH THE AID OF ANESTHESIA VENTILATION TO INCREASE OXYGENATION. THERE WAS NO ADVERSE PATIENT EFFECTS. NO PERFUSION RECORDS AVAILABLE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8854903 ↗
Other Medtronic, Inc. products
- CapSureFix Novus MRI SureScan29 models · 87,337 reports
- Medtronic Sprint Quattro Lead67 models · 45,880 reports · 1 recall
- Intellis AdaptiveStim3 models · 35,469 reports
- SYNCHROMED II11 models · 32,260 reports · 4 recalls
- CAPSUREFIX NOVUS17 models · 27,547 reports
- INTERSTIM30 models · 24,415 reports · 1 recall
- Sprint Quattro Secure S MRI SureScan7 models · 23,549 reports
- Medtronic LINQ Heart Monitor11 models · 19,366 reports · 3 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page