Skip to content

Arctic Sun 5000Recalls & reported problems across all models

Medivance, Inc. · 3 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names at least one Arctic Sun 5000 model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
11,704
Deaths reported
5
Injuries reported
680
Recalls
1

We found 3 models of Arctic Sun 5000 by Medivance, Inc. in the FDA's device registry (GUDID). Together they are named in 11,704 adverse event reports in the FDA's MAUDE database, received between Jun 29, 2017 and Sep 29, 2026 — classified as 5 deaths, 680 injuries, 11,019 malfunctions.

The most frequently reported specific device problem is “Insufficient Cooling” (2,047 reports).

1 FDA recall record name at least one of these models by its device identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 111 reportsDec 2023: 133 reportsJan 2024: 136 reportsFeb 2024: 146 reportsMar 2024: 114 reportsApr 2024: 105 reportsMay 2024: 100 reportsJun 2024: 84 reportsJul 2024: 105 reportsAug 2024: 120 reportsSep 2024: 156 reportsOct 2024: 94 reportsNov 2024: 78 reportsDec 2024: 64 reportsJan 2025: 76 reportsFeb 2025: 53 reportsMar 2025: 64 reportsApr 2025: 59 reportsMay 2025: 60 reportsJun 2025: 68 reportsJul 2025: 58 reportsAug 2025: 43 reportsSep 2025: 47 reportsOct 2025: 61 reportsNov 2025: 58 reportsDec 2025: 63 reportsJan 2026: 40 reportsFeb 2026: 63 reportsMar 2026: 55 reportsApr 2026: 54 reportsMay 2026: 32 reportsJun 2026: 11 reportsJul 2026: 9 reportsAug 2026: 0 reportsSep 2026: 11 reportsOct 2026: 0 reports202420252026156
Last 36 months. The most recent month may be incomplete.

Arctic Sun 5000 recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Jun 27, 2024Open, Classified

    If temperature management system fails to reach correct target water temperature while device is operating in patient control mode, due to sudden patient temperature changes/interruption in water flow/blockage of air flow by an obstruction or dirty filter, then system may not alert, and alarm absence may lead to hypothermia or hyperthermia, so a software update will be released to correct issue.

    Z-2457-2024 · Root cause: Software design · names 3 of these models · Recall details · FDA record ↗

Most reported Arctic Sun 5000 problems

Device problems

  • Insufficient Cooling2,047
  • Defective Component1,976
  • Failure to Pump1,591
  • Device Sensing Problem1,380
  • Insufficient Heating1,065
  • Gas/Air Leak873
  • Temperature Problem862
  • Restricted Flow rate859

Patient problems

  • No Clinical Signs, Symptoms or Conditions7,438
  • No Known Impact Or Consequence To Patient2,311
  • No Consequences Or Impact To Patient2,119
  • No Patient Involvement1,275
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available788
  • Alteration In Body Temperature481
  • Patient Problem/Medical Problem329
  • Chills160

Arctic Sun 5000 models

3 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent Arctic Sun 5000 adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Medivance, Inc. products

All Medivance, Inc. devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page