Skip to content

Alere Cholestech LDX® 10-990Recalls & reported problems

Alere San Diego, Inc. · Alere Cholestech LDX® TC HDL GLU · Class 2

All models of Alere Cholestech LDX (6)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
0
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

No MAUDE adverse event reports in our data reference this device identifier directly.

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Recalls naming this device

  1. Jun 24, 2022Class IIOpen, Classified

    Cholesterol test cassettes package insert is to be updated with new bilirubin interference information because it was found that samples containing bilirubin, at concentrations greater than 2.0 mg/dL (conjugated) or 1.6 mg/dL (unconjugated), can result in low recovery of Triglyceride (TRG) and High-Density Lipoproteins (HDL).

    Z-1508-2022 · Root cause: Device Design · Recall details · FDA record ↗

Other Glucose Oxidase, Glucose devices

All Glucose Oxidase, Glucose devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page