Amplia MRI Quad CRT-D SureScanRecalls & reported problems across all models
Medtronic, Inc. · 8 models in the FDA device registry
Data updated Oct 18, 2026 from official FDA records · Editorial policy
3 open recalls name at least one Amplia MRI Quad CRT-D SureScan model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1,423
- Deaths reported
- 32
- Injuries reported
- 916
- Recalls
- 6
We found 8 models of Amplia MRI Quad CRT-D SureScan by Medtronic, Inc. in the FDA's device registry (GUDID). Together they are named in 1,423 adverse event reports in the FDA's MAUDE database, received between Dec 9, 2016 and Sep 9, 2026 — classified as 32 deaths, 916 injuries, 475 malfunctions.
The most frequently reported specific device problem is “Premature Discharge of Battery” (201 reports).
6 FDA recall records name at least one of these models by its device identifier, 3 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works
Reports received per month, all models
Amplia MRI Quad CRT-D SureScan recalls
FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.
- May 10, 2023Class IOpen, Classified
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
Z-1764-2023 · Root cause: Under Investigation by firm · names 2 of these models · Recall details · FDA record ↗
- May 10, 2023Class IOpen, Classified
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
Z-1763-2023 · Root cause: Under Investigation by firm · names 2 of these models · Recall details · FDA record ↗
- Feb 3, 2021Class IOpen, Classified
A small percentage of implanted cardiac devices, from a well-defined subset, may experience a shortened Recommended Replacement Time (RRT) to End of Service (EOS) interval following an earlier-than-expected RRT observation
Z-1306-2021 · Root cause: Device Design · names 2 of these models · Recall details · FDA record ↗
- Dec 19, 2017Class ITerminated
Possible prevention of high and low voltage therapy in Medtronic Implantable Cardioverter Defibrillators (ICDs) and Cardiac Resynchronization Therapy Defibrillator (CRT-D).
Z-0582-2018 · Root cause: Process design · names 3 of these models · Recall details · FDA record ↗
- Apr 12, 2017Class IITerminated
An unexpected transient mode switch behavior was detected by Medtronic during systems testing that affects Amplia' "', Claria'"', and Compia'"' Quad CRT-D MRI devices.
Z-2117-2017 · Root cause: Software design · Recall details · FDA record ↗
- Dec 9, 2016Class IITerminated
During internal testing conducted as part of next generation product development, it was discovered that the AdaptivCRT (aCRT) feature of Amplia MRI and Claria MRI CRT -D devices may appear to be enabled when it is not active.
Z-0962-2017 · Root cause: Software Design Change · names 2 of these models · Recall details · FDA record ↗
Most reported Amplia MRI Quad CRT-D SureScan problems
Device problems
- Adverse Event Without Identified Device or Use Problem424
- Premature Discharge of Battery201
- Device Sensing Problem165
- Inappropriate/Inadequate Shock/Stimulation142
- Pacing Problem126
- Electrical /Electronic Property Problem109
- Appropriate Device Problem Term/Code Not Available69
- Mechanical Problem63
Patient problems
- No Clinical Signs, Symptoms or Conditions335
- Unspecified Infection305
- No Known Impact Or Consequence To Patient244
- Shock from Patient Lead(s)134
- No Consequences Or Impact To Patient79
- No Patient Involvement56
- Tachycardia35
- Bradycardia33
Amplia MRI Quad CRT-D SureScan models
8 models, most reported first. Each model has its own page with its reports, recalls and device details.
- Amplia MRI™ Quad CRT-D SureScan™ DTMB1QQDefibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)4 recalls18 deaths648 reports
- Amplia MRI™ Quad CRT-D SureScan™ DTMB1QQDefibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)1 recall5 deaths268 reports
- Amplia MRI™ Quad CRT-D SureScan™ DTMB1QQDefibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)1 recall2 deaths177 reports
- Amplia MRI™ Quad CRT-D SureScan™ DTMB1Q1Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)3 recalls2 deaths68 reports
- Amplia MRI™ Quad CRT-D SureScan™ DTMB1QQDefibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)1 recall3 deaths107 reports
- Amplia MRI™ Quad CRT-D SureScan™ DTMB1Q1Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)1 recall1 death78 reports
- Amplia MRI™ Quad CRT-D SureScan™ DTMB1Q1Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)1 recall1 death61 reports
- Amplia MRI™ Quad CRT-D SureScan™ DTMB1Q1Defibrillator, automatic implantable cardioverter, with cardiac resynchronization (CRT-D)16 reports
Recent Amplia MRI Quad CRT-D SureScan adverse event reports
The 15 most recent reports across all models, in the reporter's own words as published by the FDA.
