Skip to content

AlarisRecalls & reported problems across all models

Carefusion 303, Inc. · 25 models in the FDA device registry

Data updated Oct 18, 2026 from official FDA records · Editorial policy

14 open recalls name at least one Alaris model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
1,236,345
Deaths reported
146
Injuries reported
1,136
Recalls
16

We found 25 models of Alaris by Carefusion 303, Inc. in the FDA's device registry (GUDID). Together they are named in 1,236,345 adverse event reports in the FDA's MAUDE database, received between Oct 23, 2017 and Sep 30, 2026 — classified as 146 deaths, 1,136 injuries, 1,235,063 malfunctions.

The most frequently reported specific device problem is “Crack” (710,671 reports).

16 FDA recall records name at least one of these models by its device identifier, 14 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 4,586 reportsDec 2023: 3,096 reportsJan 2024: 2,108 reportsFeb 2024: 2,737 reportsMar 2024: 2,132 reportsApr 2024: 2,101 reportsMay 2024: 1,519 reportsJun 2024: 1,647 reportsJul 2024: 1,178 reportsAug 2024: 1,073 reportsSep 2024: 1,659 reportsOct 2024: 1,484 reportsNov 2024: 1,815 reportsDec 2024: 1,049 reportsJan 2025: 1,064 reportsFeb 2025: 811 reportsMar 2025: 917 reportsApr 2025: 654 reportsMay 2025: 714 reportsJun 2025: 663 reportsJul 2025: 760 reportsAug 2025: 847 reportsSep 2025: 564 reportsOct 2025: 439 reportsNov 2025: 571 reportsDec 2025: 487 reportsJan 2026: 335 reportsFeb 2026: 452 reportsMar 2026: 526 reportsApr 2026: 508 reportsMay 2026: 438 reportsJun 2026: 288 reportsJul 2026: 262 reportsAug 2026: 249 reportsSep 2026: 314 reportsOct 2026: 0 reports2024202520264,586
Last 36 months. The most recent month may be incomplete.

Alaris recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Feb 11, 2026Class IIOpen, Classified

    Due to product labeling not indicating which Transport Layer Security (TLS) version that must be enabled on hospital network to ensure secure Wi-Fi communication with Point-of-Care Units

    Z-1645-2026 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  2. Oct 17, 2025Class IOpen, Classified

    If infusion pump is dropped or severely jarred this may damage the pump module bezel assembly, which can cause under-infusion, over-infusion, unregulated flow, or pump module failure to calibrate; so dropped/jarred pumps should be removed from use, tested, and inspected by qualified service personnel prior to reuse; tip sheet, cleaning and disinfecting procedure and reference guide updated.

    Z-0430-2026 · Root cause: Device Design · names 3 of these models · Recall details · FDA record ↗

  3. Jul 8, 2025Class IOpen, Classified

    Infusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion bolos volume.

    Z-2179-2025 · Root cause: Under Investigation by firm · names 3 of these models · Recall details · FDA record ↗

  4. Feb 18, 2025Class IOpen, Classified

    Software issue that may result in outdated automated programming request(APR) being sent to the progressive care unit (PCU).

    Z-1304-2025 · Root cause: Under Investigation by firm · names 2 of these models · Recall details · FDA record ↗

  5. Aug 15, 2024Class IIOpen, Classified

    Due to affected devices that may have bent female IUI connector and may not be operable and potentially cause delays in the start of the infusion or monitoring.

    Z-3142-2024 · Root cause: Process control · names 2 of these models · Recall details · FDA record ↗

  6. Aug 15, 2024Class IIOpen, Classified

    Due to affected devices that may have bent female IUI connector and may not be operable and potentially cause delays in the start of the infusion or monitoring.

    Z-3141-2024 · Root cause: Process control · names 4 of these models · Recall details · FDA record ↗

  7. Aug 15, 2024Class IIOpen, Classified

    Due to affected devices that may have bent female IUI connector and may not be operable and potentially cause delays in the start of the infusion or monitoring.

    Z-3140-2024 · Root cause: Process control · names 2 of these models · Recall details · FDA record ↗

  8. Aug 15, 2024Class IIOpen, Classified

    Due to affected devices that may have bent female IUI connector and may not be operable and potentially cause delays in the start of the infusion or monitoring.

    Z-3139-2024 · Root cause: Process control · names 4 of these models · Recall details · FDA record ↗

  9. Aug 15, 2024Class IIOpen, Classified

    Due to affected devices that may have bent female IUI connector and may not be operable and potentially cause delays in the start of the infusion or monitoring.

    Z-3138-2024 · Root cause: Process control · names 3 of these models · Recall details · FDA record ↗

  10. May 21, 2024Class IIOpen, Classified

    Due to a software issue the PC unit may not connect to the server which could impact wireless data transmission to and from the server.

    Z-2262-2024 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  11. Sep 15, 2023Class IOpen, Classified

    Alaris PCA Module 8120, Patient Controlled Analgesia infusion pump: compatible syringes labeling contains syringes that have not been validated.

    Z-0186-2024 · Root cause: Under Investigation by firm · names 3 of these models · Recall details · FDA record ↗

  12. Sep 15, 2023Class IOpen, Classified

    Alaris PCA Module 8120, Patient Controlled Analgesia infusion pump: compatible syringes labeling contains syringes that have not been validated.

    Z-0185-2024 · Root cause: Under Investigation by firm · names 3 of these models · Recall details · FDA record ↗

  13. Sep 15, 2023Class IOpen, Classified

    Alaris PCA Module 8120, Patient Controlled Analgesia infusion pump: compatible syringes labeling contains syringes that have not been validated.

    Z-0184-2024 · Root cause: Under Investigation by firm · names 2 of these models · Recall details · FDA record ↗

  14. Jan 24, 2023Class IIOpen, Classified

    Infusion pump PCs with specific software/network cards/IP addresses can have network/wireless connection failure. Could cause inability to receive new datasets and inability to transfer retrospective analytics data to connected systems. Users with interfaces to electronic medication record or admin discharge transfer, may be unable to receive/send data; infusions may need to be manually programmed

    Z-1192-2023 · Root cause: Software Design Change · names 2 of these models · Recall details · FDA record ↗

  15. Nov 17, 2017Class IITerminated

    The syringe plunger grippers may fail to automatically close around the syringe plunger press when the gripper control knob has been closed.

    Z-0323-2018 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

  16. Nov 17, 2017Class IITerminated

    The syringe plunger grippers may fail to automatically close around the syringe plunger press when the gripper control knob has been closed.

    Z-0322-2018 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗

Most reported Alaris problems

Device problems

  • Crack710,671
  • Corroded177,886
  • Break118,606
  • Display Difficult to Read82,807
  • Failure to Read Input Signal52,362
  • Device Sensing Problem44,126
  • Contamination24,945
  • Physical Resistance/Sticking18,964

Patient problems

  • Insufficient Information642,526
  • No Clinical Signs, Symptoms or Conditions516,921
  • No Patient Involvement76,521
  • No Consequences Or Impact To Patient2,262
  • No Information1,379
  • No Known Impact Or Consequence To Patient1,117
  • Low Blood Pressure/ Hypotension387
  • High Blood Pressure/ Hypertension77

Alaris models

25 models, most reported first. Each model has its own page with its reports, recalls and device details.