Alaris PCA Module, Model 8120Recalls & reported problems across all models
Carefusion 303, Inc. · 2 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
2 open recalls name at least one Alaris PCA Module, Model 8120 model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 918
- Deaths reported
- 1
- Injuries reported
- 2
- Recalls
- 2
We found 2 models of Alaris PCA Module, Model 8120 by Carefusion 303, Inc. in the FDA's device registry (GUDID). Together they are named in 918 adverse event reports in the FDA's MAUDE database, received between Dec 22, 2023 and Sep 30, 2026 — classified as 1 death, 2 injuries, 914 malfunctions.
The most frequently reported specific device problem is “Break” (257 reports).
2 FDA recall records name at least one of these models by its device identifier, 2 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Alaris PCA Module, Model 8120 recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Aug 15, 2024Open, Classified
Due to affected devices that may have bent female IUI connector and may not be operable and potentially cause delays in the start of the infusion or monitoring.
Z-3139-2024 · Root cause: Process control · names 2 of these models · Recall details · FDA record ↗
- Sep 15, 2023Open, Classified
Alaris PCA Module 8120, Patient Controlled Analgesia infusion pump: compatible syringes labeling contains syringes that have not been validated.
Z-0184-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported Alaris PCA Module, Model 8120 problems
Device problems
- Break257
- Calibration Problem172
- Device Alarm System103
- Inaccurate Delivery97
- Failure to Read Input Signal68
- Device Sensing Problem63
- No Apparent Adverse Event47
- Electrical /Electronic Property Problem43
Patient problems
- No Clinical Signs, Symptoms or Conditions894
- Insufficient Information25
- Pain2
- Chest Pain1
- Decreased Level of Consciousness1
- Hallucination1
- Loss of consciousness1
Alaris PCA Module, Model 8120 models
2 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Alaris™ PCA Module, Model 8120Pump, infusion, pca2 recalls1 death917 reports
- Alaris™ PCA Module, Model 8120Pump, infusion1 recall1 report
Recent Alaris PCA Module, Model 8120 adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionSep 30, 2026Break
It was reported that the device had missing claws. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50945191 ↗MalfunctionSep 30, 2026Excess Flow or Over-Infusion
It was reported the PCA continued to infuse following an alarm being triggered. The device continued to infuse at the programmed dose and rate. There was patient involvement; however, outcome is unknown.
Patient problems: Insufficient Information
MAUDE report 50943542 ↗MalfunctionSep 30, 2026Naturally Worn
It was reported that the device had broken wiper assembly. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50943537 ↗MalfunctionSep 29, 2026Electrical /Electronic Property Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50935968 ↗MalfunctionSep 29, 2026Circuit Failure
It was reported that a failure was observed during a planned preventive maintenance or recall remediation service event. There was no reported patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50932694 ↗MalfunctionSep 28, 2026Inadequate or Insufficient Training
It was reported that the 8120 Error 358.2000. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50913809 ↗MalfunctionSep 25, 2026Inaccurate Delivery
It was reported that the Patient Controlled Analgesia Pump Module was allegedly displaying a Volume to be Infused of 33mL or other varying volumes rather than the expected 30mL on a Hydromorphone infusion. However, when the customer attempted to recreate the scenario, they were unable to reproduce the issue. There was patient involvement, but no harm.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50894665 ↗MalfunctionSep 24, 2026Break; Deformation Due to Compressive Stress
It was reported that the device had plunger claw is broken. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50883633 ↗MalfunctionSep 24, 2026Device Alarm System
It was reported that the device had error code 358.2000. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50881314 ↗MalfunctionSep 24, 2026Break
It was reported that the device had unit is unable to be attached to a PCU. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50881311 ↗MalfunctionSep 24, 2026Device Alarm System
It was reported that the device had error code 356.6710. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50880522 ↗MalfunctionSep 24, 2026Break
It was reported that the device plunger sensor only reads closed and not open also the flange sensor seems to work every other turn. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50880514 ↗MalfunctionSep 24, 2026Calibration Problem
It was reported that the device had failed binary switches. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50879992 ↗MalfunctionSep 24, 2026Break
It was reported that the PCA Pump was in a vandalism situation and it appeared there was an object inside the locking mechanism.Customer provided additional information indicating:Only the locking key got damaged , and the individual could not access the drug. The hospital Biomed later accessed the drug syringe while the Director of Pharmacy observed them open the plastic cover. forcing the plastic lever to open. There was no patient Impact.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50877442 ↗MalfunctionSep 23, 2026Break
It was reported that the device had failed binary switches. There was no patient involvement.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50868105 ↗
Other Carefusion 303, Inc. products
- BD Alaris Infusion System29 models · 1,242,913 reports · 16 recalls
- Alaris25 models · 1,236,345 reports · 16 recalls
- BD Pyxis26 models · 81,193 reports · 66 recalls
- Pyxis5 models · 7,862 reports · 16 recalls
- BD Alaris PCU Model 80153 models · 4,071 reports · 2 recalls
- BD Alaris Syringe Module, Model 81102 models · 3,183 reports · 2 recalls
- BD Alaris PCU Module, Model 80152 models · 3,078 reports · 3 recalls
- BD Alaris ETCO2 Module, Model 83002 models · 956 reports
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.