Alere Cholestech LDXRecalls & reported problems across all models
Alere San Diego, Inc. · 6 models in the FDA device registry
Data updated Oct 16, 2026 from official FDA records · Editorial policy
1 open recall names at least one Alere Cholestech LDX model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls
- 1
We found 6 models of Alere Cholestech LDX by Alere San Diego, Inc. in the FDA's device registry (GUDID). Together they are named in 1 adverse event report in the FDA's MAUDE database, received between Dec 11, 2020 and Dec 11, 2020 — classified as 1 malfunction.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (1 report).
1 FDA recall record name at least one of these models by its device identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Alere Cholestech LDX recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- Jun 24, 2022Open, Classified
Cholesterol test cassettes package insert is to be updated with new bilirubin interference information because it was found that samples containing bilirubin, at concentrations greater than 2.0 mg/dL (conjugated) or 1.6 mg/dL (unconjugated), can result in low recovery of Triglyceride (TRG) and High-Density Lipoproteins (HDL).
Z-1508-2022 · Root cause: Device Design · names 6 of these models · Recall details · FDA record ↗
Most reported Alere Cholestech LDX problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings1
Patient problems
- Insufficient Information1
Alere Cholestech LDX models
6 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Alere Cholestech LDX® 10-991Glucose Oxidase, Glucose1 recall1 report
- Alere Cholestech LDX® 10-987Enzymatic Esterase--Oxidase, Cholesterol1 recall0 reports
- Alere Cholestech LDX® 10-989Colorimetric Method, Triglycerides1 recall0 reports
- Alere Cholestech LDX® 10-990Glucose Oxidase, Glucose1 recall0 reports
- Alere Cholestech LDX® 99021Glucose Oxidase, Glucose1 recall0 reports
- Alere Cholestech LDX® 14-531Glucose Oxidase, Glucose1 recall0 reports
Recent Alere Cholestech LDX adverse event reports
The 1 most recent reports across all models, in the reporter's own words as published by the FDA.
MalfunctionDec 11, 2020Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
ON (B)(6) 2020 A PATIENT HAVING A WELLNESS CHECK HAD A DISCREPANCY WITH LDX RESULTS DISPLAYING N/A, SPECIFICALLY TC, NON HDL, TC/HDL AND LDL. THE TEST WAS RERUN WITH THE SAME SAMPLE AND LOT ALTHOUGH A DIFFERENT CASSETTE WITH DIFFERENT RESULTS. THERE WAS NO LAB/CONFIRMATORY TESTING PROVIDED. (ALL RESULTS IN MG/DL). THERE WAS NO ADVERSE EVENT REPORTED. ALTHOUGH REQUESTED, NO FURTHER INFORMATION HAS BEEN PROVIDED. TEST 1: TC:N/A; TRG:289; HDL:<15; NON-HDL: N/A; TC/HDL: N/A; LDL: N/A; GLUCOSE:178 TEST 2: TC:166; TRG: 186; HDL: 51; NON-HDL: NOT PROVIDED; TC/HDL: NOT PROVIDED; LDL: 78; GLUCOSE:64
Patient problems: Insufficient Information
MAUDE report 10993890 ↗
Other Alere San Diego, Inc. products
All Alere San Diego, Inc. devices →Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page