Skip to content

Arctic Sun StatRecalls & reported problems across all models

Medivance, Inc. · 2 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names at least one Arctic Sun Stat model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
1,717
Deaths reported
0
Injuries reported
37
Recalls
1

We found 2 models of Arctic Sun Stat by Medivance, Inc. in the FDA's device registry (GUDID). Together they are named in 1,717 adverse event reports in the FDA's MAUDE database, received between Oct 31, 2020 and Sep 26, 2026 — classified as 37 injuries, 1,680 malfunctions.

The most frequently reported specific device problem is “Insufficient Heating” (297 reports).

1 FDA recall record name at least one of these models by its device identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 35 reportsDec 2023: 33 reportsJan 2024: 28 reportsFeb 2024: 64 reportsMar 2024: 37 reportsApr 2024: 30 reportsMay 2024: 39 reportsJun 2024: 33 reportsJul 2024: 31 reportsAug 2024: 34 reportsSep 2024: 52 reportsOct 2024: 28 reportsNov 2024: 34 reportsDec 2024: 26 reportsJan 2025: 46 reportsFeb 2025: 42 reportsMar 2025: 33 reportsApr 2025: 23 reportsMay 2025: 39 reportsJun 2025: 34 reportsJul 2025: 34 reportsAug 2025: 28 reportsSep 2025: 36 reportsOct 2025: 39 reportsNov 2025: 26 reportsDec 2025: 28 reportsJan 2026: 33 reportsFeb 2026: 27 reportsMar 2026: 18 reportsApr 2026: 15 reportsMay 2026: 14 reportsJun 2026: 6 reportsJul 2026: 10 reportsAug 2026: 1 reportsSep 2026: 8 reportsOct 2026: 0 reports20242025202664
Last 36 months. The most recent month may be incomplete.

Arctic Sun Stat recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Jun 27, 2024Open, Classified

    If temperature management system fails to reach correct target water temperature while device is operating in patient control mode, due to sudden patient temperature changes/interruption in water flow/blockage of air flow by an obstruction or dirty filter, then system may not alert, and alarm absence may lead to hypothermia or hyperthermia, so a software update will be released to correct issue.

    Z-2457-2024 · Root cause: Software design · names 2 of these models · Recall details · FDA record ↗

Most reported Arctic Sun Stat problems

Device problems

  • Insufficient Heating297
  • Device Sensing Problem217
  • Inaccurate Flow Rate204
  • Defective Component170
  • Insufficient Cooling165
  • Gas/Air Leak161
  • Computer Operating System Problem146
  • Circuit Failure116

Patient problems

  • No Clinical Signs, Symptoms or Conditions1,580
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available117
  • Chills33
  • Convulsion/Seizure3
  • No Code Available3
  • Fever2
  • Insufficient Information1
  • No Consequences Or Impact To Patient1

Arctic Sun Stat models

2 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent Arctic Sun Stat adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Medivance, Inc. products

All Medivance, Inc. devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page