InjurySep 9, 2026Adverse Event Without Identified Device or Use Problem
It was reported that the patient presented to the emergency department with sceptic shock due to Methicillin-resistant Staphylococcus aureus bacteremia, requiring vasopressor administration. Transesophageal Echocardiography (TEE) confirmed vegetation on the Right Ventricular (RV) lead and Right Atrial (RA) lead. Computerized tomography (CT) scan could not discern if the patient was experiencing pneumonia or sceptic pulmonary emboli. The Cardiac Resynchronization Therapy Defibrillator (CRT-D) system was explanted. No further patient complications have been reported as a result of this event.
Patient problems: Bacteremia; Drug Resistant Bacterial Infection; Medical device site infection; Pneumonia; Pulmonary Embolism; Sepsis; Septic Shock
MAUDE report 50691810 ↗MalfunctionSep 3, 2026Electromagnetic Interference; Pacing Problem
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50592016 ↗InjuryAug 24, 2026Defibrillation/Stimulation Problem; Device Sensing Problem
It was reported that the Cardiac Resynchronization Therapy Defibrillator (CRT-D) inappropriately withheld therapy for events that were misclassified as Ventricular Fibrillation (VF) by the Atrial Flutter (AF) by the device discriminator. The device discriminator was turned off. The device remains in use. No further patient complications have been reported as a result of this event.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50419005 ↗MalfunctionJul 7, 2026Over-Sensing
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) TRIGGERED A LEAD INTEGRITY ALERT (LIA) DUE TO ELEVATED SENSING INTEGRITY COUNTER (SIC) AND HIGH RATE NON-SUSTAINED EPISODES. THE PATIENT WAS SEEN IN CLINIC AND NO PRODUCT ISSUES WERE OBSERVED, THE PHYSICIAN NOTED THAT THE LIA WAS NOT OF CONCERN AS IT WAS TRIGGERED DUE TO THE PATIENT PLUGGING IN A DRYER THAT HAS A LIKELY SHORT. THE CRTD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25734802 ↗MalfunctionJul 2, 2026Device Sensing Problem; Data Problem
IT WAS REPORTED THAT THE RIGHT VENTRICULAR (RV) LEAD EXHIBITED T-WAVE OVERSENSING (TWOS) POST PACE, RESULTING IN INHIBITION/DELAY OF PACING. THE RV LEAD REMAINS IN USE. IT WAS ALSO REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) DETECTED SUPRA VENTRICULAR TACHYCARDIA (SVT) VERSUS A MONITORED VENTRICULAR TACHYCARDIA (VT) WITH 1 TO1 SUPRA VENTRICULAR TACHYCARDIA (SVT) ATRIAL EVENTS FALLING INTO THE BLANKING PERIOD. ADDITIONALLY, THE CRT-D DEVICE INTERROGATION REPORTED DISPLAYED AN OBSERVATION WHEN THE DEVICE FEATURE THAT AUTOMATICALLY MONITORS PACING THRESHOLDS AT PERIODIC INTERVALS IS PROGRAMMED OFF. THE CRT-D REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25696412 ↗InjuryJun 30, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) PATIENT HAD EXPERIENCED A STROKE. THE DEVICE REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Stroke/CVA
MAUDE report 25673680 ↗InjuryMay 27, 2026Pacing Problem; Inappropriate/Inadequate Shock/Stimulation; Device Sensing Problem
IT WAS REPORTED THAT THE PATIENT RECEIVED INAPPROPRIATE HIGH VOLTAGE THERAPY AND ANTI-TACHYCARDIA PACING (ATP) FOR SUPRA VENTRICULAR TACHYCARDIA (SVT) THAT WAS DETECTED AS VENTRICULAR TACHYCARDIA (VT). ADDITIONALLY, IT WAS NOTED THAT THE EPISODE DID NOT MEET DISCRIMINATOR CRITERIA; THEREFORE, THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) DELIVERED THERAPY PER PROGRAMMING. THE CRT-D REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Shock from Patient Lead(s)
