Adverse event reports
19,366
Deaths reported
3
Injuries reported
987
Recalls
3

We found 11 models of Medtronic LINQ Heart Monitor by Medtronic, Inc. in the FDA's device registry (GUDID). Together they are named in 19,366 adverse event reports in the FDA's MAUDE database, received between Dec 10, 2014 and Sep 30, 2026 — classified as 3 deaths, 987 injuries, 18,376 malfunctions.

The most frequently reported specific device problem is “Under-Sensing” (5,672 reports).

3 FDA recall records name at least one of these models by its device identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 153 reportsDec 2023: 99 reportsJan 2024: 177 reportsFeb 2024: 165 reportsMar 2024: 174 reportsApr 2024: 156 reportsMay 2024: 323 reportsJun 2024: 137 reportsJul 2024: 185 reportsAug 2024: 158 reportsSep 2024: 188 reportsOct 2024: 259 reportsNov 2024: 180 reportsDec 2024: 46 reportsJan 2025: 203 reportsFeb 2025: 169 reportsMar 2025: 164 reportsApr 2025: 105 reportsMay 2025: 201 reportsJun 2025: 214 reportsJul 2025: 208 reportsAug 2025: 361 reportsSep 2025: 109 reportsOct 2025: 184 reportsNov 2025: 202 reportsDec 2025: 147 reportsJan 2026: 166 reportsFeb 2026: 142 reportsMar 2026: 160 reportsApr 2026: 282 reportsMay 2026: 122 reportsJun 2026: 137 reportsJul 2026: 109 reportsAug 2026: 112 reportsSep 2026: 81 reportsOct 2026: 0 reports202420252026361
Last 36 months. The most recent month may be incomplete.

Medtronic LINQ Heart Monitor recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Nov 3, 2023Open, Classified

    It may create the potential for amplified noise and/or overall signal reduction of the ICM, which may interfere with intended recordings of heart rhythms. This noise pattern is different from occasional noise due to device position/migration, patient activity, or external electromagnetic interference.

    Z-0421-2024 · Root cause: Process change control · names 3 of these models · Recall details · FDA record ↗

  2. Jan 18, 2022Terminated

    Devices may lose functionality due to susceptibility to moisture ingress.

    Z-0782-2022 · Root cause: Nonconforming Material/Component · names 2 of these models · Recall details · FDA record ↗

  3. Jun 1, 2021Open, Classified

    Reveal LINQ with TruRhythm ICMs, that undergo a partial electrical reset appear to be programmed "ON", but are no longer able to detect and report Brady and Pause events to clinicians.

    Z-2102-2021 · Root cause: Software design · names 2 of these models · Recall details · FDA record ↗

Most reported Medtronic LINQ Heart Monitor problems

Device problems

  • Under-Sensing5,672
  • Over-Sensing3,912
  • Reset Problem2,269
  • Signal Artifact/Noise1,943
  • Premature Discharge of Battery1,652
  • Device Sensing Problem1,573
  • Inappropriate or Unexpected Reset1,172
  • Output Problem970

Patient problems

  • No Clinical Signs, Symptoms or Conditions10,432
  • No Known Impact Or Consequence To Patient6,871
  • No Consequences Or Impact To Patient951
  • Unspecified Infection439
  • Pain285
  • Erosion267
  • Discomfort125
  • Erythema94

Medtronic LINQ Heart Monitor models

11 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent Medtronic LINQ Heart Monitor adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Medtronic, Inc. products

All Medtronic, Inc. devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page