MAUDE report 25306097 ↗InjuryMay 20, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE PATIENT WAS IN THE HOSPITAL FOR A PROCEDURE, WHICH WAS CANCELLED DUE TO A BLOOD CLOT IN THEIR HEART. THE PATIENT WAS PUT ON AN ANTICOAGULANT MEDICATION, AND A DAY LATER, THEY "FELT AN ITCHING FEELING INSIDE THE PACEMAKER." THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Embolism/Embolus; Itching Sensation
MAUDE report 25237497 ↗MalfunctionMay 19, 2026Failure to Interrogate; Battery Problem
THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) WAS RETURNED TO THE MANUFACTURER, ANALYZED, AND TESTED OUT OF SPECIFICATION. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25218738 ↗MalfunctionMay 18, 2026Premature Discharge of Battery; Battery Problem
IT WAS REPORTED BY THE PATIENTS REPRESENTATIVE THAT THE PATIENT WENT IN FOR AN MAGNETIC RESONANCE IMAGING (MRI) AND FOUND OUT THEIR CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) BATTERY WAS DEAD. THE CRT-D REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25205832 ↗InjuryApr 21, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS EXPLANTED DUE TO INFECTION. IT WAS NOTED THAT THE INFECTION ORGANISM WAS METHICILLIN-RESISTANT STAPHYLOCOCCUS AUREUS. THE PATIENT EXPERIENCED SEPSIS, FEVER, AND VEGETATION ON THE LEADS. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Drug Resistant Bacterial Infection; Fever; Medical device site infection; Sepsis
MAUDE report 24941928 ↗InjuryApr 9, 2026Premature Discharge of Battery
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) EXPERIENCED PREMATURE BATTERY DEPLETION. THE CRTD WAS EXPLANTED AND REPLACED. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24832383 ↗MalfunctionApr 8, 2026Electromagnetic Interference
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) EXHIBITED NOISE/OVERSENSING SUGGESTIVE OF ELECTROMAGNETIC INTERFERENCE (EMI). NOISE EPISODES ARE OCCURRING AROUND THE SAME TIME EACH DAY. THE CRT-D REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24818617 ↗InjuryApr 8, 2026Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) SYSTEM WAS EXPLANTED DUE TO A POCKET INFECTION. THERE WAS VEGETATION OBSERVED ON THE RIGHT VENTRICULAR (RV) LEAD, AND THE PATIENT HAD A POSITIVE CULTURE FOR KLEBSIELLA PNEUMONIAE. DURING EXPLANT, THE LEFT VENTRICULAR (LV) LEAD FRACTURED DISTALLY BEFORE THE ENTIRE LEAD COULD BE FULLY REMOVED. MOST OF THE LV LEAD WAS REMOVED EXCEPT FOR THE TIP, WHICH REMAINS IN THE PATIENT. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: Unspecified Infection
MAUDE report 24817263 ↗MalfunctionApr 2, 2026Device Sensing Problem
IT WAS REPORTED THAT THE CARDIAC RESYNCHRONIZATION THERAPY DEFIBRILLATOR (CRT-D) PATIENT WHILE RIDING ON A BUS ON A BUMPY ROAD THE CRTD EXHIBITED INTERMITTENT DETECTION FOR AN ARRYTHMIA. THE PATIENT WAS PUT ON MEDICATION. THE DEVICE REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24768234 ↗
Other Medtronic, Inc. products
- CapSureFix Novus MRI SureScan29 models · 87,337 reports
- Medtronic Sprint Quattro Lead67 models · 45,880 reports · 1 recall
- Intellis AdaptiveStim3 models · 35,469 reports
- SYNCHROMED II11 models · 32,260 reports · 4 recalls
- CAPSUREFIX NOVUS17 models · 27,547 reports
- INTERSTIM30 models · 24,415 reports · 1 recall
- Sprint Quattro Secure S MRI SureScan7 models · 23,549 reports
- Medtronic LINQ Heart Monitor11 models · 19,366 reports · 3 recalls
Sources and updates
- Models: FDA GUDID, file dated Oct 13, 2026. Models are grouped by the brand name the manufacturer registered.
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page